Tinaquant Haptoglobin ver2 3005593322Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04015630916580), received between Dec 5, 2022 and Dec 5, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Low Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 5, 2022Low Test Results; Low Test Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR SAMPLES FROM ONE PATIENT TESTED WITH HAPT2 (TINA-QUANT HAPTOGLOBIN VER.2) ASSAY ON A COBAS 8000 C 502 MODULE. THE FIRST SAMPLE INITIALLY RESULTED IN A HAPT2 VALUE OF 0.02 G/L. 24 HOURS LATER, THE SAMPLE WAS REPEATED TWICE, RESULTING IN VALUES OF 4.76 G/L AND 4.95 G/L. ON (B)(6) 2022, A SECOND SAMPLE FROM THE SAME PATIENT WAS MEASURED. THE SAMPLE INITIALLY RESULTED IN A HAPT2 VALUE OF 4.77 G/L. THE SAMPLE WAS REPEATED THREE TIMES, RESULTING IN VALUES OF 4.92 G/L, 4.76 G/L AND 4.82 G/L. THE INITIAL QUESTIONABLE LOW RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THE REAGENT LOT WAS 642662 WITH AN EXPIRATION DATE OF 31-MAR-2024.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15920590 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page