Tinaquant IgG Gen2 3507432190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630919710), received between Jun 13, 2024 and Sep 15, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings1
- Non Reproducible Results1
- Unable to Obtain Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2026Non Reproducible Results
There was an allegation of questionable Tina-quant IgG Gen.2 results from the cobas 8000 cobas c 502 module.The samples were repeated on 29-July-2026.Sample 1 initial result was 13.9 g/L, and the repeat result was 7.87 g/L.Sample 2 initial result was 15.02 g/L, and the repeat result was 2 6.9 g/L.Sample 3 initial result was 17.9 g/L, and the repeat result was 10.76 g/L.Sample 4 initial result was 4.48 g/L, and the repeat result was 9.91 g/L.Sample 5 initial result was 7.79 g/L, and the repeat result was 11.05 g/L.Sample 6 initial result was 6.58 g/L, and the repeat result was 8.92 g/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50771793 ↗MalfunctionJun 13, 2024Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings
THERE WAS AN ALLEGATION OF QUESTIONABLE IGG-2 TINA-QUANT IGG GEN.2 RESULTS FOR 1 PATIENT ON A COBAS 8000 COBAS C 502 MODULE. THE DOCTOR QUESTIONED THE CSF IGG RESULTS FOR THE PATIENT. THEY CLAIMED THE RESULTS WERE IMPLAUSIBLE AS THE PATIENT WAS RECEIVING AN IGG SUBSTITUTION. THE INITIAL CSF IGG RESULT WAS 76.8 MG/L. THE SAMPLE WAS REPEATED TWICE AND THE RESULTS WERE 79.61 MG/L AND 79.61 MG/L. A SERUM SAMPLE WAS COLLECTED AT THE SAME TIME AS THE CSF SAMPLE. THE INITIAL SERUM IGG RESULT WAS 38.3 G/L AND THE REPEAT RESULT WAS 39.4 G/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19530994 ↗
Other IGG, ANTIGEN, ANTISERUM, CONTROL devices
- VIDAS® 99735Biomerieux SA · Igg, Antigen, Antiserum, Control1 recall36 reports
- BN II System 11254820Siemens Healthcare Diagnostics Products GmbH · IGG, ANTIGEN, ANTISERUM, CONTROL17 reports
- Tina-quant IgG Gen.2 08057915190Roche Diagnostics GmbH · IGG, ANTIGEN, ANTISERUM, CONTROL2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page