pH Electrode 3111717180Recalls & reported problems
Roche Diagnostics GmbH · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04015630014248), received between Oct 24, 2022 and Oct 24, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “High Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- High Test Results1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 24, 2022High Test Results; No Apparent Adverse Event; High Test Results; No Apparent Adverse Event
THERE WAS AN ALLEGATION OF QUESTIONABLE ROCHE OMNI ELECTRODE PH RESULTS FROM 3 PATIENT SAMPLES ON A COBAS B 221 ROCHE OMNI S6 SYSTEM. FOR SAMPLE 1, THE PH RESULT WAS 7.45. THE RESULT ON AN AVL ANALYZER WAS 7.33. FOR SAMPLE 2, THE PH RESULT WAS 7.39. THE RESULT ON AN AVL ANALYZER WAS 7.28. FOR SAMPLE 3, THE PH RESULT WAS 7.43. THE RESULT ON AN AVL ANALYZER WAS 7.30. THE INITIAL RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE REPEAT RESULTS FROM THE AVL ANALYZER WERE DEEMED CORRECT. THE ELECTRODE SERIAL NUMBER AND EXPIRATION DATE WERE REQUESTED BUT NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15662816 ↗
Other ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH devices
- GEM Premier 5000 00024019255Instrumentation Laboratory Company · Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph5 recalls5 deaths7 reports
- ABL800 FLEX 393-800Radiometer Medical ApS · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls81 reports
- ABL90 FLEX Analyzer 393-090Radiometer Medical ApS · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall112 reports
- i-STAT 03P88-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH105 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 300 TEST 00055430010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls3 deaths3 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 450 TEST 00055445010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls2 reports
- GEM Premier 5000 PAK BG/COOX 75 TEST 00055407504Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 600 TEST 00055360010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.