Skip to content

Endo-Model Modular Knee Prosthesis System 15-2950/04Recalls & reported problems

WALDEMAR LINK GmbH & Co. KG · Modular Stem · Class 2 · Implant

All models of Endo-Model Modular Knee Prosthesis System (64)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04026575319015), received between Jan 9, 2025 and Jan 9, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Failure of Implant1
  • Metal Related Pathology1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 9, 2025Fracture; Fracture

    SIDE: LEFT. DIAGNOSIS: FEMORAL SHAFT FRACTURE WITH POSSIBLE METALLOSIS? REVISION OF THE CEMENTED 120MM FEMORAL SHAFT, THE SIZE M FEMORAL COMPONENT, THE SIZE 3 FEMORAL SEGMENT AND THE TIBIAL PLATEAU. THE TIBIA WAS FIRMLY ANCHORED. THE FEMORAL SHAFT IS BROKEN OR FRAGMENTED AT THE DISTAL END OR AT THE ATTACHMENT TO THE FEMORAL CONE IN SEVERAL PLACES. REMOVAL OF THE SHAFT WAS DIFFICULT BECAUSE NO LOOSENING WAS DETECTABLE. THE HOLE FOR SCREW FIXATION WAS WIDENED WITH A DRILL TO ALLOW A REMOVAL HOOK TO BE ATTACHED AND THE STEM TO BE REMOVED WITH A HAMMER. AFTER REMOVAL OF THE STEM, IRRIGATION AND INSERTION OF LOW-VISCOSITY CEMENT, THE FOLLOWING WAS IMPLANTED: FEMORAL COMPONENT 15-2817/12 REPLACEMENT PLATEAU 150027/15 MODULAR STEM 15-2950/06 FEMORAL SEGMENT 15-2978/02 A VERY SATISFACTORY RESULT WAS ACHIEVED.[CUSTOMER].

    Patient problems: Failure of Implant; Metal Related Pathology

    MAUDE report 21111586 ↗

Other Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices

All Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page