Sophysa SPVBRecalls & reported problems

Sophysa · POLARIS SPV ADJUSTABLE VALVE/BURR HOLE RES. · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03760124131321), received between Nov 29, 2016 and Oct 18, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Obstruction of Flow” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Obstruction of Flow1

Patient problems

  • Hydrocephalus2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 18, 2024Obstruction of Flow; Obstruction of Flow

    THE PATIENT HAS BEEN IMPLANTED WITH THE POLARIS VALVE. DUE TO SUSPECTED SHAFT DYSFUNCTION, IT WAS REPLACED FROM SPVB TO SPVA VALVE ON (B)(6) 2024. THE VENTRICULAR CATHETER WAS ALSO REPLACED. THE REASON IS THE DIFFICULTY IN CHANGING THE PRESSURE OF THE VALVE BEFORE THE SURGERY AND THE VISIBLE DEPOSITS INSIDE THE REMOVED VALVE.

    Patient problems: Hydrocephalus

    MAUDE report 20479074 ↗
  • MalfunctionNov 29, 2016Adverse Event Without Identified Device or Use Problem

    A (B)(6) PATIENT PRESENTED SHUNT FAILURE SYMPTOMS. THE VALVE WAS ADJUSTED ON HIGH PRESSURE AND THE FLOW WAS EXCELLENT. THE SYMPTOMS IMPROVED BUT RECURRED. THE USER BELIEVES THAT VALVE TURNED DOWN FROM THE ORIGINAL SETTING TO A LOWER SETTING. THE VALVE WAS ADJUSTED ON HIGH PRESSURE AND THE PROXIMAL FLOW WAS GOOD BUT AN X-RAY EXAMINATION CONFIRMED THAT THE VALVE WAS SET AT A LOWER POSITION. PATIENT WAS TAKEN TO THE OPERATING ROOM. THE VALVE WAS DISCONNECTED FROM THE CATHETERS AND IT SHOWED A GOOD PROXIMAL AND DISTAL FLOW.

    Patient problems: Hydrocephalus

    MAUDE report 6131919 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →