Maquet Sterigrip ARD567203900Recalls & reported problems
Maquet Sas · STG PSX 01 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 03700712405120), received between Jan 18, 2024 and Jan 22, 2024 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Crack” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Crack12
- Detachment of Device or Device Component12
Patient problems
- No Clinical Signs, Symptoms or Conditions12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 22, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
ON 9TH JANUARY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL EQUIPMENT - MAQUET STERIGRIP. IT WAS STATED THE MOUNTING WAS BROKEN WITH MISSING PARTICLES ON STERIGRIP, AND AS A RESULT THERE WAS A RISK OF DETACHMENT OF HANDLE. IT WAS CONFIRMED BY PHOTOGRAPHIC EVIDENCE THE SCREW SUPPORTS FROM HANDLE WERE BROKEN WITH MISSING PARTICLES. IT WAS FOUND THAT THE LIKELY CAUSE OF THE PROBLEM WOULD BE FAILURE TO COMPLY WITH THE MAXIMUM NUMBER OF USES/STERILIZATIONS RECOMMENDED BY THE MANUFACTURER. ACCORDING TO THE CUSTOMER'S CALCULATION, THE NUMBER OF TIMES USED/STERILIZED IS ON AVERAGE AROUND 104 TIMES, THUS FAR EXCEEDING THE RECOMMENDED AMOUNT- ACCORDING TO THE IFU 01781 EN 18 PAGE 92, THE HANDLES HAVE A MAXIMUM NUMBER OF 50 RECOMMENDED STERILIZATION CYCLES, WITH THE RISK OF FALLING OFF THE SUPPORT IF EXCEEDED. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES OR PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18550159 ↗MalfunctionJan 22, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
ON 9TH JANUARY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL EQUIPMENT - MAQUET STERIGRIP. IT WAS STATED THE MOUNTING WAS BROKEN WITH MISSING PARTICLES ON STERIGRIP, AND AS A RESULT THERE WAS A RISK OF DETACHMENT OF HANDLE. IT WAS CONFIRMED BY PHOTOGRAPHIC EVIDENCE THE SCREW SUPPORTS FROM HANDLE WERE BROKEN WITH MISSING PARTICLES. IT WAS FOUND THAT THE LIKELY CAUSE OF THE PROBLEM WOULD BE FAILURE TO COMPLY WITH THE MAXIMUM NUMBER OF USES/STERILIZATIONS RECOMMENDED BY THE MANUFACTURER. ACCORDING TO THE CUSTOMER'S CALCULATION, THE NUMBER OF TIMES USED/STERILIZED IS ON AVERAGE AROUND 104 TIMES, THUS FAR EXCEEDING THE RECOMMENDED AMOUNT- ACCORDING TO THE IFU 01781 EN 18 PAGE 92, THE HANDLES HAVE A MAXIMUM NUMBER OF 50 RECOMMENDED STERILIZATION CYCLES, WITH THE RISK OF FALLING OFF THE SUPPORT IF EXCEEDED. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES OR PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18550102 ↗MalfunctionJan 22, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
ON 9TH JANUARY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL EQUIPMENT - MAQUET STERIGRIP. IT WAS STATED THE MOUNTING WAS BROKEN WITH MISSING PARTICLES ON STERIGRIP, AND AS A RESULT THERE WAS A RISK OF DETACHMENT OF HANDLE. IT WAS CONFIRMED BY PHOTOGRAPHIC EVIDENCE THE SCREW SUPPORTS FROM HANDLE WERE BROKEN WITH MISSING PARTICLES. IT WAS FOUND THAT THE LIKELY CAUSE OF THE PROBLEM WOULD BE FAILURE TO COMPLY WITH THE MAXIMUM NUMBER OF USES/STERILIZATIONS RECOMMENDED BY THE MANUFACTURER. ACCORDING TO THE CUSTOMER'S CALCULATION, THE NUMBER OF TIMES USED/STERILIZED IS ON AVERAGE AROUND 104 TIMES, THUS FAR EXCEEDING THE RECOMMENDED AMOUNT- ACCORDING TO THE IFU 01781 EN 18 PAGE 92, THE HANDLES HAVE A MAXIMUM NUMBER OF 50 RECOMMENDED STERILIZATION CYCLES, WITH THE RISK OF FALLING OFF THE SUPPORT IF EXCEEDED. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES OR PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18550100 ↗MalfunctionJan 22, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
