Maquet PowerLED II ARDPWT229139ARecalls & reported problems
Maquet Sas · PWDII70SF VP AIM XO K3 1416 · Class 2
All models of Maquet PowerLED II (992)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03700712406837), received between Sep 5, 2023 and Nov 29, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component5
- Device Contaminated at the User Facility4
- Fluid/Blood Leak4
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 29, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ON 22ND NOVEMBER, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II 700. IT WAS STATED THE IMPACT SHOCK MOVED THE REAR COVER SIDE WAYS AND SPLIT THE KIT. THEN, AS THE USER MOVED THE SURGICAL LIGHT UP AND DOWN, THE DUST COVER MOVED FORWARD. THE TECHNICIAN INFORMED THE METAL PLATE INSIDE THE COVER DETACHED. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS OR PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18225806 ↗MalfunctionSep 5, 2023Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility; Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility
ON (B)(6) 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II. IT WAS STATED THE INNER SEAL HAS PULLED AWAY FROM THE CORNER ALLOWING MOISTURE INTO THE LIGHT PANELS CAUSING STREAKING AND DEBRIS INTERNALLY. BASED ON PHOTOGRAPHIC EVIDENCE PARTICLES MAY BE MISSING FROM THE SEAL. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS CONTACT OF WATER WITH LIVE PARTS MAY CAUSE ELECTRIC SHOCK AND PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17682452 ↗MalfunctionSep 5, 2023Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility; Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility
ON (B)(6) 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II. IT WAS STATED THE INNER SEAL HAS PULLED AWAY FROM THE CORNER ALLOWING MOISTURE INTO THE LIGHT PANELS CAUSING STREAKING AND DEBRIS INTERNALLY. BASED ON PHOTOGRAPHIC EVIDENCE PARTICLES MAY BE MISSING FROM THE SEAL. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS CONTACT OF WATER WITH LIVE PARTS MAY CAUSE ELECTRIC SHOCK AND PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17682450 ↗MalfunctionSep 5, 2023Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility; Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility
ON 7TH AUGUST, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II. IT WAS STATED THE INNER SEAL HAS PULLED AWAY FROM THE CORNER ALLOWING MOISTURE INTO THE LIGHT PANELS CAUSING STREAKING AND DEBRIS INTERNALLY. BASED ON PHOTOGRAPHIC EVIDENCE PARTICLES MAY BE MISSING FROM THE SEAL. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS CONTACT OF WATER WITH LIVE PARTS MAY CAUSE ELECTRIC SHOCK AND PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17682447 ↗MalfunctionSep 5, 2023Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility; Fluid/Blood Leak; Detachment of Device or Device Component; Device Contaminated at the User Facility
ON 7TH AUGUST, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II. IT WAS STATED THE INNER SEAL HAS PULLED AWAY FROM THE CORNER ALLOWING MOISTURE INTO THE LIGHT PANELS CAUSING STREAKING AND DEBRIS INTERNALLY. BASED ON PHOTOGRAPHIC EVIDENCE PARTICLES MAY BE MISSING FROM THE SEAL. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS CONTACT OF WATER WITH LIVE PARTS MAY CAUSE ELECTRIC SHOCK AND PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17682445 ↗
Other Lamp, surgical devices
- ROLITE VOLISTA STANDOP ARDROL309023AMaquet Sas · Lamp, surgical2 recalls0 reports
- ROLITE VOLISTA STANDOP ARDROL309018AMaquet Sas · Lamp, surgical2 recalls0 reports
- Maquet PowerLED II ARDPWT259215AMaquet Sas · Lamp, surgical1 recall30 reports
- iLED7 Surgical Light 4068310Baxter Medical Systems GmbH + Co. KG · Lamp, surgical1 recall5 reports
- Maquet PowerLED II ARDPWT629089AMaquet Sas · Lamp, surgical1 recall3 reports
- Maquet PowerLED II ARDPWT259043AMaquet Sas · Lamp, surgical1 recall3 reports
- Maquet PowerLED II ARDPWT229064AMaquet Sas · Lamp, Surgical1 recall3 reports
- Maquet PowerLED II ARDPWT209058AMaquet Sas · Lamp, surgical1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page