Skip to content

Maquet PowerLED II ARDPWT229035ARecalls & reported problems

Maquet Sas · PWDII75DF VP AIM K4 0810 · Class 2

All models of Maquet PowerLED II (992)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03700712408404), received between Jan 5, 2023 and Jan 5, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Misassembled” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Misassembled1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 5, 2023Misassembled; Misassembled

    ON 3RD JANUARY, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II. IT WAS STATED THE SUPPORT WHERE THE DEVICE IS ANCHORED SHOWED FATIGUE IN THE WELDING TO THE STRUCTURE, CAUSING THE UNEVEN CONNECTION. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS THIS UNEVEN CONNECTION COULD LEAD TO FALL OF THE DEVICE AND AS A RESULT OF THAT, COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16101344 ↗

Other Lamp, surgical devices

All Lamp, surgical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page