MANUFACTURER'S REFERENCE NUMBER: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18550099 ↗MalfunctionJan 22, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
ON 9TH JANUARY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL EQUIPMENT - MAQUET STERIGRIP. IT WAS STATED THE MOUNTING WAS BROKEN WITH MISSING PARTICLES ON STERIGRIP, AND AS A RESULT THERE WAS A RISK OF DETACHMENT OF HANDLE. IT WAS CONFIRMED BY PHOTOGRAPHIC EVIDENCE THE SCREW SUPPORTS FROM HANDLE WERE BROKEN WITH MISSING PARTICLES. IT WAS FOUND THAT THE LIKELY CAUSE OF THE PROBLEM WOULD BE FAILURE TO COMPLY WITH THE MAXIMUM NUMBER OF USES/STERILIZATIONS RECOMMENDED BY THE MANUFACTURER. ACCORDING TO THE CUSTOMER'S CALCULATION, THE NUMBER OF TIMES USED/STERILIZED IS ON AVERAGE AROUND 104 TIMES, THUS FAR EXCEEDING THE RECOMMENDED AMOUNT- ACCORDING TO THE IFU 01781 EN 18 PAGE 92, THE HANDLES HAVE A MAXIMUM NUMBER OF 50 RECOMMENDED STERILIZATION CYCLES, WITH THE RISK OF FALLING OFF THE SUPPORT IF EXCEEDED. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES OR PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18550098 ↗MalfunctionJan 22, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
MANUFACTURER'S REFERENCE NUMBER (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18550097 ↗MalfunctionJan 22, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
MANUFACTURER'S REFERENCE NUMBER: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18550096 ↗MalfunctionJan 22, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
ON 9TH JANUARY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL EQUIPMENT - MAQUET STERIGRIP. IT WAS STATED THE MOUNTING WAS BROKEN WITH MISSING PARTICLES ON STERIGRIP, AND AS A RESULT THERE WAS A RISK OF DETACHMENT OF HANDLE. IT WAS CONFIRMED BY PHOTOGRAPHIC EVIDENCE THE SCREW SUPPORTS FROM HANDLE WERE BROKEN WITH MISSING PARTICLES. IT WAS FOUND THAT THE LIKELY CAUSE OF THE PROBLEM WOULD BE FAILURE TO COMPLY WITH THE MAXIMUM NUMBER OF USES/STERILIZATIONS RECOMMENDED BY THE MANUFACTURER. ACCORDING TO THE CUSTOMER'S CALCULATION, THE NUMBER OF TIMES USED/STERILIZED IS ON AVERAGE AROUND 104 TIMES, THUS FAR EXCEEDING THE RECOMMENDED AMOUNT- ACCORDING TO THE IFU 01781 EN 18 PAGE 92, THE HANDLES HAVE A MAXIMUM NUMBER OF 50 RECOMMENDED STERILIZATION CYCLES, WITH THE RISK OF FALLING OFF THE SUPPORT IF EXCEEDED. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES OR PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18550095 ↗MalfunctionJan 22, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
ON 9TH JANUARY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL EQUIPMENT - MAQUET STERIGRIP. IT WAS STATED THE MOUNTING WAS BROKEN WITH MISSING PARTICLES ON STERIGRIP, AND AS A RESULT THERE WAS A RISK OF DETACHMENT OF HANDLE. IT WAS CONFIRMED BY PHOTOGRAPHIC EVIDENCE THE SCREW SUPPORTS FROM HANDLE WERE BROKEN WITH MISSING PARTICLES. IT WAS FOUND THAT THE LIKELY CAUSE OF THE PROBLEM WOULD BE FAILURE TO COMPLY WITH THE MAXIMUM NUMBER OF USES/STERILIZATIONS RECOMMENDED BY THE MANUFACTURER. ACCORDING TO THE CUSTOMER'S CALCULATION, THE NUMBER OF TIMES USED/STERILIZED IS ON AVERAGE AROUND 104 TIMES, THUS FAR EXCEEDING THE RECOMMENDED AMOUNT- ACCORDING TO THE IFU 01781 EN 18 PAGE 92, THE HANDLES HAVE A MAXIMUM NUMBER OF 50 RECOMMENDED STERILIZATION CYCLES, WITH THE RISK OF FALLING OFF THE SUPPORT IF EXCEEDED. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES OR PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18550094 ↗MalfunctionJan 22, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
