1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03700712412593), received between Oct 7, 2024 and Oct 7, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Unexpected Shutdown” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 8, 2023Class IIOpen, Classified

    A potential for a light system to fall in the operating room.

    Z-0757-2024 · Root cause: Equipment maintenance · Recall details · FDA record ↗

Most reported problems

Device problems

  • Unexpected Shutdown1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 7, 2024Unexpected Shutdown; Unexpected Shutdown

    ON 25TH SEPTEMBER, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHT - MAQUET POWERLED II. IT WAS STATED THAT THERE WAS A POWER FAILURE AND THE BATTERY SUPPLY DID NOT WORK. THEREFORE THE LIGHTS WENT OUT DURING THE OPERATION AND THE SURGEON WAS IN THE DARK. ONLY WHEN THE EMERGENCY POWER GENERATOR WAS STARTED UP FOR THE OPERATION DID THEY LIGHT UP AGAIN. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS THE EXTINCTION OF SURGICAL LIGHT SYSTEM DURING PROCEDURE MAY CAUSE SERIOUS INJURY IN CASE OF REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20380732 ↗

FDA clearance and classification

The FDA regulates product code FTD (lamp, surgical) under 21 CFR 878.4580 as a Class II device. 21 CFR 878.4580 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Maquet PowerLED II ARDPWT239221A

Has the Maquet PowerLED II ARDPWT239221A been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0757-2024 (Class II), began on Nov 8, 2023: A potential for a light system to fall in the operating room. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Maquet PowerLED II ARDPWT239221A?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Oct 7, 2024. That report lists device problem “Unexpected Shutdown” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Maquet PowerLED II ARDPWT239221A?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Maquet PowerLED II ARDPWT239221A FDA approved?

Devices of this type (product code FTD) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Maquet Sas.

Is the Maquet PowerLED II ARDPWT239221A MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the Maquet PowerLED II ARDPWT239221A, and is it still sold?

The FDA's device database lists Maquet Sas as the company responsible for it (brand Maquet PowerLED II, model ARDPWT239221A). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Lamp, surgical devices

All Lamp, surgical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page