Skip to content

Ethanol Gen2 5967104190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04015630927678), received between Apr 25, 2024 and Apr 25, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Non Reproducible Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 25, 2024Non Reproducible Results; Non Reproducible Results

    THE INITIAL REPORTER RECEIVED A QUESTIONABLE ETHANOL GEN.2 RESULT FOR ONE PATIENT TESTED ON A COBAS 8000 C702 MODULE. THE INITIAL RESULT WAS REPORTED TO THE EMERGENCY ROOM. THE DOCTOR QUESTIONED THE RESULT AS THE PATIENT SMELLED OF ALCOHOL AND THE PATIENT'S COMPANION CONFIRMED THE AMOUNT OF ALCOHOL THE PATIENT TOOK, PROMPTING THE REPORTER TO RERUN THE QC AND THE PATIENT SAMPLE. THE INITIAL RESULT WITH THE PATIENT SAMPLE IN A SAMPLE CUP WAS 6.5 MG/DL (NEGATIVE). THE FIRST REPEAT RESULT USING THE PRIMARY TUBE WAS >300 MG/DL (326.3 MG/DL). THE SECOND REPEAT WITH THE PATIENT SAMPLE IN A SAMPLE CUP WAS >300 MG/DL (336.3 MG/DL). THE RESULT OF A NEW PATIENT SAMPLE WAS >300 MG/DL (309.7MG/DL).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19182805 ↗

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page