FX V135 Shoulder Prosthesis 204-0806Recalls & reported problems
Fx Solutions · FX V135 HUMERAL CUP 135/145° STANDARD PE/TA6V Ø40 +6 · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03701037315347), received between Dec 3, 2025 and Dec 12, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Improper or Incorrect Procedure or Method1
Patient problems
- Inadequate Osseointegration1
- Joint Laxity1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION OCCURRED APPROXIMATELY 6 WEEKS AFTER THE PRIMARY SURGERY, WHICH OCCURRED ON (B)(6) 2025. NO FALL OR OTHER TRAUMA REPORTED. REVISION OCCURRED DUE TO INSTABILITY. A 40 MM + 6 135/145* HUMERAL STANDARD CUP WAS EXPLANTED. THIS ITEM WAS REPLACED WITH A 40 MM + 3 135/145 HUMERAL STABILITY CUP AND A 9 MM SPACER.
Patient problems: Joint Laxity
MAUDE report 23791409 ↗InjuryDec 3, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
REVISION OCCURRED APPROXIMATELY 8 WEEKS AFTER THE PRIMARY SURGERY, WHICH OCCURRED ON (B)(6) 2025. NO FALL OR OTHER TRAUMA REPORTED. REVISION OCCURRED DUE TO USER ERROR. THE SURGEON WAS NOT HAPPY WITH THE STABILITY OF THE INITIAL CONSTRUCT, AS THE GLENOSPHERE WAS PUT IN "WITH A LOT OF INCLINATION", ACCORDING TO THE DISTRIBUTOR; THE GLENOSPHERE WAS NOT ADJUSTED DURING THE REVISION. A 40 MM + 6 135/145* HUMERAL STABILITY CUP WAS EXPLANTED. THIS ITEM WAS REPLACED WITH A 40 MM + 3 135/145* HUMERAL STABILITY CUP AND A 9 MM SPACER.
Patient problems: Inadequate Osseointegration
MAUDE report 23706078 ↗
Other Shoulder Prosthesis, Reverse Configuration devices
- DJO SURGICAL 508-36-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration508 reports
- DJO SURGICAL 508-32-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration433 reports
- DJO SURGICAL 508-32-103Encore Medical, L.P. · shoulder prosthesis, reverse configuration421 reports
- DJO SURGICAL 509-00-036Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall259 reports
- DJO SURGICAL 508-32-104Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall203 reports
- DJO SURGICAL 509-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration233 reports
- DELTA XTEND 130760000Depuy (Ireland) · shoulder prosthesis, reverse configuration223 reports
- DJO SURGICAL 508-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration122 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page