Skip to content

OptiView DAB Detection Kit 06396500001Recalls & reported problems

Ventana Medical Systems, Inc. · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

3 FDA recall records name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Aug 2, 2018Terminated

    Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.

    Z-2996-2018 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Dec 19, 2017Terminated

    Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.

    Z-0729-2018 · Root cause: Process control · Recall details · FDA record ↗

  3. Dec 19, 2017Terminated

    Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.

    Z-0730-2018 · Root cause: Process control · Recall details · FDA record ↗

Other IMMUNOHISTOCHEMISTRY REAGENTS AND KITS devices

All IMMUNOHISTOCHEMISTRY REAGENTS AND KITS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page