Elecsys Troponin T STAT Assay 4660307190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04015630921621), received between Aug 17, 2018 and Apr 14, 2023 — classified as 6 malfunctions.
The most frequently reported specific device problem is “High Test Results” (3 reports).
4 other recalls cover devices cleared under the same FDA submission (K082699); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- High Test Results3
- Incorrect, Inadequate or Imprecise Result or Readings3
- Non Reproducible Results3
- Incorrect Measurement1
- Insufficient Device Problem Information1
- Unable to Obtain Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 14, 2023Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH THE ELECSYS TROPONIN T STAT ASSAY ON COBAS 6000 E601 MODULE SERIAL NUMBER (B)(4). THE SAMPLE INITIALLY RESULTED IN A TROPONIN T VALUE OF 80.56 NG/ML AND THIS VALUE WAS REPORTED OUTSIDE OF THE LABORATORY. A SECOND SAMPLE COLLECTED FROM THE PATIENT RESULTED IN A TROPONIN T VALUE OF 9.10 NG/ML, SO THE DOCTOR QUESTIONED THE INITIAL VALUE FROM THE COMPLAINED SAMPLE. THE SAMPLE WAS REPEATED TWICE ON THE SAME ANALYZER, RESULTING IN TROPONIN T VALUES OF 14.89 NG/ML AND 17.37 NG/ML. THE 14.89 NG/ML VALUE WAS DEEMED CORRECT. THE SAMPLE WAS ALSO REPEATED ON A SECOND E601 ANALYZER (UNKNOWN SERIAL NUMBER), RESULTING IN A TROPONIN T VALUE OF 18.66 NG/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16746183 ↗MalfunctionFeb 28, 2023Non Reproducible Results; Non Reproducible Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR FIVE PATIENT SAMPLES TESTED WITH THE ELECSYS TROPONIN T HS STAT ASSAY ON A COBAS E 411 ANALYZER (DISK SYSTEM). PATIENT SAMPLE 1: THE SAMPLE INITIALLY RESULTED IN A TROPONIN T HS STAT VALUE OF 23.86 PG/ML. THE SAMPLE WAS REPEATED THREE TIMES, RESULTING IN VALUES OF 11.19 PG/ML, 35.52 PG/ML AND 37.43 PG/ML. THE RESULT DEEMED CORRECT WAS 8.3 NG/L. PATIENT SAMPLE 2: THE SAMPLE INITIALLY RESULTED IN A TROPONIN T HS STAT VALUE OF 18.6 PG/ML. THE SAMPLE WAS REPEATED THREE TIMES, RESULTING IN VALUES OF 27.9 PG/ML, 12.9 PG/ML, AND 19.6 PG/ML. IT IS UNKNOWN WHICH RESULT WAS DEEMED CORRECT. PATIENT SAMPLE 3: THE SAMPLE INITIALLY RESULTED IN A TROPONIN T HS STAT VALUE OF 27.85 PG/ML AND REPEATED AS 12.91 PG/ML. THE RESULT DEEMED CORRECT WAS 78.8 NG/L PATIENT 4: THE SAMPLE INITIALLY RESULTED IN A TROPONIN T HS STAT VALUE OF 16.62 PG/ML. THE RESULT DEEMED CORRECT WAS 8.6 NG/L. PATIENT 5: THE SAMPLE INITIALLY RESULTED IN A TROPONIN T HS STAT VALUE OF 12.25 PG/ML AND REPEATED AS 21.86 PG/ML. THE RESULT DEEMED CORRECT WAS 12.5 NG/L. THE QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. IT WAS ASKED BUT IT IS UNKNOWN IF THE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16454133 ↗MalfunctionDec 6, 2022High Test Results; Insufficient Device Problem Information; Non Reproducible Results; High Test Results; Insufficient Device Problem Information; Non Reproducible Results
THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS TROPONIN T STAT ASSAY RESULTS FROM FIVE PATIENT SAMPLES TESTED ON THE COBAS 6000 E601 MODULE. THE INITIAL RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE PATIENT SAMPLES WERE RERUN ON THEIR OTHER E 601. THE REPEAT RESULTS WERE REPORTED TO BE LESS THAN THE LEVEL OF DETECTION WHICH IS 0.010 NG/ML. ON (B)(6) 2022: SAMPLE (B)(6)- THE INITIAL RESULT FROM THE MODULE WAS 0.035 NG/ML. THE REPEAT RESULT WAS LESS THAN THE LEVEL OF DETECTION OF 0.010 NG/ML. ON (B)(6) 2022: SAMPLE (B)(6)- THE INITIAL RESULT FROM THE MODULE WAS 0.036 NG/ML. THE REPEAT RESULT WAS LESS THAN THE LEVEL OF DETECTION OF 0.010 NG/ML. SAMPLE (B)(6)- THE INITIAL RESULT FROM THE MODULE WAS 0.89 NG/ML. THE REPEAT RESULT WAS LESS THAN THE LEVEL OF DETECTION OF 0.010 NG/ML. ON (B)(6) 2022: SAMPLE (B)(6)- THE INITIAL RESULT FROM THE MODULE WAS 0.024 NG/ML. THE REPEAT RESULT WAS LESS THAN THE LEVEL OF DETECTION OF 0.010 NG/ML. ON AN UNSPECIFIED DATE: SAMPLE (B)(6) - THE INITIAL RESULT FROM THE MODULE WAS 1.8 NG/ML. THE REPEAT RESULT WAS LESS THAN THE LEVEL OF DETECTION OF 0.010 NG/ML. THE REPEAT RESULTS WERE DEEMED CORRECT. THE REAGENT LOT NUMBER AND THE EXPIRATION DATE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15926317 ↗MalfunctionJan 26, 2022Incorrect Measurement; Incorrect Measurement; Incorrect, Inadequate or Imprecise Result or Readings
THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS TROPONIN T STAT ASSAY RESULTS NOT CORRESPONDING TO THE CLINICAL PICTURE FOR TWO SAMPLES FROM ONE PATIENT TESTED ON THE COBAS 8000 - COBAS E 602 MODULE WITH SERIAL NUMBER (B)(4). THE INITIAL RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. UPON RECEIVING THE RESULT, THE CLINICIAN ASKED THE PATIENT TO GO TO THE EMERGENCY ROOM. AT THE EMERGENCY ROOM, THE PATIENT WAS TESTED FOR TROPONIN I USING AN ABBOTT ARCHITECT ANALYZER. A NEW SAMPLE WAS OBTAINED FROM THE PATIENT FOR THE TROPONIN T AND TROPONIN I ASSAYS. THE REPORTER RECALIBRATED THE TROPONIN T ASSAY AND RERUN THE SAMPLE. THIS SAMPLE WAS ALSO SENT OUT TO A SISTER FACILITY THAT ALSO USES A ROCHE ANALYZER AND AN ABBOTT ARCHITECT ANALYZER FOR THE TROPONIN T AND TROPONIN I ASSAYS RESPECTIVELY. ON (B)(6) 2021 AT 9:00 PM, THE TROPONIN T INITIAL RESULT FROM THE E 602 MODULE WAS 4.1 NG/ML. THE TROPONIN I INITIAL RESULT FROM THE ABBOTT ARCHITECT WAS <0.01 (THE UNIT OF MEASUREMENT WAS NOT SPECIFIED). THE REPORTER STATED THAT THIS TROPONIN T RESULT DID NOT MATCH THE CLINICAL PICTURE OF THE PATIENT; THEY BELIEVED THAT THE TROPONIN I WAS THE ACCURATE RESULT. ON (B)(6) 2021 AT 9:00 AM, THE TROPONIN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13358383 ↗MalfunctionJul 14, 2021Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; High Test Results
THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS TROPONIN T STAT ASSAY RESULTS FOR THREE PATIENTS TESTED ON A COBAS 6000 E 601 MODULE SERIAL NUMBER (B)(4). THE INITIAL TROPONIN RESULTS WERE REPORTED OUTSIDE THE LABORATORY. THE CUSTOMER PERFORMED REPEAT TESTING WITH THE SAMPLES ON A DIFFERENT COBAS 6000 E 601 MODULE. AT 19:45, PATIENT 1'S INITIAL TROPONIN RESULT WAS 0.04 NG/ML. AT 20:33, PATIENT 1'S REPEAT TROPONIN RESULT WAS 0.02 NG/ML. AT 19:46, PATIENT 2'S INITIAL TROPONIN RESULT WAS 0.04 NG/ML. AT 20:33, PATIENT 2'S REPEAT TROPONIN RESULT WAS 0.01 NG/ML. AT 20:33, PATIENT 3'S INITIAL TROPONIN RESULT WAS 0.04 NG/ML. AT 20:33, PATIENT 3'S REPEAT TROPONIN RESULT WAS 0.01 NG/ML. THE CUSTOMER DETERMINED THE REPEAT TROPONIN RESULTS WERE CORRECT AND CORRECTED REPORTS WERE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12165318 ↗MalfunctionAug 17, 2018Non Reproducible Results; Non Reproducible Results
