Elecsys Ferritin 3737551160Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04015630938940), received between Oct 28, 2021 and Jun 2, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results3
  • Non Reproducible Results3
  • High Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 2, 2025Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results

    THE INITIAL REPORTER RECEIVED A QUESTIONABLE RESULT ELECSYS FERRITIN (FERRITIN) ASSAY RESULT FROM ONE PATIENT SAMPLE TESTED ON THE COBAS E 411 ANALYZER (RACK SYSTEM). THE INITIAL RESULT OF THE PATIENT'S FIRST SAMPLE WAS 0.846 NG/ML WITH A DATA FLAG. THE REPEAT RESULT WAS 332.0 NG/ML. THE RESULT OF THE PATIENT'S SECOND SAMPLE WAS 334.2 NG/ML. THE NURSE QUESTIONED THE INITIAL RESULT, PROMPTING THE PATIENT'S SAMPLE TO BE RERUN. THE REPEAT RESULT WAS DEEMED CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22125323 ↗
  • MalfunctionOct 3, 2023Low Test Results; Low Test Results

    THERE WAS AN ALLEGATION OF A QUESTIONABLE FERRITIN ELECSYS V2 RESULT FROM THE COBAS E411 RACK ANALYZER. THE INITIAL RESULT WAS 1.42 NG/ML AND WAS REPORTED OUTSIDE OF THE LABORATORY AS <2 NG/ML. THE DOCTOR QUESTIONED THE RESULT AND THE REPEAT RESULT WAS 74.18 NG/ML. THE REPEAT RESULT WAS BELIEVED CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17860670 ↗
  • MalfunctionOct 3, 2023Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results

    THE INITIAL REPORTER RECEIVED A QUESTIONABLE RESULT ELECSYS FERRITIN (FERRITIN) FROM ONE PATIENT SAMPLE TESTED ON THE COBAS E411 RACK. THE INITIAL RESULT WAS NOT REPORTED OUTSIDE THE LABORATORY. A RERUN WAS REQUESTED BY THE LABORATORY RESULTS VALIDATION DEPARTMENT DUE TO AN INITIAL RESULT NOT MATCHING THE PATIENT HISTORY. THE PATIENT'S OTHER SAMPLE WAS ALSO RUN. ON (B)(6) 2023: THE INITIAL RESULT OF THE INITIAL SAMPLE WAS 0.638 NG/ML WITH A DATA FLAG. THE RESULT OF THE OTHER SAMPLE WAS 331.8 NG/ML. ON (B)(6) 2023: THE REPEAT RESULT OF THE INITIAL SAMPLE WAS 336.4 NG/ML. THE INITIAL RESULT WAS DEEMED AN INCORRECT RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17859481 ↗
  • MalfunctionOct 28, 2021High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results

    THE INITIAL REPORTER RECEIVED A QUESTIONABLE ELECSYS FERRITIN RESULT FOR ONE PATIENT SAMPLE WITH THE COBAS E411 IMMUNOASSAY ANALYZER SERIAL NUMBER (B)(4). THE INITIAL RESULT WAS 25.8 NG/ML. THIS RESULT WAS REPORTED OUTSIDE THE LABORATORY AND QUESTIONED AS IT DID NOT MATCH MEDICAL HISTORY. ON (B)(4) 2021, A NEW SAMPLE WAS TESTED WITH A RESULT OF 18.16 NG/ML. ON (B)(4) 2021, THE SAMPLE WAS SENT TO AN EXTERNAL LABORATORY AND TESTED WITH AN ARCHITECT WITH A RESULT OF 7.1 NG/ML. ON (B)(4) 2021, THE SAMPLE WAS REPEATED ON ANOTHER E411 AT THE CUSTOMER SITE WITH A RESULT OF 16.78 NG/ML.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12714786 ↗

Other Radioimmunoassay (two-site solid phase), ferritin devices

All Radioimmunoassay (two-site solid phase), ferritin devices →