Sophysa · SOPHY MINI PROG.VALVE/8PRESS. · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03760124131093), received between Sep 8, 2016 and Nov 20, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Decrease in Pressure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Decrease in Pressure1
  • Device Dislodged or Dislocated1
  • Inaccurate Delivery1
  • Infusion or Flow Problem1
  • Obstruction of Flow1
  • Pressure Problem1

Patient problems

  • Hydrocephalus2
  • Nerve Damage1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 20, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A CHILDREN WAS HAVING A CONTROL MR SCAN AFTER THE PLACEMENT OF A PERITONEAL VENTRICLE SHUNT (FROM THE BRAIN, UNDER THE SKIN TO THE ABDOMINAL REGION). THE PATIENT WAS PLACED ON THE MR TABLE AND POSITIONED. RIGHT AFTER BEING POSITIONED THE PATIENT INFORMED HIS MOTHER (THAT WAS ON THE ROOM WITH HIM) THAT HE FELT MOVEMENT OF THE SHUNT IN THE ABDOMINAL REGION. THE PATIENT WAS REMOVED FROM THE SYSTEM AND EVALUATED BY THE TEAM THAT HAD OPERATED HIM. THE INFORMATION WE HAVE AT THE MOMENT IS THAT NO SERIOUS INJURY HAS OCCURRED, BUT THE PATIENT WILL HAVE TO BE SUBMITTED TO ANOTHER SURGERY TO CORRECT THE PLACEMENT OF THE SHUT IN THE ABDOMINAL REGION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20733421 ↗
  • InjuryOct 30, 2024Inaccurate Delivery; Inaccurate Delivery

    A PATIENT IN CHINA RECEIVED A SOPHY SM8 VALVE IMPLANT IN (B)(6) 2020 FOLLOWING AN ICH AND OPEN-SKULL SURGERY. TWO WEEKS AFTER THE PROCEDURE, IN EARLY SEPTEMBER, THE PATIENT'S DOCTOR SUSPECTED VALVE OVER DRAINAGE. THE VALVE'S RELEASE PRESSURE WAS INITIALLY SET TO 110 MMH2O AND WAS THEN ADJUSTED TO 140MMH2O. HOWEVER, TWO WEEKS AFTER THIS FIRST ADJUSTMENT, THE DOCTOR OBSERVED THAT THE RELEASE PRESSURE REMAINED TOO LOW AND RE-ADJUSTED IT TO 200MMH2O ON (B)(6) 2024. DESPITE THIS ADJUSTMENT, A PRESSURE MEASUREMENT TEST CONDUCTED BY THE HOSPITAL ON (B)(6) 2024, INDICATED THAT THE ACTUAL PRESSURE WAS ONLY 170MMH2O. THE OVER DRAINAGE IS NOT RESOLVED AND THE LOW PRESSURE HAS POTENTIALLY CAUSED SEVERE NEURAL DAMAGE. CURRENTLY THE VALVE REMAINS IMPLANTED IN THE PATIENT.

    Patient problems: Nerve Damage

    MAUDE report 20566303 ↗
  • MalfunctionOct 13, 2016Decrease in Pressure; Pressure Problem; Decrease in Pressure; Pressure Problem

    SUSPISION OF CHRONIC HYPODRAINAGE, DESPITE THE VALVE BEING IN THE LOW POSITION. THE VALVE HAD TO BE EXPLANTED.

    Patient problems: Hydrocephalus

    MAUDE report 6026372 ↗
  • MalfunctionSep 8, 2016Obstruction of Flow; Infusion or Flow Problem; Obstruction of Flow; Infusion or Flow Problem

    THE VALVE DOES NOT FUNCTION, THERE IS NO CSF FLOW.

    Patient problems: Hydrocephalus

    MAUDE report 5935468 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →