Skip to content

Tinaquant IgG Gen2 3507432190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630919710), received between Jun 13, 2024 and Sep 15, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Non Reproducible Results1
  • Unable to Obtain Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2026Non Reproducible Results

    There was an allegation of questionable Tina-quant IgG Gen.2 results from the cobas 8000 cobas c 502 module.The samples were repeated on 29-July-2026.Sample 1 initial result was 13.9 g/L, and the repeat result was 7.87 g/L.Sample 2 initial result was 15.02 g/L, and the repeat result was 2 6.9 g/L.Sample 3 initial result was 17.9 g/L, and the repeat result was 10.76 g/L.Sample 4 initial result was 4.48 g/L, and the repeat result was 9.91 g/L.Sample 5 initial result was 7.79 g/L, and the repeat result was 11.05 g/L.Sample 6 initial result was 6.58 g/L, and the repeat result was 8.92 g/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50771793 ↗
  • MalfunctionJun 13, 2024Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THERE WAS AN ALLEGATION OF QUESTIONABLE IGG-2 TINA-QUANT IGG GEN.2 RESULTS FOR 1 PATIENT ON A COBAS 8000 COBAS C 502 MODULE. THE DOCTOR QUESTIONED THE CSF IGG RESULTS FOR THE PATIENT. THEY CLAIMED THE RESULTS WERE IMPLAUSIBLE AS THE PATIENT WAS RECEIVING AN IGG SUBSTITUTION. THE INITIAL CSF IGG RESULT WAS 76.8 MG/L. THE SAMPLE WAS REPEATED TWICE AND THE RESULTS WERE 79.61 MG/L AND 79.61 MG/L. A SERUM SAMPLE WAS COLLECTED AT THE SAME TIME AS THE CSF SAMPLE. THE INITIAL SERUM IGG RESULT WAS 38.3 G/L AND THE REPEAT RESULT WAS 39.4 G/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19530994 ↗

Other IGG, ANTIGEN, ANTISERUM, CONTROL devices

All IGG, ANTIGEN, ANTISERUM, CONTROL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page