Ilesto 7 VR-T DF-1 US 383578Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BIOTRONIK Ilesto 7 VR-T is an implantable defibrillator, a device placed in the body to keep the heart from beating in an abnormal rhythm. Most of the 28 reports describe units signaling that the battery had reached end of life sooner than expected, prompting doctors to remove and replace them; a few also mention the device not reading out properly. Two injuries were reported and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
28
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 04035479125349), received between Apr 3, 2019 and Dec 19, 2025 — classified as 2 injuries, 26 malfunctions.

The most frequently reported specific device problem is “Premature Elective Replacement Indicator” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Elective Replacement Indicator12
  • Premature End-of-Life Indicator11
  • Battery Problem4
  • Failure to Interrogate2
  • Adverse Event Without Identified Device or Use Problem1
  • Unable to Obtain Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions19
  • No Known Impact Or Consequence To Patient7
  • Pocket Erosion1
  • Shock from Patient Lead(s)1

Recent adverse event reports

The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 19, 2025Premature End-of-Life Indicator; Premature End-of-Life Indicator

    THIS DEVICE WAS EXPLANTED DUE TO EOS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23849642 ↗
  • MalfunctionNov 17, 2025Premature End-of-Life Indicator; Premature End-of-Life Indicator

    DEVICE EXPLANTED DUE TO EOS INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23568980 ↗
  • MalfunctionJun 11, 2025Premature End-of-Life Indicator; Premature End-of-Life Indicator

    THIS DEVICE WAS EXPLANTED DUE TO EOS WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22185133 ↗
  • MalfunctionMar 10, 2025Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21568983 ↗
  • MalfunctionDec 5, 2024Premature End-of-Life Indicator; Premature End-of-Life Indicator

    DEVICE EXPLANTED DUE TO EOS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20854641 ↗
  • MalfunctionNov 11, 2024Premature End-of-Life Indicator; Premature End-of-Life Indicator

    THIS DEVICE WAS EXPLANTED DUE TO EOS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20660233 ↗
  • MalfunctionOct 7, 2024Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE WAS EXPLANTED DUE TO ERI WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20392270 ↗
  • MalfunctionAug 22, 2024Premature End-of-Life Indicator; Premature End-of-Life Indicator

    THIS DEVICE WAS EXPLANTED DUE TO EOS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20031607 ↗
  • MalfunctionJun 18, 2024Premature End-of-Life Indicator; Premature End-of-Life Indicator

    THIS DEVICE WAS INTERROGATED ON (B)(6) 2024 AND WAS FOUND TO BE IN EOS. DEVICE CURRENTLY REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19558584 ↗
  • InjuryJul 26, 2023Premature End-of-Life Indicator; Premature End-of-Life Indicator

    PATIENT RECEIVED MULTIPLE INAPPROPRIATE SHOCKS DUE TO NOISE ON THE ASSOCIATED RV LEAD. PATIENT WENT TO THE ER WHERE DEVICE WAS INTERROGATED AND FOUND TO BE EOS. DEVICE CURRENTLY REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 17400856 ↗
  • MalfunctionMar 1, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE SHOWS ERI WITH UNEXPECTED BATTERY BEHAVIOR. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16464900 ↗
  • MalfunctionFeb 21, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16414421 ↗
  • MalfunctionDec 13, 2022Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15971910 ↗
  • MalfunctionJun 6, 2022Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE AT ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14604975 ↗
  • MalfunctionMay 13, 2022Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE IS AT ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. DEVICE CURRENTLY REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14401742 ↗
  • MalfunctionOct 11, 2021Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED. ERI REPORTED ON HM WITH BATTERY VOLTAGE 3.02. NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12608423 ↗
  • MalfunctionAug 18, 2021Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE SHOWED 53 PERCENT BATTERY LONGEVITY ON 25-SEP-2020, AND AN ERI ALERT ON (B)(6) 2020. DEVICE WAS EXPLANTED AND RETURNED FOR ANALYSIS. THE DEVICE ANALYSIS INDICATED A BATTERY ISSUE. NO ADVERSE PATIENT EVENTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12339895 ↗
  • MalfunctionJul 29, 2021Failure to Interrogate; Premature End-of-Life Indicator; Failure to Interrogate; Premature End-of-Life Indicator

    DEVICE REMAINS IMPLANTED. DEVICE APPEARS TO BE AT EOS AND IS NOT RESPONDING TO INTERROGATION. DEVICE WILL BE CHANGED OUT AS SOON AS POSSIBLE. NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12244916 ↗
  • MalfunctionJul 19, 2021Failure to Interrogate; Failure to Interrogate

    DEVICE IS NOT ABLE TO BE INTERROGATED. THE PATIENT SHOWED UP TO THE ER AFTER RECEIVING MULTIPLE SHOCKS AND AFTER THAT THE DEVICE WAS NOT ABLE TO BE INTERROGATED. IT WAS RECOMMENDED TO CHANGE OUT THE DEVICE. DEVICE EXPLANTED ON (B)(6) 2021.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12193524 ↗
  • MalfunctionJun 7, 2021Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE AT ERI WITH UNEXPECTED BATTERY BEHAVIOR. THE DEVICE REMAINS IMPLANTED AT THIS TIME. NO ADVERSE PATIENT EVENTS HAVE BEEN REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11946445 ↗
  • MalfunctionAug 27, 2020Premature End-of-Life Indicator; Premature End-of-Life Indicator

    DEVICE EXPLANTED DUE TO EOS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10460754 ↗
  • MalfunctionMay 13, 2020Battery Problem; Battery Problem

    (B)(6), DEVICE SHOWED 77 PERCENT BATTERY ONE MONTH AGO. TODAY ON HOME MONITORING AND AN IN OFFICE CHECK THE DEVICE IS AT ERI. THE DEVICE WAS EXPLANTED AND REPLACED ON (B)(6) AND RETURNED FOR ANALYSIS. 5/13/2020, BASED ON THE ANALYSIS, THIS IS REPORTABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10052106 ↗
  • MalfunctionFeb 18, 2020Unable to Obtain Readings; Unable to Obtain Readings

    THIS DEVICE WAS UNABLE TO BE INTERROGATED (B)(6) 2020 OR (B)(6) 2020. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9723624 ↗
  • InjuryFeb 17, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO EROSION. NEW UPGRADED SYSTEM WAS IMPLANTED SAME DAY. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED. HOSPITAL RETAINED SYSTEM AND SHOULD RETURN IT FOR ANALYSIS.

    Patient problems: Pocket Erosion

    MAUDE report 9717470 ↗
  • MalfunctionSep 18, 2019Premature Elective Replacement Indicator; Battery Problem; Premature Elective Replacement Indicator; Battery Problem

    THIS DEVICE REPORTED ERI STATUS ON (B)(6) 2019. IN (B)(6) OF 2018, IN OFFICE INTERROGATION SHOWED BATTERY LONGEVITY OF 85%. THE DEVICE WAS EXPLANTED ON (B)(6) 2019 DUE TO ERI. BASED ON THE ANALYSIS RESULTS RECEIVED FROM THE MANUFACTURER ON (B)(4) 2019, THIS EVENT IS REPORTABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9082633 ↗

Other Implantable cardioverter defibrillator (non-CRT) devices

All Implantable cardioverter defibrillator (non-CRT) devices →