Skip to content

Endo-Model Modular Knee Prosthesis System 15-8521/06Recalls & reported problems

WALDEMAR LINK GmbH & Co. KG · Rotational Knee Joint Prosthesis,Femoral component, condylus replacement · Class 2 · Implant

All models of Endo-Model Modular Knee Prosthesis System (64)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04026575383191), received between Jan 16, 2023 and Jan 17, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Naturally Worn1
  • Positioning Problem1

Patient problems

  • Failure of Implant1
  • Unspecified Infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 17, 2024Naturally Worn; Positioning Problem; Naturally Worn; Positioning Problem

    (B)(6). PATIENT PRESENTED WITH DIFFICULTY WALKING, COMBINED WITH CLICKING AND LOCKING IN THE RIGHT KNEE SUBSEQUENT TO A ON (B)(6) 2022 TKA (REVISION). X-RAYS SHOWS SOME OFFSET OF FEMUR WITH RESPECT TO THE TIBIAL PLATEAU. BOTH THE FEMORAL AND TIBIAL COMPONENTS APPEAR SECURELY FIXED ON X-RAY. REVISION SURGERY IS SCHEDULED FOR ON (B)(6) 2024. ADDITIONAL INFORMATION WILL BE UPDATED AFTER SURGERY. UPDATE 03/13/2024: (B)(6) CONFIRMED THAT THE REVISION PLANNED FOR ON (B)(6) 2024 TOOK PLACE. DR. (B)(6) REPLACED THE FEMORAL COMPONENT AND TIBIAL POLY. THE EXPLANTED DEVICES WERE NOT RELEASED BY THE HOSPITAL. A REQUEST HAS BEEN MADE FOR THE OPERATIVE REPORT. (B)(6) NOTED THAT THIS PATIENT HAS NO EXTENSOR MECHANISM. INTRAOPERATIVELY, HE OBSERVED THAT THE POLYETHYLENE THIMBLE IN THE POST HAD WORN THROUGH ON THE ANTERIOR SIDE SEEMINGLY FROM THE SHEAR FORCE OF THE FEMUR ON THE POST RELATED TO LACK OF EXTENSOR MECHANISM. NO PHOTOS WERE AVAILABLE. ON (B)(6) 2024, USAGE FORM PROVIDED.[CUSTOMER].

    Patient problems: Failure of Implant

    MAUDE report 18525512 ↗
  • InjuryJan 16, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    DUE TO EXCESSIVE SWELLING, DR. (B)(6) UNDERTOOK A REVISION SURGERY ON THE PATIENT ON (B)(6) 2023 TO DEBRIDE/WASH OUT THE JOINT AND REPLACE THE PLATEAU, AND DISCOVERED CONDYLE CAP FROM THE FEMORAL COMPONENT CONDYLAR REPLACEMENT LOOSE IN THE JOINT SPACE, SITTING IN FRONT OF THE KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 16172828 ↗

Other Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices

All Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page