Skip to content

Maquet PowerLED II ARDPWT609051ARecalls & reported problems

Maquet Sas · PWDII50SF CL K3 SAX12 · Class 2

All models of Maquet PowerLED II (992)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03700712423292), received between Feb 19, 2024 and Jan 12, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 8, 2023Class IIOpen, Classified

    A potential for a light system to fall in the operating room.

    Z-0757-2024 · Root cause: Equipment maintenance · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 12, 2026Detachment of Device or Device Component

    GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED II. IT WAS STATED THE COVER OF THE SPRING ARM OF THE SURGICAL LIGHTING DEVICE FELL OFF DURING SURGERY. UPON CHECKING THE COVER, THERE WERE IMPACT MARKS ON IT. IT APPEARS THAT PART OF THE COVER (THE TABS THAT FIT INTO PLACE) HAS BEEN DAMAGED DUE TO SOME KIND OF IMPACT. THE ISSUE DID NOT CAUSE ANY HEALTH ISSUES FOR THE PATIENT. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24035140 ↗
  • MalfunctionFeb 19, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ON 14TH FEBRUARY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR SURGICAL LIGHTS ¿ POWERLED II. AS IT WAS STATED THE COVER HAS COME OFF. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18729929 ↗

Other Lamp, surgical devices

All Lamp, surgical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page