LinkSymphoKnee 880-307/21Recalls & reported problems
WALDEMAR LINK GmbH & Co. KG · Femoral Augment, Distal · Class 2 · Implant
All models of LinkSymphoKnee (117)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04026575257393), received between Aug 2, 2023 and Aug 2, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Component Misassembled” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
9 other recalls cover devices cleared under the same FDA submission (K202924); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Apr 3, 2025Class IIOpen, Classified
The device was delivered with a preassembled Femoral Augment screw that was missing its thread.
Z-1699-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Dec 23, 2022Class IIOpen, Classified
Fixation screw was inserted through the wrong side of the augment (distal to proximal, as opposed to proximal to distal, which is the correct orientation).
Z-2067-2023 · Root cause: Component change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Component Misassembled2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 2, 2023Component Misassembled; Component Misassembled
SCREWS THAT ATTACH AUGMENT TO FEM COMPONENT WERE PUT INTO AUGMENTS BACKWARDS. SCREWS SHOULD BE FED FROM PROX TO DISTAL. THERE WERE FED THROUGH THE AUGMENT FROM DISTAL TO PROXIMAL THUS CAUSING THE SURGEON TO HAVE TO REMOVE THEM AND REINSERT THEM. THE PATIENT IN NO WAY WAS AFFECTED AND THE SURGERY PROCEEDED AS PLANNED. DR.(B)(6) DOES NOT WANT OR NEED ANY FOLLOW UP. HAS SPECIFICALLY ASKED TO JUST INFORM LINK SO THEY CAN CORRECT THE PROBLEM MOVING FORWARD. [CUSTOMER]
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17441422 ↗MalfunctionAug 2, 2023Component Misassembled; Component Misassembled
DR. (B)(6) WAS IN A SURGERY ON (B)(6) 2023 AND HAD AUGMENTS THAT WERE PACKAGED WITH THE SCREWS ASSEMBLED FROM THE INCORRECT SIDE. SCREWS WERE REMOVED AND PUT INTO OTHER SIDE AND AUGMENTS WERE IMPLANTED. NO CONSEQUENCE TO THE PATIENT. [CUSTOMER].
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17441407 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K202924. One submission often covers many models, so check the product and lot details in each record.
- Z-2541-2025 · Jul 31, 2025 · Open, ClassifiedLinkSymphoKnee System, Tibial Component, Modular, Symmetric, Sz. 5, CoCrMo/TiNbN, cemented, Fixed. REF Number: 880-100/50
- Z-1702-2025 · Apr 3, 2025 · Open, ClassifiedTibial Augment; Item Numbers: (1) 880-331/11 (Sz. 1-2 - cemented x-sm), (2) 880-331/12 (Sz. 1-2 - cemented sm), (3) 880-331/13 (Sz. 1-2 - cemented md), (4) 880-331/21 (Sz. 1-2 - ce
- Z-1700-2025 · Apr 3, 2025 · Open, ClassifiedPosterior Femoral Augment; Item Numbers: (1) 880-310/11 (Sz. 0 - cemented sm), (2) 880-310/21 (Sz. 0 - cemented lg), (3) 880-311/11 (Sz. 1-2 - cemented sm), (4) 880-311/21 (Sz. 1-2
- Z-1701-2025 · Apr 3, 2025 · Open, ClassifiedL-Shaped Femoral Augment; Item Numbers: (1) 880-320/12 (Sz. 0 - cemented sm), (2) 880-320/22 (Sz. 0 - cemented lg), (3) 880-321/12 (Sz. 1-2 - cemented sm), (4) 880-321/22 (Sz. 1-2
- Z-1373-2025 · Feb 14, 2025 · Open, ClassifiedModular Stem cylindrical, Tilastan, press-fit cementless, L= 80 mm, DIA = 12 mm, Product Code: 880-601/12
- Z-1372-2025 · Feb 14, 2025 · Open, ClassifiedModular Stem cylindrical, Tilastan, press-fit cementless, L= 80 mm, DIA = 11 mm, Product Code: 880-601/11
- Z-1374-2025 · Feb 14, 2025 · Open, ClassifiedModular Stem cylindrical, Tilastan, press-fit cementless, L= 80 mm, DIA = 13 mm, Product Code: 880-601/13
- Z-1674-2023 · Apr 25, 2023 · Open, ClassifiedLINK SymphoKnee System L-Shaped Femoral Augment, Medial-Right/Lateral-Left, Sz 5-6 Tilastan, cemented. Intended for primary and revision total knee replacement. Item Number: 880-32
- Z-2066-2023 · Dec 23, 2022 · Open, ClassifiedLinkSymphoKnee Distal Femoral Augment - Tilastan - Size 7-8, Medial-Right/Lateral Left, 5mm- Intended for primary and revision total knee replacement in skeletally mature patients.
Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- Scorpio 73-3710Howmedica Osteonics Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls1 report
- Scorpio 73-3110Howmedica Osteonics Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls0 reports
- Arthrex® AR-503-TTTEArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall14 reports
- Arthrex® AR-503-TTTJArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall9 reports
- ENGAGE 1-10011-300Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- Arthrex® AR-503-TTTGArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- ENGAGE 1-10003-400Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall6 reports
- ENGAGE 1-10012-550Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page