Skip to content

Tinaquant IgA Gen2 3507343190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630919680), received between Oct 14, 2022 and May 7, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (1 report).

1 other recall cover devices cleared under the same FDA submission (K040435); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE INITIAL REPORTER RECEIVED QUESTIONABLE TINA-QUANT IGA GEN.2 (IGA) RESULTS FROM MULTIPLE PATIENT SAMPLES TESTED ON THE COBAS 6000 C 501 (UL) V. THE FOLLOWING ARE EXAMPLES OF DISCREPANT RESULTS: PATIENT SAMPLE 1 THE INITIAL RESULT FROM THE MODULE WAS 356 MG/DL. THE REPEAT RESULT FROM ANOTHER C 501 MODULE WAS 169 MG/DL. THE REPEAT RESULT WAS DEEMED CORRECT. PATIENT SAMPLE 2 THE INITIAL RESULT FROM THE MODULE WAS 437 MG/DL. THE REPEAT RESULT FROM ANOTHER C 501 MODULE WAS 287 MG/DL. PATIENT SAMPLE 3 THE INITIAL RESULT FROM THE MODULE WAS 348 MG/DL. THE REPEAT RESULT FROM ANOTHER C 501 MODULE WAS 252 MG/DL. PATIENT SAMPLE 4 THE INITIAL RESULT FROM THE MODULE WAS 52 MG/DL. THE REPEAT RESULT FROM ANOTHER C 501 MODULE WAS 151 MG/DL. SOME RESULTS HAD INVALIDATING FLAGS, PROMPTING THE PATIENT SAMPLES TO BE RERUN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21975791 ↗
  • MalfunctionOct 14, 2022High Test Results; High Test Results

    THE INITIAL REPORTER RECEIVED A QUESTIONABLE IGA-2 TINA-QUANT IGA GEN.2 RESULT FOR ONE PATIENT SAMPLE TESTED ON THE COBAS 8000 COBAS C 502 MODULE. THE INITIAL RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THE RESULT WAS DIFFERENT FROM THE PREVIOUS RESULT OF THE PATIENT FROM 4 DAYS PRIOR. THE INITIAL RESULT WAS 71 MG/DL. THE FIRST REPEAT RESULT WAS 0 MG/DL WITH A DATA FLAG. THE SECOND REPEAT RESULT WAS 3 MG/DL WITH A DATA FLAG. THE THIRD REPEAT RESULT WAS 0 MG/DL WITH A DATA FLAG. THE FOURTH REPEAT RESULT WAS 2 MG/DL WITH A DATA FLAG. THE REPORTER RELEASED A RESULT OF <5 MG/DL. THE REPEAT RESULT WAS DEEMED CORRECT. THE REAGENT LOT NUMBER IS 58129801 WITH AN EXPIRATION DATE OF 31-AUG-2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15608084 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K040435. One submission often covers many models, so check the product and lot details in each record.

  • Z-1979-2019 · Jun 3, 2019 · Terminated
    The Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample to form an antigen/antibody complex which is measured turbidimetrical

Other IGA, ANTIGEN, ANTISERUM, CONTROL devices

All IGA, ANTIGEN, ANTISERUM, CONTROL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page