LinkSymphoKnee - Instruments 881-509/00Recalls & reported problems
WALDEMAR LINK GmbH & Co. KG · Patella prosthesis trial - LinkSymphoKnee System · Class 1
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04026575310203), received between Jun 28, 2024 and Jun 28, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Inaccurate Information” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jun 27, 2024Open, Classified
The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process.
Z-2445-2024 · Root cause: Process change control · FDA record ↗
Most reported problems
Device problems
- Inaccurate Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 28, 2024Inaccurate Information; Inaccurate Information
FOR PROF. (B)(6), THE PATELLA SELECTED WITH THE SIZING TEMPLATE APPEARED TOO LARGE. A REMEASUREMENT OF THE SIZING TEMPLATE HAS OBVIOUSLY RESULTED IN A REVERSED LABELING ON THE SIZING TEMPLATE. [CUSTOMER].
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19631399 ↗
Other TEMPLATE devices
- ATTUNE 2544-01-014Depuy (Ireland) · TEMPLATE802 reports
- ATTUNE 2545-00-546Depuy (Ireland) · TEMPLATE1 recall301 reports
- ARTICUL/EZE 2531-51-000Depuy (Ireland) · TEMPLATE323 reports
- DELTA XTEND 2307-95-000Depuy (Ireland) · TEMPLATE315 reports
- ATTUNE 2545-00-547Depuy (Ireland) · TEMPLATE1 recall231 reports
- ATTUNE 2544-00-520Depuy (Ireland) · TEMPLATE279 reports
- ATTUNE 2545-00-545Depuy (Ireland) · TEMPLATE1 recall221 reports
- ARTICUL/EZE 2530-91-000Depuy (Ireland) · TEMPLATE271 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.