MANUFACTURER'S REFERENCE NUMBER (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18550093 ↗MalfunctionJan 22, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
ON 9TH JANUARY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL EQUIPMENT - MAQUET STERIGRIP. IT WAS STATED THE MOUNTING WAS BROKEN WITH MISSING PARTICLES ON STERIGRIP, AND AS A RESULT THERE WAS A RISK OF DETACHMENT OF HANDLE. IT WAS CONFIRMED BY PHOTOGRAPHIC EVIDENCE THE SCREW SUPPORTS FROM HANDLE WERE BROKEN WITH MISSING PARTICLES. IT WAS FOUND THAT THE LIKELY CAUSE OF THE PROBLEM WOULD BE FAILURE TO COMPLY WITH THE MAXIMUM NUMBER OF USES/STERILIZATIONS RECOMMENDED BY THE MANUFACTURER. ACCORDING TO THE CUSTOMER'S CALCULATION, THE NUMBER OF TIMES USED/STERILIZED IS ON AVERAGE AROUND 104 TIMES, THUS FAR EXCEEDING THE RECOMMENDED AMOUNT- ACCORDING TO THE IFU 01781 EN 18 PAGE 92, THE HANDLES HAVE A MAXIMUM NUMBER OF 50 RECOMMENDED STERILIZATION CYCLES, WITH THE RISK OF FALLING OFF THE SUPPORT IF EXCEEDED. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES OR PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18550092 ↗MalfunctionJan 18, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
ON 9TH JANUARY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL EQUIPMENT - MAQUET STERIGRIP. IT WAS STATED THE MOUNTING WAS BROKEN WITH MISSING PARTICLES ON STERIGRIP, AND AS A RESULT THERE WAS A RISK OF DETACHMENT OF HANDLE. IT WAS CONFIRMED BY PHOTOGRAPHIC EVIDENCE THE SCREW SUPPORTS FROM HANDLE WERE BROKEN WITH MISSING PARTICLES. IT WAS FOUND THAT THE LIKELY CAUSE OF THE PROBLEM WOULD BE FAILURE TO COMPLY WITH THE MAXIMUM NUMBER OF USES/STERILIZATIONS RECOMMENDED BY THE MANUFACTURER. ACCORDING TO THE CUSTOMER'S CALCULATION, THE NUMBER OF TIMES USED/STERILIZED IS ON AVERAGE AROUND 104 TIMES, THUS FAR EXCEEDING THE RECOMMENDED AMOUNT- ACCORDING TO THE IFU 01781 EN 18 PAGE 92, THE HANDLES HAVE A MAXIMUM NUMBER OF 50 RECOMMENDED STERILIZATION CYCLES, WITH THE RISK OF FALLING OFF THE SUPPORT IF EXCEEDED. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES OR PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18533561 ↗
Other Light, Surgical, Accessories devices
- NA Sterile Lighthandle CoverSteris Corporation · Light, surgical, accessories1 recall93 reports
- Harmony iQ Harmony iQ 3600Black Diamond Video, Inc. · Light, surgical, accessories20 reports
- HexaVue HexaVue-SLC 1.0Black Diamond Video, Inc. · Light, surgical, accessories9 reports
- HexaVue HexaVue IPBlack Diamond Video, Inc. · Light, surgical, accessories7 reports
- NA LB83Steris Corporation · Light, surgical, accessories6 reports
- HexaVue HexaVue FocusBlack Diamond Video, Inc. · Light, surgical, accessories5 reports
- Harmony iQ Harmony iQ 2800Black Diamond Video, Inc. · Light, surgical, accessories3 reports
- Aspen® Flexible Light Handle Cover LT-F01BAspen Surgical Products, Inc. · Light, surgical, accessories3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page