Tinaquant IgA Gen2 3507343190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630919680), received between Oct 14, 2022 and May 7, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (1 report).
1 other recall cover devices cleared under the same FDA submission (K040435); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE INITIAL REPORTER RECEIVED QUESTIONABLE TINA-QUANT IGA GEN.2 (IGA) RESULTS FROM MULTIPLE PATIENT SAMPLES TESTED ON THE COBAS 6000 C 501 (UL) V. THE FOLLOWING ARE EXAMPLES OF DISCREPANT RESULTS: PATIENT SAMPLE 1 THE INITIAL RESULT FROM THE MODULE WAS 356 MG/DL. THE REPEAT RESULT FROM ANOTHER C 501 MODULE WAS 169 MG/DL. THE REPEAT RESULT WAS DEEMED CORRECT. PATIENT SAMPLE 2 THE INITIAL RESULT FROM THE MODULE WAS 437 MG/DL. THE REPEAT RESULT FROM ANOTHER C 501 MODULE WAS 287 MG/DL. PATIENT SAMPLE 3 THE INITIAL RESULT FROM THE MODULE WAS 348 MG/DL. THE REPEAT RESULT FROM ANOTHER C 501 MODULE WAS 252 MG/DL. PATIENT SAMPLE 4 THE INITIAL RESULT FROM THE MODULE WAS 52 MG/DL. THE REPEAT RESULT FROM ANOTHER C 501 MODULE WAS 151 MG/DL. SOME RESULTS HAD INVALIDATING FLAGS, PROMPTING THE PATIENT SAMPLES TO BE RERUN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21975791 ↗MalfunctionOct 14, 2022High Test Results; High Test Results
THE INITIAL REPORTER RECEIVED A QUESTIONABLE IGA-2 TINA-QUANT IGA GEN.2 RESULT FOR ONE PATIENT SAMPLE TESTED ON THE COBAS 8000 COBAS C 502 MODULE. THE INITIAL RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THE RESULT WAS DIFFERENT FROM THE PREVIOUS RESULT OF THE PATIENT FROM 4 DAYS PRIOR. THE INITIAL RESULT WAS 71 MG/DL. THE FIRST REPEAT RESULT WAS 0 MG/DL WITH A DATA FLAG. THE SECOND REPEAT RESULT WAS 3 MG/DL WITH A DATA FLAG. THE THIRD REPEAT RESULT WAS 0 MG/DL WITH A DATA FLAG. THE FOURTH REPEAT RESULT WAS 2 MG/DL WITH A DATA FLAG. THE REPORTER RELEASED A RESULT OF <5 MG/DL. THE REPEAT RESULT WAS DEEMED CORRECT. THE REAGENT LOT NUMBER IS 58129801 WITH AN EXPIRATION DATE OF 31-AUG-2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15608084 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K040435. One submission often covers many models, so check the product and lot details in each record.
- Z-1979-2019 · Jun 3, 2019 · TerminatedThe Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample to form an antigen/antibody complex which is measured turbidimetrical
Other IGA, ANTIGEN, ANTISERUM, CONTROL devices
- IgA IA3832Randox Laboratories Limited · Iga, Antigen, Antiserum, Control1 recall0 reports
- Tina-quant IgA Gen.2 08057885190Roche Diagnostics GmbH · IGA, ANTIGEN, ANTISERUM, CONTROL4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page