ARCAD CS041618Recalls & reported problems

Novastep · Compressive staple 18-16-18 · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03700879501291), received between Oct 25, 2016 and Oct 25, 2016 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break2
  • Improper or Incorrect Procedure or Method1
  • Patient-Device Incompatibility1

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 25, 2016Break; Patient-Device Incompatibility

    HOSPITAL CALLED NOVASTEP SALES REP ON 29TH SEPTEMBER 2016 AND REPORTED ARCAD 18 COMPRESSIVE STAPLE BREAKAGE. ARCAD 18-16-18 COMPRESSIVE STAPLE WAS DETECTED AS BROKEN BY THE SURGEON DURING THE ANALYSIS OF XRAY PERFORMED DURING A PATIENT FOLLOW UP VISIT. WEIGHT OF THE PATIENT ASSOCIATED TO THE USE OF ONLY TWO ARCAD 18 STAPLES TO PROCEED MTP ARTHRODESIS (TOO WEAK) IS EXPECTED TO BE THE ROOT CAUSE OF THIS BREAKAGE. ASSEMBLY SHOULD HAVE BEEN DONE USING BIGGER STAPLES OR ADDITIONAL DEVICES ASSOCIATED WITH TWO ARCAD 18 COMPRESSIVE STAPLES.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6053978 ↗
  • InjuryOct 25, 2016Break; Improper or Incorrect Procedure or Method

    DURING A SALES REP VISIT ON (B)(6) 2016 IT WAS REPORTED TO NOVASTEP SALES REP. THAT DURING THE LAST 12 MONTH 3 CASES HAVE CONDUCTED TO ARCAD 18 COMPRESSIVE STAPLE BREAKAGE. THIS IS THE SAME SURGEON FOR THESE 3 CASES. FIRST CASE: ARCAD 18-16-18 COMPRESSIVE STAPLE WAS DETECTED AS BROKEN BY THE SURGEON DURING THE ANALYSIS OF XRAY PERFORMED DURING A PATIENT FOLLOW UP VISIT. THE SURGEON EXPECTS THAT THE USE OF ONLY TWO ARCAD 18 STAPLES TO PROCEED MTP ARTHRODESIS TO BE TOO WEAK. ASSEMBLY SHOULD HAVE BEEN DONE USING BIGGER STAPLES OR ADDITIONAL DEVICES ASSOCIATED WITH TWO ARCAD 18 COMPRESSIVE STAPLES. THE OTHERS 2 CASES HAVE BEEN REPORTED IN REPORT 3010673777-2016-00001, 3010673777-2016-00002.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6053719 ↗

Other Staple, Fixation, Bone devices

All Staple, Fixation, Bone devices →