Skip to content

Maquet PowerLED II ARDPWT679062ARecalls & reported problems

Maquet Sas · PWDII70DF VP4K CL K3 SAX08 · Class 2

All models of Maquet PowerLED II (992)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03700712431372), received between Feb 11, 2026 and Feb 11, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 8, 2023Class IIOpen, Classified

    A potential for a light system to fall in the operating room.

    Z-0757-2024 · Root cause: Equipment maintenance · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 11, 2026Detachment of Device or Device Component

    GETINGE BECAME AWARE OF AN ISSUE INVOLVING ONE OF THE SURGICAL LIGHTS ¿ POWERLED II. IT WAS REPORTED THAT THE CONNECTION BETWEEN THE SUSPENSION ARM AND THE MAIN TUBE WAS FOUND TO BE LOOSE DUE TO THE LACK OF A CIRCLIP. NO INJURIES WERE REPORTED; HOWEVER, WE DECIDED TO REPORT THE ISSUE OUT OF AN ABUNDANCE OF CAUTION, AS A LOOSE SUSPENSION ARM MAY LEAD TO DETACHMENT OF THE DEVICE FROM THE CEILING AND COULD RESULT IN SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24312470 ↗

Other Lamp, surgical devices

All Lamp, surgical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page