Skip to content

Elecsys PTH STAT Immunoassay 4892470160Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630939244), received between Oct 4, 2023 and Nov 22, 2023 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Low Test Results1
  • Mechanical Problem1
  • Non Reproducible Results1

Patient problems

  • Hormonal Imbalance1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 22, 2023Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results

    WE RECEIVED AN ALLEGATION OF QUESTIONABLE HIGH RESULTS FOR 1 PATIENT TESTED FOR ELECSYS PTH STAT IMMUNOASSAY (PTH STAT) DURING A SCHEDULED SURGERY TO REMOVE A PARATHYROID ADENOMA ON A COBAS 8000 E 602 MODULE. THE HIGH RESULTS ALLEGEDLY RESULTED IN A LONGER SURGERY AND ADDITIONAL TISSUE BEING REMOVED BY THE SURGEON. PTH HAD BEEN TESTED AT AN OUTSIDE LABORATORY BY AN UNKNOWN METHOD IN 2023 AND THE RESULT WAS REPORTEDLY 137 PG/ML. THE PATIENT ALLEGEDLY UNDERWENT REMOVAL OF THE ADENOMA WITH INITIAL DROP IN HER PARATHYROID LEVELS FROM 84 PG/ML TO 59 PG/ML. SINCE THE LEVEL REMAINED ELEVATED ADDITIONAL PARATHYROID TISSUE WAS SURGICALLY REMOVED ALONG WITH HALF OF HER THYROID AND HER THYMUS. SUBSEQUENTLY, TO WAS REPORTED THAT THE PATIENT UNDERWENT TESTING AT AN OUTSIDE LABORATORY ON A SIEMENS INSTRUMENT WITH A PARATHYROID HORMONE LEVEL OF LESS THAN 6.4 PG/ML. REPORTEDLY, THE PATIENT THEN UNDERWENT AN ADDITIONAL SURGERY TO REIMPLANT SOME OF HER PARATHYROID TISSUE INTO HER ABDOMEN TO PREVENT DEFICIENCY OF PARATHYROID HORMONE WHICH WAS SUCCESSFUL. ON AN UNKNOWN DATE, THE PATIENT REPORTEDLY WENT FOR A FOLLOW-UP LABORATORY TEST AT A DIFFERENT FACILITY AND THE PTH RESULT FROM COMPETITOR METHOD 1

    Patient problems: Hormonal Imbalance

    MAUDE report 18191411 ↗
  • MalfunctionOct 4, 2023Mechanical Problem; Low Test Results; Non Reproducible Results; Mechanical Problem; Low Test Results; Non Reproducible Results

    WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT'S PLASMA SAMPLE TESTED WITH ELECSYS PARATHYROID HORMONE (PTH) STAT ASSAY ON A COBAS E411 IMMUNOASSAY ANALYZER WHEN COMPARED TO A DIFFERENT COBAS E411 IMMUNOASSAY ANALYZER. INITIAL RESULT: < 1.2 PG/ML. THE PHYSICIAN QUESTIONED THE RESULT. THE SAMPLE WAS REPEATED. REPEAT RESULT: 208.4 PG/ML.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17871237 ↗