Skip to content

LUCEA LED® ARDLCA309004A (UDI 03700712400736)Recalls & reported problems

Maquet Sas · LCA 50 MOBILE · Class 2

All models of LUCEA LED (33)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03700712400736), received between Mar 17, 2023 and Nov 14, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K113679); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Peeled/Delaminated3
  • Crack2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2024Crack; Peeled/Delaminated; Detachment of Device or Device Component; Crack; Peeled/Delaminated; Detachment of Device or Device Component

    MANUFACTURER'S REFERENCE NUMBER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20681700 ↗
  • MalfunctionNov 23, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    MANUFACTURER'S REFERENCE NUMBER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18197993 ↗
  • MalfunctionSep 19, 2023Crack; Peeled/Delaminated; Detachment of Device or Device Component; Crack; Peeled/Delaminated; Detachment of Device or Device Component

    ON (B)(6), 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - LUCEA 50. BASED ON PHOTOGRAPHIC EVIDENCE THE HEADLIGHT COVER WAS CRACKED WITH MISSING PARTICLES, THE HEADLIGHT WAS HANGING ON CABLES AND THE PAINT WAS CHIPPING FROM THE ARM. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS OR PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17771401 ↗
  • MalfunctionMar 17, 2023Peeled/Delaminated; Peeled/Delaminated

    ON 17TH FEBRUARY 2023 GETINGE BECAME AWARE OF A NON-REPORTABLE ISSUE WITH ONE OF OUR SURGICAL LIGHTS - LUCEA 50. ON 13TH MARCH 2023, SERVICE TECHNICIAN ARRIVED ON SITE AND DETECTED REPORTABLE ISSUE. ACCORDING TO PHOTOGRAPHIC EVIDENCE, THE PAINT WAS DAMAGED ON THE SPRING ARM AND PARTICLES WERE MISSING. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PAINT PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16563319 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113679. One submission often covers many models, so check the product and lot details in each record.

  • Z-0759-2024 · Nov 8, 2023 · Open, Classified
    Maquet Lucea - Lucea10/40, Lucea50/100 OR Light Systems, Model Numbers and Catalog Numbers LCA 10 - ARD2LCA00040C, LCA 10 - ARDLCA109000C, LCA 10 - ARDLCA109002C, LCA 10 - ARDLCA10

Other Light, surgical, floor standing devices

All Light, surgical, floor standing devices →

Sources and updates