Maquet PowerLED II ARDPWT209011ARecalls & reported problems

Maquet Sas · PWDII75SFDF AIM K3 1011 · Class 2

All models of Maquet PowerLED II (992)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03700712407230), received between Aug 11, 2023 and Aug 11, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Moisture Damage” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Nov 8, 2023Class IIOpen, Classified

    A potential for a light system to fall in the operating room.

    Z-0757-2024 · Root cause: Equipment maintenance · Recall details · FDA record ↗

Most reported problems

Device problems

  • Moisture Damage1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 11, 2023Moisture Damage; Moisture Damage

    ON (B)(6) 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR SURGICAL LIGHTS ¿ POWERLED II. THE RECEIVED CUSTOMER ALLEGATION REGARDED FLICKERING LIGHT AND THE ISSUE WAS CONSIDERED AS A NON-REPORTABLE. ON (B)(6) 2023, THE GETINGE SERVICE TECHNICIAN ARRIVED ON SITE AND DISCOVERED CORROSION INSIDE THE DEVICE ON THE METAL STRUCTURE AND THE LED SEGMENT BOARD WITH INDICATION OF THE PRESENCE OF DRIED LIQUID. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS CONTACT OF WATER WITH LIVE PARTS MAY CAUSE ELECTRIC SHOCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17507800 ↗

Other Lamp, Surgical devices

All Lamp, Surgical devices →