Tina-quant CReactive Protein IV 07876432190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04015630946853), received between Feb 28, 2023 and Feb 19, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Non Reproducible Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Non Reproducible Results2
- Incorrect Measurement1
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 19, 2026Non Reproducible Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH TINA-QUANT C-REACTIVE PROTEIN I ON A COBAS C 111 ANALYZER. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLE INITIALLY RESULTED IN A CRP VALUE OF 335.6 MG/L. THE REAGENT PACK WAS NEARLY DEPLETED AT THE TIME OF MEASUREMENT. A NEW REAGENT PACK WAS LOADED AND THE SAMPLE WAS REPEATED TWICE, RESULTING IN CRP VALUES OF 28.14 MG/L AND 28.29 MG/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24388222 ↗MalfunctionSep 18, 2023Incorrect Measurement; Non Reproducible Results; Incorrect Measurement; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE CRPL4 C-REACTIVE PROTEIN GEN.4 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS C 111 ANALYZER. ON (B)(6) 2023, THE INITIAL CRPL4 RESULT WAS 52 MG/DL. ON (B)(6) 2023, THE PATIENT PRESENTED BACK AT THE EMERGENCY ROOM WITH ABDOMINAL PAIN AND THIS PROMPTED THE CUSTOMER TO REPEAT THE ORIGINAL SAMPLE. THE SAMPLE WAS REPEATED THREE TIMES AND THE RESULTS WERE 0.67 MG/DL, 58 MG/DL, AND 133 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17764674 ↗MalfunctionFeb 28, 2023Low Test Results; Low Test Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT TESTED WITH CRPL4 C-REACTIVE PROTEIN GEN.4 ON A COBAS C 111 ((B)(4)). NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLE INITIALLY RESULTED IN A CRP VALUE OF 0.71 MG/L. THE SAMPLE WAS REPEATED TWICE, RESULTING IN VALUES OF 75.85 MG/L AND 78.31 MG/L. THE 75.85 MG/L VALUE WAS REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16456974 ↗
Other SYSTEM, TEST, C-REACTIVE PROTEIN devices
- Atellica CH Revised C-Reactive Protein (RCRP) 11537223Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, C-REACTIVE PROTEIN2 recalls0 reports
- Optilite C-Reactive Protein Reagent NK044.OPT.AThe Binding Site Group Limited · System, Test, C-Reactive Protein1 recall0 reports
- Tina-quant CReactive Protein IV 07876033190Roche Diagnostics GmbH · SYSTEM, TEST, C-REACTIVE PROTEIN7 reports
- Tina-quant CReactive Protein IV 07876424190Roche Diagnostics GmbH · SYSTEM, TEST, C-REACTIVE PROTEIN6 reports
- Dimension® Flex® reagent cartridge RCRP 10444903Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, C-REACTIVE PROTEIN6 reports
- Tina-quant C-Reactive Protein IV 08057591190Roche Diagnostics GmbH · SYSTEM, TEST, C-REACTIVE PROTEIN5 reports
- CReactive Protein Latex) 20764930322Roche Diagnostics GmbH · SYSTEM, TEST, C-REACTIVE PROTEIN2 reports
- CReactive Protein Gen3 4956842190Roche Diagnostics GmbH · SYSTEM, TEST, C-REACTIVE PROTEIN1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page