Alkaline Phosphatase acc to IFCC Gen2 3333752190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04015630919512), received between Feb 22, 2022 and Jul 26, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Non Reproducible Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Non Reproducible Results3
- High Test Results2
- Incorrect, Inadequate or Imprecise Result or Readings2
- Material Integrity Problem2
- Chemical Problem1
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 26, 2024Incorrect, Inadequate or Imprecise Result or Readings; Non Reproducible Results; Incorrect, Inadequate or Imprecise Result or Readings; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ALP2 ALKALINE PHOSPHATASE ACC. TO IFCC GEN.2 AND UREAL UREA/BUN RESULTS FOR 2 PATIENT SAMPLES ON A COBAS 8000 C702 MODULE. THIS MEDWATCH WILL COVER ALP2. REFER TO MEDWATCH WITH A1 PATIENT IDENTIFIER PT-00070199 FOR INFORMATION ON THE UREA RESULTS. FOR SAMPLE 1, THE INITIAL ALP RESULT WAS 9 U/L. THE RESULT DID NOT MATCH THE PATIENT'S CLINICAL HISTORY WHICH PROMPTED THE CUSTOMER TO REPEAT THE SAMPLE. THE REPEAT RESULT WAS 144 U/L. FOR SAMPLE 2, THE INITIAL UREA RESULT WAS 0.5 MMOL/L. THE RESULT WAS LOWER THAN THE MEASURING RANGE WHICH PROMPTED THE CUSTOMER TO REPEAT THE SAMPLE. THE REPEAT RESULT WAS 11.6 MMOL/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19840717 ↗MalfunctionJan 19, 2024Chemical Problem; Material Integrity Problem; Non Reproducible Results; Chemical Problem; Material Integrity Problem; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ALP2 ALKALINE PHOSPHATASE ACC. TO IFCC GEN.2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS 6000 C501 MODULE. THE INITIAL ALP2 RESULT WAS <3 IU/ML. THE SAMPLE WAS REPEATED AND THE SAME RESULT WAS OBTAINED. A NEW SAMPLE WAS COLLECTED THE NEXT DAY AND THE RESULT WAS 155 IU/ML. THE RESULT OF 155 IU/ML WAS DEEMED CORRECT AS IT MATCHED THE PATIENT'S PREVIOUS RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18544472 ↗MalfunctionSep 25, 2023High Test Results; Material Integrity Problem; High Test Results; Material Integrity Problem
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 4 PATIENTS' SAMPLES TESTED WITH ALKALINE PHOSPHATASE GEN.2 (ALP2) ASSAY ON A COBAS INTEGRA 400 PLUS ANALYZER. SAMPLE 1: INITIAL RESULT: 275 U/L. RERUN RESULT: 74.14 U/L. SAMPLE 2: INITIAL RESULT: 265 U/L. RERUN RESULT: 67.98 U/L. SAMPLE 3: INITIAL RESULT: 308 U/L. RERUN RESULT: 80.26 U/L. SAMPLE 4: INITIAL RESULT: 250 U/L. RERUN RESULT: 64.58 U/L. THE PHYSICIAN QUESTIONED THE RESULTS AND THE SAMPLES WERE RERUN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17806551 ↗MalfunctionSep 11, 2023Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results; Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results
HOLD PN 9/11. WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 2 PATIENTS' PLASMA SAMPLES TESTED WITH ALKALINE PHOSPHATASE GEN.2 (ALP2) ASSAY ON A COBAS INTEGRA 400 PLUS ANALYZER. SAMPLE 1 (PATIENT 1): INITIAL RESULT: 15 U/L. 1ST REPEAT RESULT: 67 U/L. 2ND REPEAT RESULT: 425 U/L. SAMPLE 2 (PATIENT 2): INITIAL RESULT: 15 U/L. 1ST REPEAT RESULT: 392 U/L. THE PHYSICIAN QUESTIONED THE RESULTS. THE SAMPLES WERE THEN REPEATED. THE RESULTS OF 425 U/L FOR SAMPLE 1 AND 392 U/L FOR SAMPLE 2 WERE DEEMED TO BE CORRECT BASED ON THE PATIENTS' PREVIOUS RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17719128 ↗MalfunctionNov 10, 2022Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ALP2 ALKALINE PHOSPHATASE ACC. TO IFCC GEN.2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS PRO C 503 ANALYTICAL UNIT. THE INITIAL ALP2 RESULT WAS 337 U/L. THE REPEAT RESULT WAS 182 U/L. THE SAMPLE WAS REPEATED AGAIN ON ANOTHER ANALYZER AND THE RESULT WAS 190 U/L. THE INITIAL RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY. THE REAGENT LOT NUMBER IS 654864 AND THE EXPIRATION DATE IS 31-MAY-2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15770603 ↗MalfunctionFeb 22, 2022High Test Results; High Test Results
THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT TESTED FOR ALP2 ALKALINE PHOSPHATASE ACC. TO IFCC GEN.2 (ALP2) ON A COBAS INTEGRA 400 PLUS. THE INITIAL RESULT WAS 359.4 U/L. THE REPEAT RESULT WAS 214.2 U/L. THESE RESULTS DID NOT MATCH THE PREVIOUS ALP2 RESULT FROM THE PATIENT OF 77 U/L FROM (B)(6) 2022. THE CUSTOMER PERFORMED PRECISION TESTING WITH THE SAMPLE AND OBTAINED RESULTS BETWEEN 86.7 U/L AND 178.3 U/L. THE SAMPLE WAS REPEATED ON A DIFFERENT INTEGRA 400 PLUS WITH RESULTS OF 77.9 U/L, 75.7 U/L, 77.3 U/L AND 74.9 U/L. A 2ND SAMPLE WAS OBTAINED AND THE RESULTS WERE 124.7 U/L, 76.6 U/L, 83.4 U/L AND 74.6 U/L AND 76.2 U/L. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE INTEGRA 400 PLUS SERIAL NUMBER WAS (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13580140 ↗
FDA clearance and classification
Cleared through Special 510(k) K033185 on Oct 14, 2003 (decision: Substantially Equivalent). Submitted by Roche Diagnostics Corp. as “Cobas Integra Alp Ifcc Gen.2”. The FDA review took 13 days. FDA record ↗
The FDA regulates product code CJE (nitrophenylphosphate, alkaline phosphatase or isoenzymes) under 21 CFR 862.1050 as a Class II device. 21 CFR 862.1050 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Alkaline Phosphatase acc to IFCC Gen2 3333752190
Has the Alkaline Phosphatase acc to IFCC Gen2 3333752190 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04015630919512), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Alkaline Phosphatase acc to IFCC Gen2 3333752190?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Feb 22, 2022 and Jul 26, 2024. In those reports, the most common specific device problems are “Non Reproducible Results” (3), “High Test Results” (2) and “Incorrect, Inadequate or Imprecise Result or Readings” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Alkaline Phosphatase acc to IFCC Gen2 3333752190?
None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Alkaline Phosphatase acc to IFCC Gen2 3333752190 FDA approved?
It was cleared through Special 510(k) K033185 on Oct 14, 2003. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Alkaline Phosphatase acc to IFCC Gen2 3333752190, and is it still sold?
The FDA's device database lists Roche Diagnostics GmbH as the company responsible for it (brand Alkaline Phosphatase acc to IFCC Gen2, model 3333752190). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES devices
- VITROS 6844296Ortho-Clinical Diagnostics, Inc. · NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES4 recalls0 reports
- VITROS 1053180Ortho-Clinical Diagnostics, Inc. · NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES1 recall0 reports
- Alkaline phosphatase acc. to IFCC Gen.2 08056757190Roche Diagnostics GmbH · NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES21 reports
- Alkaline Phosphatase acc to IFCC Gen2 3333701190Roche Diagnostics GmbH · NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES10 reports
- Alinity 08P2030Abbott GmbH · NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page