A ROCHE REPRESENTATIVE HAD BEEN AT THE CUSTOMER SITE TO PERFORM TECHNICAL TASKS ON MULTIPLE ANALYZERS INCLUDING THIS COBAS 6000 E 601 MODULE. THE ELECSYS TROPONIN T STAT ASSAY (TROPONIN T STAT) HAD BEEN REMOVED FROM THE MODULE DURING THIS TIME REQUIRING THE CUSTOMER TO RECALIBRATE AND RUN QC ONCE IT WAS PUT BACK ON. THE CUSTOMER THEN BEGAN TESTING PATIENT SAMPLES. AT APPROXIMATELY 5:15 P.M. THE LABORATORY WAS NOTIFIED THAT A TREND HAD BEEN NOTICED FOR TROPONIN T STAT RESULTS FOR PATIENT SAMPLES. ALL TROPONIN T STAT RESULTS WERE EITHER 0.04 NG/ML OR HIGHER WHEN THE EXPECTED RESULTS WERE 0.01 NG/ML OR BELOW. THE CUSTOMER STOPPED RUNNING PATIENT SAMPLES AND RAN QC WHICH WAS OUTSIDE OF THE ACCEPTABLE RANGE. THE CUSTOMER STATED 23 PATIENTS HAD BEEN RUN AND TROPONIN T STAT RESULTS HAD BEEN REPORTED OUTSIDE OF THE LABORATORY BETWEEN 11:30 A.M. AND 12:00 P.M. THE CUSTOMER STARTING REPEATING AND CORRECTING PATIENT RESULTS BETWEEN 5:30 P.M. AND 10:00 P.M. REFER TO ATTACHED DATA FOR PATIENT RESULTS. THE CUSTOMER MADE NEW CALIBRATOR MATERIAL AND RECALIBRATED THE ASSAY. AFTER RECALIBRATING, THE QC RESULTS WERE WITHIN RANGE. THE CUSTOMER WAS NOT AWARE OF ANY PATIENTS BEING TREATED BASED ON THE I
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7792859 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082699. One submission often covers many models, so check the product and lot details in each record.
- Z-1596-2021 · Apr 7, 2021 · Open, ClassifiedElecsys Troponin I Immunoassay: Catalog# 05094810160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.
- Z-1595-2021 · Apr 7, 2021 · Open, ClassifiedElecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.
- Z-1388-2012 · Mar 12, 2012 · TerminatedElecsys Troponin I STAT Immunoassay For the in-vitro quantitative determination of cardiac troponin I in human serum and plasma. Elecsys Troponin I is intended to aid in the diagn
- Z-1387-2012 · Mar 12, 2012 · TerminatedElecsys Troponin I Immunoassay For the in-vitro quantitative determination of cardiac troponin I in human serum and plasma. Elecsys Troponin I is intended to aid in the diagnosis
Other IMMUNOASSAY METHOD, TROPONIN SUBUNIT devices
- i-STAT 03P90-25Abbott Point Of Care Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT474 reports
- Atellica IM TNIH 10997841Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT463 reports
- Access hsTnI B52669Beckman Coulter, Inc. · Immunoassay Method, Troponin Subunit2 recalls170 reports
- i-STAT 06P23-25Abbott Point Of Care Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT195 reports
- VITROS 6802301Ortho Clinical Diagnostics · IMMUNOASSAY METHOD, TROPONIN SUBUNIT1 recall139 reports
- Atellica IM TNIH 10997840Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT166 reports
- Dimension EXL LOCI Troponin I High Sensitive 10471068Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT1 recall76 reports
- Access AccuTnI+3 A98143Beckman Coulter, Inc. · Immunoassay Method, Troponin Subunit1 death106 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026