Elecsys AFP 07026706190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630939909), received between Jun 5, 2025 and Dec 16, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 16, 2025Low Test Results; Low Test Results

    THERE WAS AN ALLEGATION OF A QUESTIONABLE ELECSYS AFP RESULT FROM THE COBAS E 801 ANALYTICAL UNIT FOR ONE PATIENT. THE INITIAL RESULT WAS 1063 NG/ML. THE REPEAT RESULT AFTER 1:10 DILUTION WAS 12,100 NG/ML, ACCOMPANIED BY A DATA FLAG. A SECOND REPEAT RESULT AFTER 1:100 DILUTION WAS 121,000 NG/ML, ACCOMPANIED BY A DATA FLAG. A THIRD REPEAT RESULT AFTER 1:400 DILUTION WAS 484,000 NG/ML, ACCOMPANIED BY A DATA FLAG. A FOURTH REPEAT RESULT AFTER A MANUAL 1:10 DILUTION, AND THEN A 1:400 DILUTION ON THE INSTRUMENT, WAS 410,480 NG/ML. THE SAMPLE WAS REPEATED AFTER THE DOCTOR QUESTIONED IT. THE PATIENT HAD PREVIOUS RESULTS IN JULY THAT WERE EXTREMELY HIGH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23816117 ↗
  • MalfunctionJun 5, 2025High Test Results; High Test Results

    WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS NOT MATCHING THE PATIENT'S CLINICAL PICTURE FOR 1 PATIENT SAMPLE WHEN TESTED WITH ELECSYS ALPHA 1-FETOPROTEIN (AFP) ASSAY ON A COBAS E801 ANALYTICAL UNIT. INITIAL RESULT: 91.3 NG/ML. THE CUSTOMER QUESTIONED THE INITIAL RESULT AS IT DID NOT MATCH THE PATIENT'S CLINICAL DIAGNOSIS AND REPEATED THE SAMPLE. NO QUESTIONABLE RESULT WAS REPORTED OUTSIDE THE LABORATORY. AFP RESULT ON WAKO CHEMILUMINESCENCE: 2.7 NG/ML, AND THIS RESULT WAS CONSIDERED NORMAL (REFERENCE RANGE: 0-7 NG/ML). AFP-L3/AFP: <0.5 % (REFERENCE RANGE: 0-9.99). PIVKA RESULT ON FUJI CHEMILUMINESCENCE ANALYZER (FUJIREBIO LUMIPULSE G1200): 22 MAU/ML (REFERENCE RANGE: 0-40 MAU/ML). THE PATIENT'S MAGNETIC RESONANCE IMAGING (MRI) EXAMINATION SHOWED NO ABNORMALITY. THE CUSTOMER BELIEVED THAT THE WAKO RESULT WAS CORRECT. THE CUSTOMER SUSPECTED AN INTERFERENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22148851 ↗