Itrevia 7 DR-T DF-1 392412Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Itrevia 7 DR-T is an implantable dual-chamber defibrillator, a pacemaker-like device that shocks the heart when needed. The 36 reports mostly describe units signaling they had reached end of service sooner than expected, with unusual battery behavior leading to devices being taken out. Reports mention 1 death and 9 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
36
Deaths reported
1
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 04035479128807), received between Jul 31, 2018 and Feb 12, 2026 — classified as 1 death, 9 injuries, 26 malfunctions.

The most frequently reported specific device problem is “Premature Elective Replacement Indicator” (22 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Elective Replacement Indicator22
  • Premature End-of-Life Indicator6
  • Adverse Event Without Identified Device or Use Problem5
  • Ambient Noise Problem1
  • Battery Problem1
  • Failure to Deliver Shock/Stimulation1
  • Inappropriate/Inadequate Shock/Stimulation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions27
  • Unspecified Infection4
  • Death1
  • No Known Impact Or Consequence To Patient1
  • Pocket Erosion1
  • Shock from Patient Lead(s)1
  • Syncope/Fainting1

Recent adverse event reports

The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 12, 2026Premature End-of-Life Indicator

    THIS DEVICE WAS EXPLANTED DUE TO EOS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24322422 ↗
  • InjuryFeb 10, 2026Premature End-of-Life Indicator

    THIS DEVICE WAS EXPLANTED DUE TO EOS. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24303974 ↗
  • MalfunctionJun 26, 2025Premature End-of-Life Indicator; Premature End-of-Life Indicator

    THIS DEVICE WAS EXPLANTED DUE TO EOS. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22342119 ↗
  • MalfunctionSep 11, 2024Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20193300 ↗
  • MalfunctionAug 27, 2024Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20075536 ↗
  • MalfunctionJul 10, 2024Premature End-of-Life Indicator; Premature End-of-Life Indicator

    THIS DEVICE WAS EXPLANTED DUE TO EOS WITH UNUSUAL BATTERY BEHAVIOR. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19716056 ↗
  • MalfunctionMar 6, 2024Premature End-of-Life Indicator; Premature End-of-Life Indicator

    DEVICE EXPLANTED DUE TO ERI THEN EOS INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18851777 ↗
  • MalfunctionFeb 27, 2024Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE AT ERI WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18794686 ↗
  • MalfunctionFeb 1, 2024Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18627447 ↗
  • MalfunctionDec 1, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE WAS EXPLANTED DUE TO ERI INDICATION WITH UNUSUAL BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18247133 ↗
  • MalfunctionSep 12, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17730692 ↗
  • MalfunctionMar 21, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE SHOWS ERI WITH UNEXPECTED BATTERY BEHAVIOR. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16588377 ↗
  • MalfunctionMar 2, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE SHOWS ERI WITH UNEXPECTED BATTERY BEHAVIOR. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16466509 ↗
  • MalfunctionFeb 21, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16414418 ↗
  • MalfunctionAug 23, 2022Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE WAS EXPLANTED DUE TO ERI WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15275858 ↗
  • MalfunctionMay 16, 2022Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14412439 ↗
  • MalfunctionFeb 15, 2022Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI. NO ADVERSE EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13539039 ↗
  • MalfunctionJan 31, 2022Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE WAS EXPLANTED AND REPLACED DUE TO ERI WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13394414 ↗
  • MalfunctionJan 14, 2022Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE WAS EXPLANTED DUE TO ERI WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13258474 ↗
  • MalfunctionDec 14, 2021Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE WAS EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS HAVE BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13002147 ↗
  • InjuryDec 4, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS DEVICE WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 12930709 ↗
  • MalfunctionNov 12, 2021Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO GOING FROM 25 PERCENT TO ERI OVERNIGHT. NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE ADDED TO THIS FILE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12804821 ↗
  • MalfunctionSep 15, 2021Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE IS AT EARLY ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS HAVE BEEN REPORTED. THE DEVICE WILL BE CHANGED OUT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12474096 ↗
  • MalfunctionAug 31, 2021Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE IS AT ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS HAVE BEEN REPORTED. THE DEVICE CURRENTLY REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12395684 ↗
  • MalfunctionJun 23, 2021Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    IT WAS REPORTED THAT THIS DEVICE SHOWED ERI STATUS UPON INTERROGATION, AND HAD 50 PERCENT BATTERY REMAINING ON HOME MONITORING ON (B)(6) 2021. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12054262 ↗