Elecsys PTH STAT Immunoassay 4892470160Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630939244), received between Oct 4, 2023 and Nov 22, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “High Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results1
- Incorrect, Inadequate or Imprecise Result or Readings1
- Low Test Results1
- Mechanical Problem1
- Non Reproducible Results1
Patient problems
- Hormonal Imbalance1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 22, 2023Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results
WE RECEIVED AN ALLEGATION OF QUESTIONABLE HIGH RESULTS FOR 1 PATIENT TESTED FOR ELECSYS PTH STAT IMMUNOASSAY (PTH STAT) DURING A SCHEDULED SURGERY TO REMOVE A PARATHYROID ADENOMA ON A COBAS 8000 E 602 MODULE. THE HIGH RESULTS ALLEGEDLY RESULTED IN A LONGER SURGERY AND ADDITIONAL TISSUE BEING REMOVED BY THE SURGEON. PTH HAD BEEN TESTED AT AN OUTSIDE LABORATORY BY AN UNKNOWN METHOD IN 2023 AND THE RESULT WAS REPORTEDLY 137 PG/ML. THE PATIENT ALLEGEDLY UNDERWENT REMOVAL OF THE ADENOMA WITH INITIAL DROP IN HER PARATHYROID LEVELS FROM 84 PG/ML TO 59 PG/ML. SINCE THE LEVEL REMAINED ELEVATED ADDITIONAL PARATHYROID TISSUE WAS SURGICALLY REMOVED ALONG WITH HALF OF HER THYROID AND HER THYMUS. SUBSEQUENTLY, TO WAS REPORTED THAT THE PATIENT UNDERWENT TESTING AT AN OUTSIDE LABORATORY ON A SIEMENS INSTRUMENT WITH A PARATHYROID HORMONE LEVEL OF LESS THAN 6.4 PG/ML. REPORTEDLY, THE PATIENT THEN UNDERWENT AN ADDITIONAL SURGERY TO REIMPLANT SOME OF HER PARATHYROID TISSUE INTO HER ABDOMEN TO PREVENT DEFICIENCY OF PARATHYROID HORMONE WHICH WAS SUCCESSFUL. ON AN UNKNOWN DATE, THE PATIENT REPORTEDLY WENT FOR A FOLLOW-UP LABORATORY TEST AT A DIFFERENT FACILITY AND THE PTH RESULT FROM COMPETITOR METHOD 1
Patient problems: Hormonal Imbalance
MAUDE report 18191411 ↗MalfunctionOct 4, 2023Mechanical Problem; Low Test Results; Non Reproducible Results; Mechanical Problem; Low Test Results; Non Reproducible Results
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT'S PLASMA SAMPLE TESTED WITH ELECSYS PARATHYROID HORMONE (PTH) STAT ASSAY ON A COBAS E411 IMMUNOASSAY ANALYZER WHEN COMPARED TO A DIFFERENT COBAS E411 IMMUNOASSAY ANALYZER. INITIAL RESULT: < 1.2 PG/ML. THE PHYSICIAN QUESTIONED THE RESULT. THE SAMPLE WAS REPEATED. REPEAT RESULT: 208.4 PG/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17871237 ↗
Other RADIOIMMUNOASSAY, PARATHYROID HORMONE devices
- VITROS 6802892Ortho Clinical Diagnostics · RADIOIMMUNOASSAY, PARATHYROID HORMONE1 recall155 reports
- Access Intact PTH A16972Beckman Coulter, Inc. · Radioimmunoassay, Parathyroid Hormone2 recalls8 reports
- IMMULITE® 2000 Systems IPT Intact PTH 10381441Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, PARATHYROID HORMONE1 recall18 reports
- Future Diagnostics 4K-IPT-00Future Diagnostics Solutions B.V. · RADIOIMMUNOASSAY, PARATHYROID HORMONE1 recall0 reports
- IMMULITE® 2000 Systems IPT Intact PTH 10381442Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, PARATHYROID HORMONE1 recall0 reports
- ADVIA Centaur® iPTH 10492368Siemens Healthcare Diagnostics Inc. · RADIOIMMUNOASSAY, PARATHYROID HORMONE23 reports
- Elecsys PTH 07251068190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, PARATHYROID HORMONE21 reports
- VITROS 6900052Ortho Clinical Diagnostics · RADIOIMMUNOASSAY, PARATHYROID HORMONE21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page