Skip to content

STANDOP VOLISTA® ARD267900100ARecalls & reported problems

Maquet Sas · VST66DF TK AIM XS12 SPE · Class 2

All models of STANDOP VOLISTA (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 03700712406004), received between Feb 23, 2023 and Mar 18, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (8 reports).

19 other recalls cover devices cleared under the same FDA submission (K130513); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component8
  • Fracture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2026Detachment of Device or Device Component

    GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS: VOLISTA STANDOP. AS IT WAS STATED, UPON THE DEVICE INSPECTION ONE OUT OF THE SIX FIXATION SCREWS HOLDING THE DEVICE MAIN TUBE WAS LOOSE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS LOOSE SCREWS HOLDING THE DEVICE CONFIGURATION MAY LEAD TO THE DEVICE DETACHMENT FROM THE CEILING AND SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24632242 ↗
  • MalfunctionMar 18, 2026Detachment of Device or Device Component

    GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA STANDOP. AS IT WAS STATED, UPON THE DEVICE INSPECTION ONE OUT OF THE SIX FIXATION SCREWS HOLDING THE DEVICE MAIN TUBE WAS LOOSE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS LOOSE SCREWS HOLDING THE DEVICE CONFIGURATION MAY LEAD TO THE DEVICE DETACHMENT FROM THE CEILING AND SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24626399 ↗
  • MalfunctionMar 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA STANDOP. AS IT WAS STATED, UPON THE DEVICE INSPECTION ONE OUT OF THE SIX FIXATION SCREWS HOLDING THE DEVICE MAIN TUBE WERE LOOSE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS LOOSE SCREWS HOLDING THE DEVICE CONFIGURATION MAY LEAD TO THE DEVICE DETACHMENT FROM THE CEILING AND SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21503996 ↗
  • MalfunctionMar 5, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ON 28TH FEBRUARY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR SURGICAL LIGHTS ¿ VOLISTA STANDOP. AS IT WAS STATED THE DUST COVER WAS MISSING. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18835670 ↗
  • MalfunctionMar 22, 2023Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    ON (B)(6)2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA STANDOP. AS IT WAS STATED, UPON THE DEVICE INSPECTION THREE OUT OF THE SIX FIXATION SCREWS HOLDING THE DEVICE MAIN TUBE WERE BROKEN. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS BROKEN SCREWS HOLDING THE DEVICE CONFIGURATION MAY LEAD TO THE DEVICE DETACHMENT FROM THE CEILING AND SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16589148 ↗
  • MalfunctionFeb 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    ON (B)(6) 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA STANDOP. AS IT WAS STATED, UPON THE DEVICE INSPECTION ONE OUT OF THE SIX FIXATION SCREWS HOLDING THE DEVICE MAIN TUBE WERE LOOSEN. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS LOOSEN SCREWS HOLDING THE DEVICE CONFIGURATION MAY LEAD TO THE DEVICE DETACHMENT FROM THE CEILING AND SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16455895 ↗
  • MalfunctionFeb 27, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    MANUFACTURER'S REFERENCE NUMBER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16442507 ↗
  • MalfunctionFeb 23, 2023Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    ON 19TH FEBRUARY 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA STANDOP. AS IT WAS STATED, UPON THE DEVICE INSPECTION THREE OUT OF THE SIX FIXATION SCREWS HOLDING THE DEVICE MAIN TUBE WERE BROKEN. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS BROKEN SCREWS HOLDING THE DEVICE CONFIGURATION MAY LEAD TO THE DEVICE DETACHMENT FROM THE CEILING AND SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16426079 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K130513. One submission often covers many models, so check the product and lot details in each record.

  • Z-0758-2024 · Nov 8, 2023 · Open, Classified
    Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog Numbers ROL B FOR VLT400SF STP - ARDROL309001A, ROL B FOR VLT400SF STP
  • Z-1308-2022 · May 6, 2022 · Open, Classified
    VOLISTA StandOP Surgical Light, Product Code/Part Numbers ARD568811901 VLT600SF AIM STP ARD568811911 VLT600SF AIM STP ARD568811961 VLT600DF AIM STP ARDVST229000A VST60SF AIM ARDVST
  • Z-1864-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229037A Model: VST66DF
  • Z-1859-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229002A Model: VST66SF AIM
  • Z-1860-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229003A Model: VST66DF AIM
  • Z-1862-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229035A Model:VST60DF
  • Z-1853-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARD568811910 Model:VLT600SF STP
  • Z-1863-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229036A Model:VST66SF
  • Z-1856-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARD568812910 Model: VLT400SF STP
  • Z-1858-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229001A Model: VST60DF AIM
  • Z-1861-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229034A Model: VST60SF
  • Z-1855-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARD568811961 Model: VLT600DF AIM STP
  • Z-1854-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARD568811911 Model: VLT600SF AIM STP
  • Z-1857-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229000A Model: VST60SF AIM
  • Z-1852-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARD568811901 Model: VLT600SF AIM STP

…and 4 more.

Other Lamp, Surgical devices

All Lamp, Surgical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page