Skip to content

Maquet PowerLED II ARDPWT209029ARecalls & reported problems

Maquet Sas · PWDII55DF AIM K4 0810 · Class 2

All models of Maquet PowerLED II (992)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03700712407407), received between May 16, 2023 and Jul 21, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fluid/Blood Leak1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 21, 2023Fluid/Blood Leak; Fluid/Blood Leak

    ON (B)(6), 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II 500. IT WAS STATED DUE TO THE AIR-CONDITIONING WATER ENTERING THE POWER SUPPLY CABINET, ONE HEADLIGHT TURNED OFF AND SECOND TURNED ON. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS CONTACT OF WATER WITH LIVE PARTS MAY CAUSE ELECTRIC SHOCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17369197 ↗
  • MalfunctionMay 16, 2023Leak/Splash; Leak/Splash

    ON 9TH MAY, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR SURGICAL LIGHTS - POWERLED II. INITIALLY IT WAS STATED THE ONE LIGHTHEAD DID NOT WORK. BASED ON ADDITIONAL CLARIFICATION PROVIDED BY THE GETINGE TECHNICIAN, IT WAS ESTABLISHED THAT THE LOCAL ROOM AIR CONDITIONER WAS LEAKING WATER INTO THE CEILING POWER SUPPLY BOX, THERE WAS A WATER INSIDE THE BOX AND POWER SUPPLY BECAME DAMAGED. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS CONTACT OF WATER WITH LIVE PARTS MAY CAUSE ELECTRIC SHOCK. ACCORDING TO THE INFORMATION PROVIDED BY GETINGE TECHNICIAN, THAT AFTER HIS VISIT THE AFFECTED DEVICE IS OPERATIONAL AND CAN BE USED NORMALLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16940122 ↗

Other Lamp, Surgical devices

All Lamp, Surgical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page