Ilivia 7 VR-T DF4 ProMRI 404626Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Ilivia 7 VR-T DF4 ProMRI is an implantable defibrillator placed under the skin to keep the heart from beating into abnormal rhythms. Most of the 75 reports describe the unit signaling that its battery was near end of life earlier than expected, sometimes with pacing stopping during the replacement procedure; others mention infections at the implant site. Reports include 1 death and 18 injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
77
Deaths reported
1
Injuries reported
20
Recalls naming it
0

Summary

The FDA's MAUDE database contains 77 adverse event reports that list this exact device (DI 04035479142131), received between Jul 24, 2018 and Sep 24, 2026 — classified as 1 death, 20 injuries, 56 malfunctions.

The most frequently reported specific device problem is “Premature Elective Replacement Indicator” (49 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Elective Replacement Indicator49
  • Adverse Event Without Identified Device or Use Problem16
  • Battery Problem3
  • Failure to Convert Rhythm2
  • Failure to Interrogate2
  • Premature End-of-Life Indicator2
  • Defibrillation/Stimulation Problem1
  • High impedance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions57
  • Unspecified Infection12
  • No Known Impact Or Consequence To Patient2
  • Arrhythmia1
  • Implant Pain1
  • Insufficient Information1
  • Sepsis1
  • Shock from Patient Lead(s)1

Recent adverse event reports

The 25 most recent of 77 reports, in the reporter's own words as published by the FDA.

  • InjurySep 24, 2026Premature Elective Replacement Indicator

    Device explanted due to ERI indication with unexpected battery behavior. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50886058 ↗
  • InjurySep 15, 2026Premature End-of-Life Indicator

    This device was explanted due to EOS indication with unusual battery behavior. No adverse patient events were reported. Should additional information be received this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50773365 ↗
  • InjuryMar 9, 2026Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24550740 ↗
  • MalfunctionDec 22, 2025Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23872000 ↗
  • MalfunctionDec 22, 2025Premature End-of-Life Indicator; Premature End-of-Life Indicator

    THIS DEVICE WAS EXPLANTED DUE TO EOS WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23864425 ↗
  • MalfunctionAug 11, 2025Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22777914 ↗
  • MalfunctionJun 10, 2025No Pacing; No Pacing

    DEVICE AT ERI INDICATION AND WHEN IT WAS BEING CHANGED OUT ALL PACING STOPPED. WHEN THE DEVICE WAS PLACED BACK IN THE POCKET IT STARTED PACING AGAIN. THE DEVICE WAS EXPLANTED. NO ADVERSE PATIENT EVENTS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22177073 ↗
  • MalfunctionJan 23, 2025Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE SHOWS ERI WITH UNEXPECTED BATTERY BEHAVIOR. DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21218815 ↗
  • MalfunctionJan 20, 2025Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21194357 ↗
  • MalfunctionJan 13, 2025Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21137235 ↗
  • MalfunctionDec 4, 2024Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20849329 ↗
  • MalfunctionNov 18, 2024Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20711276 ↗
  • MalfunctionNov 11, 2024Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20660062 ↗
  • MalfunctionOct 29, 2024Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20557211 ↗
  • MalfunctionJun 19, 2024Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19569734 ↗
  • MalfunctionFeb 1, 2024Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18627303 ↗
  • MalfunctionDec 28, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE SHOWS ERI WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. DEVICE REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18415047 ↗
  • MalfunctionNov 21, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18183579 ↗
  • MalfunctionNov 6, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18073768 ↗
  • MalfunctionOct 16, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE AT ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS REPORTED. DEVICE REMAINS IMPLANTED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17940087 ↗
  • MalfunctionAug 31, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE SHOWS ERI WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED. DEVICE REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17660909 ↗
  • MalfunctionAug 7, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17471100 ↗
  • MalfunctionJun 21, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    THIS DEVICE WAS EXPLANTED DUE TO ERI WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17171811 ↗
  • InjuryMar 28, 2023Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    THE PATIENT CALLED TO REPORT THEY BELIEVE THEY EXPERIENCED THREE SHOCKS ON (B)(6) 2023 AROUND 2:10PM. THE DEVICE WAS INTERROGATED ON (B)(6) 2023 AND THE PATIENT REPORTED THE TECHNICIAN DID NOT SEE ANY SHOCKS WERE RECORDED. THE PATIENT IS CONCERNED ABOUT DEVICE FUNCTION AND WHY IT DID NOT RECORD WHAT THEY EXPERIENCED. THE PATIENT REPORTED THEIR DOCTOR DID NOT BELIEVE THE SHOCKS WERE REAL. APS ENCOURAGED THE PATIENT TO REACH OUT TO THEIR DOCTOR AGAIN IF THEY STILL HAVE CONCERNS. DEVICE REMAINS IMPLANTED.

    Patient problems: Implant Pain

    MAUDE report 16626847 ↗
  • MalfunctionMar 9, 2023Premature Elective Replacement Indicator; Premature Elective Replacement Indicator

    DEVICE EXPLANTED DUE TO ERI INDICATION WITH UNEXPECTED BATTERY BEHAVIOR. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16518199 ↗

Other Implantable cardioverter defibrillator (non-CRT) devices

All Implantable cardioverter defibrillator (non-CRT) devices →