Skip to content

Maquet PowerLED II ARDPWT259052ARecalls & reported problems

Maquet Sas · PWDII70DF VP AIM CL XHS021 K3 1011 · Class 2

All models of Maquet PowerLED II (992)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03700712416379), received between Feb 16, 2024 and Nov 4, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component4
  • Crack2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 4, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component

    MANUFACTURER'S REFERENCE NUMBER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20597216 ↗
  • MalfunctionSep 27, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ON 12TH SEPTEMBER, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II 700. IT WAS STATED THE RUBBER GASKET FELL OFF. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS OR PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20321170 ↗
  • MalfunctionMay 7, 2024Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component

    ON (B)(6) 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR SURGICAL LIGHTS ¿ POWERLED II 700 . AS IT WAS STATED THE TOP COVER WAS CRACKED AND IT WAS BREAKING INTO SMALL PIECES OF PLASTIC WHICH HAVE THE POTENTIAL TO FALL. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19257916 ↗
  • MalfunctionFeb 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ON 18TH DECEMBER, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II 700. BASED ON DESCRIBED ISSUES, IT WAS NOT POSSIBLE TO DETERMINE IF ANY ISSUE IS SAFETY OR RISK RELATED. FURTHER INFORMATION PROVIDED BY GETINGE EMPLOYEE ON 19TH JANUARY 2024 INDICATED THERE WAS THE SAFETY SEGMENT SLEEVE MISSING AND THE SNAP RING WAS OUT OF PLACE IN THE SPRING ARM. THERE WAS NO INJURY REPORTED, HOWEVER, BASED ON ADDITIONAL INPUT FROM GETINGE EMPLOYEE IT WAS POSSIBLE TO DETERMINE THAT THE ISSUES INVESTIGATED HEREIN ARE SAFETY AND RISK RELATED, AS SLEEVE FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION IN CASE OF REOCCURRENCE AND THE A SNAP RING BEING OUT OF PLACE IN THE SPRING ARM COULD RESULT IN A DETACHMENT OF SPRING ARM AND DOME AND AS A RESULT OF THAT, COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18718752 ↗

Other Lamp, Surgical devices

All Lamp, Surgical devices →