Humelock II Cemented Reversible Shoulder 105-3613Recalls & reported problems

Fx Solutions · HUMELOCK REVERSED ECCENTRIC GLENOSPHERE W/ SCREW CoCr/TA6V 10° TILT Ø36 mm · Class 2 · Implant

All models of Humelock II Cemented Reversible Shoulder (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 03701037300275), received between Mar 31, 2021 and Jul 31, 2026 — classified as 9 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Device Dislodged or Dislocated3
  • Appropriate Device Problem Term/Code Not Available2
  • Improper or Incorrect Procedure or Method2
  • Fracture1
  • Migration1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Insufficient Information2
  • Medical device site infection2
  • Inadequate Osseointegration1
  • Joint Dislocation1
  • Joint Laxity1
  • Unspecified Infection1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 31, 2026Adverse Event Without Identified Device or Use Problem

    Revision occurred on (b)(6) 2026 approximately 465 days after the primary surgery which occurred on (b)(6) 2025. No fall or other trauma was reported. Based on comparing usage forms only HUMELOCK REVERSED ECCENTRIC GLENOSPHERE W/ SCREW CoCr/TA6V 10° TILT Ø36mm, HUMERAL CUP 135/145° STANDARD PE/TA6V Ø36 +9 and HUMELOCK REVERSED TA6V CEMENTLESS GLENOID BASEPLATE W/ SCREW Ti/HA Ø24mm, +3mm LATERALIZED was explanted due to dislocation and replaced with REVERSED ADAPTER TA6V 135/145,° HUMERAL CUP STABILITY PE/TA6V Ø36/+3, DOUBLE TAPER TA6V +0mm 0° and CENTERED HEAD CoCr 50x19.

    Patient problems: Joint Dislocation

    MAUDE report 50104916 ↗
  • InjurySep 19, 2025Migration; Migration

    REVISION OCCURRED APPROXIMATELY 43 WEEKS AFTER THE PRIMARY FX SURGERY, WHICH OCCURRED ON (B)(6) 2024. THE PRIMARY FX CASE WAS A REVISION OF COMPETITOR. NO FALL OR OTHER TRAUMA REPORTED. REVISION OCCURRED DUE TO THE GLENOSPHERE COMING LOOSE FROM THE BASEPLATE, UNSCREWING IS SPECIFICALLY MENTIONED AS THE MECHANISM OF LOOSENING. A 36 MM ECCENTRIC GLENOSPHERE AND A 36 MM + 3 STANDARD HUMERAL CUP WERE EXPLANTED. THESE ITEMS WERE REPLACED WITH A 40 MM ECCENTRIC GLENOSPHERE AND A 40 MM + 6 HUMERAL STABILITY CUP.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23100228 ↗
  • InjuryJul 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION OCCURRED APPROXIMATELY 6 MONTHS AFTER THE PRIMARY SURGERY, WHICH OCCURRED ON (B)(6) 2024. NO FALL OR OTHER TRAUMA REPORTED. REVISION OCCURRED DUE TO INFECTION AT IMPLANT SITE. A 36 MM CENTERED GLENOSPHERE AND 36 MM + 3 HUMERAL STANDARD CUP WERE EXPLANTED. THESE ITEMS WERE REPLACED WITH A 40 MM CENTERED GLENOSPHERE AND A 40 MM + 9 STANDARD HUMERAL CUP.

    Patient problems: Medical device site infection

    MAUDE report 22512187 ↗
  • MalfunctionApr 30, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    REVISION OCCURRED APPROXIMATELY 7 MONTHS AFTER THE PRIMARY SURGERY WHICH OCCURRED ON (B)(6) 2024. NO FALL OR OTHER TRAUMA REPORTED. REVISION OCCURRED DUE TO DISSOCIATION OF GLENOSPHERE FROM THE BASEPLATE. 36 MM ECCENTRIC GLENOSPHERE AND 36 MM + 3 HUMERAL STABILITY CUP. THESE PARTS WERE REPLACED WITH A 36 MM ECCENTRIC GLENOSPHERE AND 36 MM + 9 HUMERAL STANDARD CUP.

