Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04015630919512), received between Feb 22, 2022 and Jul 26, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Non Reproducible Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Non Reproducible Results3
- High Test Results2
- Incorrect, Inadequate or Imprecise Result or Readings2
- Material Integrity Problem2
- Chemical Problem1
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 26, 2024Incorrect, Inadequate or Imprecise Result or Readings; Non Reproducible Results; Incorrect, Inadequate or Imprecise Result or Readings; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ALP2 ALKALINE PHOSPHATASE ACC. TO IFCC GEN.2 AND UREAL UREA/BUN RESULTS FOR 2 PATIENT SAMPLES ON A COBAS 8000 C702 MODULE. THIS MEDWATCH WILL COVER ALP2. REFER TO MEDWATCH WITH A1 PATIENT IDENTIFIER PT-00070199 FOR INFORMATION ON THE UREA RESULTS. FOR SAMPLE 1, THE INITIAL ALP RESULT WAS 9 U/L. THE RESULT DID NOT MATCH THE PATIENT'S CLINICAL HISTORY WHICH PROMPTED THE CUSTOMER TO REPEAT THE SAMPLE. THE REPEAT RESULT WAS 144 U/L. FOR SAMPLE 2, THE INITIAL UREA RESULT WAS 0.5 MMOL/L. THE RESULT WAS LOWER THAN THE MEASURING RANGE WHICH PROMPTED THE CUSTOMER TO REPEAT THE SAMPLE. THE REPEAT RESULT WAS 11.6 MMOL/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19840717 ↗MalfunctionJan 19, 2024Chemical Problem; Material Integrity Problem; Non Reproducible Results; Chemical Problem; Material Integrity Problem; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ALP2 ALKALINE PHOSPHATASE ACC. TO IFCC GEN.2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS 6000 C501 MODULE. THE INITIAL ALP2 RESULT WAS <3 IU/ML. THE SAMPLE WAS REPEATED AND THE SAME RESULT WAS OBTAINED. A NEW SAMPLE WAS COLLECTED THE NEXT DAY AND THE RESULT WAS 155 IU/ML. THE RESULT OF 155 IU/ML WAS DEEMED CORRECT AS IT MATCHED THE PATIENT'S PREVIOUS RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18544472 ↗MalfunctionSep 25, 2023High Test Results; Material Integrity Problem; High Test Results; Material Integrity Problem
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 4 PATIENTS' SAMPLES TESTED WITH ALKALINE PHOSPHATASE GEN.2 (ALP2) ASSAY ON A COBAS INTEGRA 400 PLUS ANALYZER. SAMPLE 1: INITIAL RESULT: 275 U/L. RERUN RESULT: 74.14 U/L. SAMPLE 2: INITIAL RESULT: 265 U/L. RERUN RESULT: 67.98 U/L. SAMPLE 3: INITIAL RESULT: 308 U/L. RERUN RESULT: 80.26 U/L. SAMPLE 4: INITIAL RESULT: 250 U/L. RERUN RESULT: 64.58 U/L. THE PHYSICIAN QUESTIONED THE RESULTS AND THE SAMPLES WERE RERUN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17806551 ↗MalfunctionSep 11, 2023Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results; Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results
HOLD PN 9/11. WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 2 PATIENTS' PLASMA SAMPLES TESTED WITH ALKALINE PHOSPHATASE GEN.2 (ALP2) ASSAY ON A COBAS INTEGRA 400 PLUS ANALYZER. SAMPLE 1 (PATIENT 1): INITIAL RESULT: 15 U/L. 1ST REPEAT RESULT: 67 U/L. 2ND REPEAT RESULT: 425 U/L. SAMPLE 2 (PATIENT 2): INITIAL RESULT: 15 U/L. 1ST REPEAT RESULT: 392 U/L. THE PHYSICIAN QUESTIONED THE RESULTS. THE SAMPLES WERE THEN REPEATED. THE RESULTS OF 425 U/L FOR SAMPLE 1 AND 392 U/L FOR SAMPLE 2 WERE DEEMED TO BE CORRECT BASED ON THE PATIENTS' PREVIOUS RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17719128 ↗MalfunctionNov 10, 2022Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ALP2 ALKALINE PHOSPHATASE ACC. TO IFCC GEN.2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS PRO C 503 ANALYTICAL UNIT. THE INITIAL ALP2 RESULT WAS 337 U/L. THE REPEAT RESULT WAS 182 U/L. THE SAMPLE WAS REPEATED AGAIN ON ANOTHER ANALYZER AND THE RESULT WAS 190 U/L. THE INITIAL RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY. THE REAGENT LOT NUMBER IS 654864 AND THE EXPIRATION DATE IS 31-MAY-2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15770603 ↗MalfunctionFeb 22, 2022High Test Results; High Test Results
THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT TESTED FOR ALP2 ALKALINE PHOSPHATASE ACC. TO IFCC GEN.2 (ALP2) ON A COBAS INTEGRA 400 PLUS. THE INITIAL RESULT WAS 359.4 U/L. THE REPEAT RESULT WAS 214.2 U/L. THESE RESULTS DID NOT MATCH THE PREVIOUS ALP2 RESULT FROM THE PATIENT OF 77 U/L FROM (B)(6) 2022. THE CUSTOMER PERFORMED PRECISION TESTING WITH THE SAMPLE AND OBTAINED RESULTS BETWEEN 86.7 U/L AND 178.3 U/L. THE SAMPLE WAS REPEATED ON A DIFFERENT INTEGRA 400 PLUS WITH RESULTS OF 77.9 U/L, 75.7 U/L, 77.3 U/L AND 74.9 U/L. A 2ND SAMPLE WAS OBTAINED AND THE RESULTS WERE 124.7 U/L, 76.6 U/L, 83.4 U/L AND 74.6 U/L AND 76.2 U/L. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE INTEGRA 400 PLUS SERIAL NUMBER WAS (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13580140 ↗
Other NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES devices
- VITROS 6844296Ortho-Clinical Diagnostics, Inc. · NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES4 recalls0 reports
- VITROS 1053180Ortho-Clinical Diagnostics, Inc. · NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES1 recall0 reports
- Alkaline phosphatase acc. to IFCC Gen.2 08056757190Roche Diagnostics GmbH · NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES21 reports
- Alkaline Phosphatase acc to IFCC Gen2 3333701190Roche Diagnostics GmbH · NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES10 reports
- Alinity 08P2030Abbott GmbH · NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.