Maquet PowerLED II ARDPWT209121ARecalls & reported problems

Maquet Sas · PWDII50SF AIM XO K3 1416 · Class 2

All models of Maquet PowerLED II (992)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03700712406486), received between Sep 5, 2023 and Jan 30, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component4
  • Device Contaminated at the User Facility4
  • Fluid/Blood Leak4
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 30, 2025Peeled/Delaminated; Peeled/Delaminated

    GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II. IT WAS STATED THE PAINT WAS CHIPPING FROM THE FORK. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PAINT PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21271952 ↗
  • MalfunctionSep 5, 2023Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility; Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility

    ON (B)(6), 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II. IT WAS STATED THE INNER SEAL HAS PULLED AWAY FROM THE CORNER ALLOWING MOISTURE INTO THE LIGHT PANELS CAUSING STREAKING AND DEBRIS INTERNALLY. BASED ON PHOTOGRAPHIC EVIDENCE PARTICLES MAY BE MISSING FROM THE SEAL. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS CONTACT OF WATER WITH LIVE PARTS MAY CAUSE ELECTRIC SHOCK AND PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17682451 ↗
  • MalfunctionSep 5, 2023Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility; Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility

    ON (B)(6) 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II. IT WAS STATED THE INNER SEAL HAS PULLED AWAY FROM THE CORNER ALLOWING MOISTURE INTO THE LIGHT PANELS CAUSING STREAKING AND DEBRIS INTERNALLY. BASED ON PHOTOGRAPHIC EVIDENCE PARTICLES MAY BE MISSING FROM THE SEAL. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS CONTACT OF WATER WITH LIVE PARTS MAY CAUSE ELECTRIC SHOCK AND PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17682449 ↗
  • MalfunctionSep 5, 2023Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility; Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility

    ON (B)(6) 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II. IT WAS STATED THE INNER SEAL HAS PULLED AWAY FROM THE CORNER ALLOWING MOISTURE INTO THE LIGHT PANELS CAUSING STREAKING AND DEBRIS INTERNALLY. BASED ON PHOTOGRAPHIC EVIDENCE PARTICLES MAY BE MISSING FROM THE SEAL. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS CONTACT OF WATER WITH LIVE PARTS MAY CAUSE ELECTRIC SHOCK AND PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17682448 ↗
  • MalfunctionSep 5, 2023Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility; Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility

    ON 7TH AUGUST, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II. IT WAS STATED THE INNER SEAL HAS PULLED AWAY FROM THE CORNER ALLOWING MOISTURE INTO THE LIGHT PANELS CAUSING STREAKING AND DEBRIS INTERNALLY. BASED ON PHOTOGRAPHIC EVIDENCE PARTICLES MAY BE MISSING FROM THE SEAL. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS CONTACT OF WATER WITH LIVE PARTS MAY CAUSE ELECTRIC SHOCK AND PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17682443 ↗

Other Lamp, surgical devices

All Lamp, surgical devices →