    Patient problems: Joint Laxity

    MAUDE report 21935081 ↗
  • InjuryOct 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT REVISED APPROX. 6 MONTHS AFTER IMPLANTATION DATE, AS THE SURGEON NOTICED AN INFECTION AT THE IMPLANT SITE. ORIGINAL 36 MM GLENOPHERE AND 36 MM + 3 STANDARD HUMERAL CUP EXPLANTED. THESE PARTS WERE REPLACED WITH 36 MM GLENOPHERE AND 36 MM + 6 STABILITY HUMERAL CUP. INFECTION IS ONGOING AFTER REVISION.

    Patient problems: Medical device site infection

    MAUDE report 17950039 ↗
  • InjuryMar 2, 2022Device Dislodged or Dislocated

    THE PATIENT WAS REVISED FOR DISLOCATION ON (B)(6) 2021. APPROXIMATELY 3 WEEKS AFTER THE FIRST SURGERY. THE SURGEON EXPLANTED 1 ECCENTRIC GLENOSPHERE (36), 1 HUMERAL SPACER (9) AND 1 HUMERAL CUP (36/9). THE SURGEON IMPLANTED 2 CORTICAL SCREW 26 AND 28, 1 HUMERAL SPACER (9) AND 1 HUMERAL CUP (36/6).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13652913 ↗
  • InjuryMar 2, 2022Device Dislodged or Dislocated

    THE PATIENT WAS REVISED ON (B)(6) 2022 FOR DISLOCATION. APPROXIMATELY 4 MONTHS AFTER THE FIRST SURGERY. THE SURGEON EXPLANTED 1 ECCENTRIC GLENOSPHERE AND 1 HUMERAL CUP 135/145° Ø36+9. THE SURGEON IMPLANTED 1 ECCENTRIC GLENOSPHERE, 1 ADAPTER AND 1 STABILITY HUMERAL CUP Ø40+3.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13652835 ↗
  • InjuryJan 5, 2022Fracture; Device Dislodged or Dislocated; Appropriate Device Problem Term/Code Not Available

    THE PATIENT WAS REVISED ON (B)(6) 2021 AFTER PRIMARY SURGERY ON (B)(6) 2021 FOR DISLOCATION DUE TO PROXIMAL HUMERAL FRACTURE. THE SURGEON EXPLANTED AN ECCENTRIC GLENOSPHERE (36), HUMERAL SPACER (9) AND HUMERAL CUP (36/9). THE SURGEON IMPLANTED 2 CORTICAL SCREWS (1 OF 28MM AND 26 MM EACH), ECCENTRIC GLENOSPHERE (40), HUMERAL SPACER (9) AND HUMERAL CUP (36/6).

    Patient problems: Insufficient Information

    MAUDE report 13172503 ↗
  • InjuryMay 10, 2021Improper or Incorrect Procedure or Method; Appropriate Device Problem Term/Code Not Available

    PATIENT REVISED FOR THIRD TIME DUE TO NON-COMPLIANCE WITH POST-SURGICAL INSTRUCTIONS ON (B)(6) 2021, APPROXIMATELY 5 WEEKS AFTER PRIMARY SURGERY. FIRST TWO REVISIONS PREVIOUSLY REPORTED. SURGEON EXPLANTED 36MM ECCENTRIC GLENOSPHERE WITH SCREW, 36/+9 STABILITY HUMERAL CUP, AND HUMELOCK REVERSED 36/14 CEMENTLESS LOCKABLE STEM. HE THEN IMPLANTED A 40MM ECCENTRIC GLENOSPHERE WITH SCREW AND HUMERAL SIDE COMPONENTS FROM ANOTHER MANUFACTURER, WHICH IS AN OFF-LABEL USE OF THE FX SHOULDER DEVICE.

    Patient problems: Insufficient Information

    MAUDE report 11796988 ↗
  • InjuryMar 31, 2021Adverse Event Without Identified Device or Use Problem

    PATIENT REVISED ON (B)(6) 2021 DUE TO INFECTION, APPROXIMATELY 1 YEAR AFTER PRIMARY SURGERY. SURGEON CONVERTED REVERSE TO A HEMI, EXPLANTING ALL COMPONENTS EXCEPT FOR STEM (36/+3 STANDARD HUMERAL CUP, 36MM ECCENTRIC GLENOSPHERE WITH SCREW, 24MM GLENOID BASEPLATE, 2 STANDARD SCREWS, AND 2 LOCKING SCREWS), AND THEN IMPLANTING +0MM DOUBLE TAPER AND 39X14 CENTERED HUMERAL HEAD.

    Patient problems: Unspecified Infection

    MAUDE report 11594202 ↗