SpineJack 0909200058Recalls & reported problems
Stryker Corporation · EXPANSION KIT 5.8MM · Class 2 · Implant
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03760253800495), received between Jul 26, 2022 and Jul 29, 2025 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Failure4
- Separation Failure2
- Fracture1
- Migration1
- Patient Device Interaction Problem1
- Positioning Problem1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 29, 2025Activation Failure; Activation Failure
THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: ACTIVATION FAILURE. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22646533 ↗MalfunctionApr 21, 2025Separation Failure; Activation Failure; Separation Failure; Activation Failure
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY FAILED TO ACTIVATE AND SEPARATE. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876247 ↗InjuryApr 7, 2025Use of Device Problem; Activation Failure; Migration; Use of Device Problem; Activation Failure; Migration
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21786551 ↗MalfunctionOct 18, 2024Separation Failure; Activation Failure; Separation Failure; Activation Failure
IT WAS REPORTED THAT DURING A BIPEDICULAR IMPLANT PROCEDURE AT T12, ONE OF THE SPINEJACK IMPLANTS EXPERIENCED A LOT OF PRESSURE DUE TO HARD BONE AND WOULD NOT EXPAND. ADDITIONALLY, THE SAME IMPLANT WOULD NOT RELEASE FROM THE RELEASE PIN. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS A UNIPEDICULAR CASE WITH ONLY CEMENT INJECTED ON ONE SIDE. NO DELAY, MEDICAL INTERVENTION, OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20480511 ↗MalfunctionJun 19, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
DURING A 2-LEVEL BIPEDICULAR PROCEDURE AT T11 AND T12 IN A PATIENT WITH SOFT BONE AND CANCER, THE OPERATOR NOTED A CHANGE IN THE INFERIOR ENDPLATE OF THE T10 LEVEL AFTER EXPANSION OF THE SPINEJACK IMPLANTS. THE CHANGE WAS DESCRIBED AS "PUSHED IN A LITTLE." THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19569755 ↗MalfunctionMay 24, 2023Positioning Problem; Positioning Problem
IT WAS REPORTED THAT DURING A BIPEDICULAR IMPLANT PROCEDURE THE EXPANSION ASSEMBLY ON ONE SIDE SEEMED TO ADVANCE FURTHER THAN EXPECTED WITHIN THE ACCESS CANNULA, PRESENTING THE POTENTIAL FOR THE IMPLANT TO ALSO REACH AN EXCESSIVE DEPTH WITHIN THE VERTEBRAL BODY. THERE WAS NO REPORT OF THE IMPLANT ACTUALLY REACHING AN EXCESSIVE DEPTH. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY-SIGNIFICANT DELAY AND NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16993811 ↗MalfunctionJul 26, 2022Fracture; Fracture
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15107756 ↗
Other Cement, Bone, Vertebroplasty devices
- CONFIDENCE SPINAL CEMENT SYSTEM 283910000Medos International Sàrl · CEMENT, BONE, VERTEBROPLASTY5 deaths163 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283913000Medos International Sàrl · CEMENT, BONE, VERTEBROPLASTY8 deaths146 reports
- KYPHON® Xpede™ Bone Cement and KYPHON® Mixer Pack CX01BMedtronic Sofamor Danek, Inc. · CEMENT, BONE, VERTEBROPLASTY3 deaths67 reports
- SYNFLATE 03.804.701SSynthes GmbH · CEMENT, BONE, VERTEBROPLASTY60 reports
- KYPHON® HV-R® Bone Cement C01AMedtronic Sofamor Danek, Inc. · CEMENT, BONE, VERTEBROPLASTY57 reports
- iVAS 0705315000Stryker Corporation · CEMENT, BONE, VERTEBROPLASTY1 recall3 reports
- Cortoss 2101-0005Stryker Corporation · CEMENT, BONE, VERTEBROPLASTY1 recall3 reports
- iVAS 0705310000Stryker Corporation · CEMENT, BONE, VERTEBROPLASTY1 recall1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.