Elecsys Cortisol II 6687733160Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630940677), received between Aug 23, 2023 and Apr 23, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results2
- Incorrect, Inadequate or Imprecise Result or Readings1
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2025High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results
THE INITIAL REPORTER RECEIVED A QUESTIONABLE ELECSYS CORTISOL II AND ELECSYS ACTH TEST SYSTEM (ACTH) RESULT FROM ONE PATIENT SAMPLE TESTED ON THE COBAS 6000 E601 MODULE. THIS MEDWATCH IS FOR THE ELECSYS CORTISOL II ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR THE ELECSYS ACTH TEST SYSTEM (ACTH) ASSAY REPORT. THE INITIAL RESULT OF THE INITIAL PATIENT SAMPLE WAS 24.27 UG/DL (24.3 UG/DL). THE REPEAT RESULT OF THE INITIAL PATIENT SAMPLE WAS 23.73 UG/DL. THE RESULT OF A NEW PATIENT SAMPLE WAS 12.51 UG/DL (12.5 UG/DL). THE PHYSICIAN QUESTIONED THE INITIAL RESULT AS IT DID NOT MATCH THE PATIENT'S PREVIOUS RESULT, PROMPTING THE PATIENT'S SAMPLE TO BE RERUN. THE RESULT OF THE NEW PATIENT SAMPLE WAS DEEMED CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21892558 ↗MalfunctionAug 23, 2023Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH ELECSYS CORTISOL II ON A COBAS E411 DISK ANALYZER. THE SAMPLE INITIALLY RESULTED IN A CORTISOL VALUE OF 7.93 UG/DL. THE DOCTOR QUESTIONED THE RESULT AND THE SAMPLE WAS REPEATED, RESULTING IN A VALUE OF 1.51 UG/DL. THE REPEAT VALUE WAS DEEMED CORRECT AND WAS EXPECTED FOR THE PATIENT. THE CUSTOMER RE-CENTRIFUGED THE SAMPLE AND LOCATED A SMALL CLOT IN IT. THE CLOT WAS REMOVED AND THE SAMPLE WAS REPEATED THREE ADDITIONAL TIMES, RESULTING IN CORTISOL VALUES OF 1.53 UG/DL, 2.25 UG/DL, AND 2.39 UG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17604100 ↗
Other FLUOROMETRIC, CORTISOL devices
- Atellica IM Cor 10995538Siemens Healthcare Diagnostics Inc. · FLUOROMETRIC, CORTISOL2 recalls0 reports
- ADVIA Centaur® COR 10994924Siemens Healthcare Diagnostics Inc. · FLUOROMETRIC, CORTISOL1 recall0 reports
- ADVIA Centaur® COR +E 10994926Siemens Healthcare Diagnostics Inc. · FLUOROMETRIC, CORTISOL1 recall0 reports
- Atellica IM Cor 11206248Siemens Healthcare Diagnostics Inc. · FLUOROMETRIC, CORTISOL1 recall0 reports
- Atellica IM Cor 10995537Siemens Healthcare Diagnostics Inc. · FLUOROMETRIC, CORTISOL1 recall0 reports
- Elecsys Cortisol II 07027150190Roche Diagnostics GmbH · FLUOROMETRIC, CORTISOL17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page