Skip to content

Sophysa SPVA-2010Recalls & reported problems

Sophysa · POLARIS SPVA KIT PREATTACHED/BO19-10 · Class 2 · Implant

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 03760124131314), received between Sep 15, 2016 and Jul 21, 2026 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Obstruction of Flow” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Obstruction of Flow3
  • Adverse Event Without Identified Device or Use Problem1
  • Device Difficult to Program or Calibrate1
  • Device Issue1
  • Inappropriate or Unexpected Reset1
  • Insufficient Device Problem Information1
  • Product Quality Problem1

Patient problems

  • Hydrocephalus3
  • Insufficient Information3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 21, 2026Obstruction of Flow

    The physician implanted the SPVA-2010. In their opinion, the valve was clogged and did not function properly. No testing was performed.

    Patient problems: Hydrocephalus

    MAUDE report 25886018 ↗
  • InjuryJul 20, 2026Obstruction of Flow

    THE PATIENT WAS IMPLANTED WITH A POLARIS VALVE ON (b)(6) 2026. DURING A FOLLOW-UP EXAMINATION CONDUCTED ON (b)(6) 2026, THE PHYSICIAN OBSERVED THAT THE VALVE WAS CLOGGED AND WAS NOT FUNCTIONING PROPERLY. AS A RESULT OF THE DEVICE MALFUNCTION, THE VALVE WAS SURGICALLY EXPLANTED AND REPLACED WITH ANOTHER VALVE.

    Patient problems: Hydrocephalus

    MAUDE report 25870291 ↗
  • MalfunctionAug 7, 2025Product Quality Problem; Product Quality Problem

    THE PHYSICIAN IMPLANTED THE BO19-10 AND HE FOUND THERE WAS NO CSF. HE PULLED OUT THE CATHETER AND FOUND THE MARKING POINT WAS INCORRECT. AFTER CHECKING THE CATHETER, IT WAS FOUND THAT THE MARKING POINTS ON THE CATHETER ARE OPPOSITE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22736559 ↗
  • MalfunctionJun 25, 2025Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    A PATIENT WAS IMPLANTED ON TUESDAY, (B)(6) 2025 WITH AN SPVA-2010 VALVE IN VENTRICULO-ATRIAL SHUNT (VAS). THE DOCTOR WAS HAVING DIFFICULTY ADJUSTING THE VALVE TO POSITION 2 (70 MMH2O). THE PATIENT PRESENTED A PICTURE OF HYPERDRAINAGE, AND THE DOCTOR WISHED TO ADJUST THE VALVE TO POSITION 3 OR 4. DESPITE NUMEROUS ATTEMPTS (USE OF A SINGLE MAGNET, DOUBLE MAGNET, WITH OR WITHOUT SELECTOR, REPEATED BACK-AND-FORTH MOVEMENTS, ETC.), IT WAS IMPOSSIBLE TO CHANGE THE VALVE POSITION. AFTER ABOUT AN HOUR OF UNSUCCESSFUL ATTEMPTS, THE DECISION WAS TAKEN TO RE-INTERVENE SURGICALLY AND REPLACE THE VALVE WITH ANOTHER SPVA. THE OPERATION WAS PERFORMED WITHOUT COMPLICATIONS ON SATURDAY, (B)(6) 2025.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 22327022 ↗
  • MalfunctionApr 28, 2025Obstruction of Flow; Obstruction of Flow

    THE VALVE DOES NOT DRAIN.

    Patient problems: Insufficient Information

    MAUDE report 21917349 ↗
  • MalfunctionFeb 19, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    A MALFUNCTION HAS BEEN OBSERVED ON A SPVA-2010 DEVICE BY THE HOSPITAL PRACTITIONERS. IT CONCERNS A VENTRICULAR SHUNT VALVE, IMPLANTED ON (B)(6) 2021, WHICH PRESENTED NO CONNECTION PROBLEMS AT THE IMPLANTATION TIME, BUT THEN SHOWED TO BE DYSFUNCTIONAL ON USE. THE VALVE HAS BEEN REPLACED.

    Patient problems: Insufficient Information

    MAUDE report 21408543 ↗
  • MalfunctionJul 22, 2024Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    DURING POST-OPERATIVING FOLLOW UP, THE VALVE HAS MALFUNCTIONED.

    Patient problems: Insufficient Information

    MAUDE report 19791857 ↗
  • MalfunctionSep 15, 2016Device Issue; Adverse Event Without Identified Device or Use Problem; Device Issue

    A PATIENT SUFFERED FROM A DILATATION OF THE VENTRICULAR SYSTEM AFTER IMPLANTATION OF THE POLARIS VALVE. NO ALTERATION OF THE DISTAL CATHETER TO THE VALVE SYSTEM AND INTRACRANIAL CATHETER WAS OBSERVED DURING THE SURGICAL ACT, HOWEVER WITHOUT OPERATING THE SYSTEM PRIOR TO THE IMPLANTATION.

    Patient problems: Hydrocephalus

    MAUDE report 5952057 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.