Roche 9180 Sodium electrode 03110419180Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630028702), received between Sep 14, 2018 and Feb 10, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results2
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 10, 2025Low Test Results; Low Test Results
THE INITIAL REPORTER COMPLAINED OF DISCREPANT LOW RESULTS FOR 1 PATIENT SAMPLE TESTED FOR SODIUM (NA) ON A (B)(6) ELECTROLYTE ANALYZER. THE INITIAL RESULT WAS 109 MMOL/L WITH A DATA FLAG. THE CUSTOMER PERFORMED CLEANING AND CONDITIONING AND THE REPEAT RESULT WAS 138 MMOL/L. THE CUSTOMER REPEATED THE SAMPLE AND THE RESULT WAS AGAIN 109 MMOL/L. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE RESULT OF 138 MMOL/L WAS BELIEVED TO BE CORRECT BASED ON THE PATIENT¿S PREVIOUS RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21344710 ↗MalfunctionSep 14, 2018Low Test Results; Low Test Results
THE CUSTOMER COMPLAINED OF DISCREPANT ISE INDIRECT NA FOR GEN.2 RESULTS FOR 1 PATIENT SAMPLE ON A ROCHE 9180 ELECTROLYTE ANALYZER COMPARED TO AN ELYTE INSTRUMENT. THE NA RESULT WAS 132 MMOL/L ON THE 9180 ELECTROLYTE ANALYZER. THE NA RESULT WAS 141 MMOL/L ON THE ELYTE INSTRUMENT. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. NO ERRONEOUS RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE CUSTOMER PERFORMED DAILY MAINTENANCE AND CALIBRATION. THE CUSTOMER REPEATED THE SAMPLE. THE REPEAT NA RESULT WAS 137 MMOL/L ON THE 9180 ELECTROLYTE ANALYZER AND 142 MMOL/L ON THE ELYTE INSTRUMENT. THE NA ELECTRODE LOT NUMBER AND EXPIRATION DATE WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7877768 ↗
Other ELECTRODE, ION SPECIFIC, SODIUM devices
- i-STAT 09P31-25Abbott Point Of Care Inc. · ELECTRODE, ION SPECIFIC, SODIUM423 reports
- VITROS 8379034Ortho-Clinical Diagnostics, Inc. · ELECTRODE, ION SPECIFIC, SODIUM1 recall76 reports
- Sodium electrode 10825468001Roche Diagnostics GmbH · ELECTRODE, ION SPECIFIC, SODIUM99 reports
- Sodium electrode 10825468001Roche Diagnostics GmbH · ELECTRODE, ION SPECIFIC, SODIUM26 reports
- Sodium Electrode 21029371001Roche Diagnostics GmbH · ELECTRODE, ION SPECIFIC, SODIUM25 reports
- GEM Premier ChemSTAT PAK Na+,K+,Ca++,Cl-, Glu, Crea, BUN, tCO2, Hct, Lac,pH,pCO2 00044345010Instrumentation Laboratory Company · Electrode, Ion Specific, Sodium19 reports
- i-STAT 03P81-25Abbott Point Of Care Inc. · ELECTRODE, ION SPECIFIC, SODIUM15 reports
- i-STAT 03P82-25Abbott Point Of Care Inc. · ELECTRODE, ION SPECIFIC, SODIUM10 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.