This BIOTRONIK BioMonitor IV is an implantable cardiac event recorder. The reports most often describe devices migrating or being expelled from the implant pocket, with wound dehiscence, pocket erosion, or discomfort; some mention sensing problems. There are 80 reports naming the identifier, with no deaths and 73 injuries, plus 7 malfunctions. No FDA recall names this device. Reports rose from 24 to 46 in the last 12 months compared with the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
82
Deaths reported
0
Injuries reported
75
Recalls naming it
0

Summary

The FDA's MAUDE database contains 82 adverse event reports that list this exact device (DI 04035479177850), received between Jun 26, 2024 and Sep 25, 2026 — classified as 75 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (25 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 2 reportsSep 2024: 3 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 2 reportsJul 2025: 7 reportsAug 2025: 5 reportsSep 2025: 3 reportsOct 2025: 1 reportsNov 2025: 5 reportsDec 2025: 2 reportsJan 2026: 2 reportsFeb 2026: 4 reportsMar 2026: 4 reportsApr 2026: 6 reportsMay 2026: 6 reportsJun 2026: 6 reportsJul 2026: 3 reportsAug 2026: 6 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem35
  • Migration or Expulsion of Device25
  • Device Sensing Problem12
  • Failure to Interrogate5
  • Device Dislodged or Dislocated4
  • Ambient Noise Problem3
  • Expulsion3
  • Failure to Sense3

Patient problems

  • Wound Dehiscence27
  • No Clinical Signs, Symptoms or Conditions25
  • Pocket Erosion10
  • Discomfort8
  • Unspecified Infection8
  • Implant Pain6
  • Anxiety1
  • Erythema1

Recent adverse event reports

The 25 most recent of 82 reports, in the reporter's own words as published by the FDA.

  • InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem

    Device was explanted per patient complaint of causing discomfort due to original placement. Should additional information be received this file will be updated.

    Patient problems: Discomfort

    MAUDE report 50899872 ↗
  • InjurySep 16, 2026Migration or Expulsion of Device; Device Dislodged or Dislocated

    This device dislodged and fell out through sutures. Date of migration unknown. Should additional information become available, this file will be updated.

    Patient problems: Wound Dehiscence

    MAUDE report 50790299 ↗
  • InjurySep 15, 2026Migration or Expulsion of Device; Device Sensing Problem

    This device was explanted due to migration resulting in loss of signal. Should additional information be received this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50779022 ↗
  • InjurySep 8, 2026Failure to Interrogate

    This device was explanted due to device could not be interrogated post implant. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50671520 ↗
  • InjuryAug 26, 2026Adverse Event Without Identified Device or Use Problem

    Device was explanted due to patient complaint of discomfort. Should additional information be received, this file will be updated.

    Patient problems: Discomfort

    MAUDE report 50457096 ↗
  • InjuryAug 20, 2026Failure to Sense; Ambient Noise Problem

    Device was explanted due to HMSC reported loss of sensing on recordings after (b)(6) 2026. sECG recordings appear noisy. Patient confirmed to have cardioversion on (b)(6) 2026. Device reset was unsuccessful. Signal remained in poor quality. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50351548 ↗
  • InjuryAug 19, 2026Migration or Expulsion of Device

    Loop recorder externalized and was partially exposed at clinic visit. Physician explanted device. Should additional information become available, this file will be updated.

    Patient problems: Wound Dehiscence

    MAUDE report 50329874 ↗
  • InjuryAug 18, 2026Migration or Expulsion of Device; Device Sensing Problem

    This device was explanted because it migrated and was not sensing. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50300015 ↗
  • InjuryAug 6, 2026Device Sensing Problem

    This device was explanted due to loss of sensing post cardioversion. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50164180 ↗
  • InjuryAug 3, 2026Migration or Expulsion of Device

    This device was reported to have fallen out of the pocket. Should additional information be received this file will be updated.

    Patient problems: Wound Dehiscence

    MAUDE report 50114330 ↗
  • InjuryJul 29, 2026Migration or Expulsion of Device

    This loop was explanted after hanging out of patients skin. Should additional information become available, this file will be updated.

    Patient problems: Wound Dehiscence

    MAUDE report 50069185 ↗
  • InjuryJul 28, 2026Migration or Expulsion of Device; Device Dislodged or Dislocated

    Device dislodged from incision while at home and migrated out of the pocket. Should additional information become available, this file will be updated.

    Patient problems: Wound Dehiscence

    MAUDE report 50056030 ↗
  • InjuryJul 28, 2026Adverse Event Without Identified Device or Use Problem

    This device was explanted due to patient has sensitive skin and wanted loop removed. Should additional information become available, this file will be updated.

    Patient problems: Discomfort

    MAUDE report 50055980 ↗
  • InjuryJun 29, 2026Device Sensing Problem

    THIS DEVICE WAS EXPLANTED DUE TO LOSS OF SIGNAL FOLLOWING A CARDIOVERSION. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25654052 ↗
  • InjuryJun 18, 2026Adverse Event Without Identified Device or Use Problem

    DEVICE WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 25548743 ↗
  • InjuryJun 10, 2026Migration or Expulsion of Device

    THIS DEVICE WAS EXPLANTED DUE TO IT BROKE THROUGH THE INCISION SITE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: Wound Dehiscence

    MAUDE report 25462783 ↗
  • MalfunctionJun 2, 2026Migration or Expulsion of Device; Device Sensing Problem

    PATIENT CAME TO THE OFFICE ON (B)(6) 2026 FOR A FOLLOW UP. THE DEVICE COULD NOT BE INTERROGATED. THE PATIENT WAS SENT FOR CHEST XRAY AND THE DEVICE COULD NOT BE SEEN ON XRAY. THE PATIENT WAS THEN SENT FOR CT SCAN. THE RESULTS OF THE CT SCAN SHOWED THAT THE DEVICE COULD NOT BE LOCATED. EVIDENCE ON HOME MONITORING SHOWS THE PATIENT IS TRANSMITTING LAST MESSAGE (B)(6) 2026 WITH THEIR OWN CARDIO MESSENGER. REVIEWING DATA SHOWS COMPLETE LOSS OF SENSING CURRENTLY. THERE ARE CONCERNS OF THE POSSIBILITY OF MIGRATION DUE TO NOT BE ABLE TO LOCATE THE DEVICE ON IMAGING. THE PATIENT CURRENTLY IS ASYMPTOMATIC.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25364109 ↗
  • InjuryJun 1, 2026Adverse Event Without Identified Device or Use Problem

    DEVICE WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 25354639 ↗
  • InjuryJun 1, 2026Migration or Expulsion of Device

    IT WAS REPORTED THAT THIS DEVICE FELL OUT OF THE PATIENTS BODY. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: Wound Dehiscence

    MAUDE report 25354630 ↗
  • InjuryMay 21, 2026Failure to Interrogate

    THIS DEVICE WAS IMPLANTED BUT THEN WAS NOT BEING IDENTIFIED AND COULD NOT BE INTERROGATED OR PROGRAMED BY SEVERAL DIFFERENT PROGRAMMERS. THE DEVICE WAS THEN EXPLANTED AND A NEW DEVICE IMPLANTED WITH NO ISSUE. NO ADVERSE PATIENT EVENTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25244341 ↗
  • InjuryMay 18, 2026Device Sensing Problem

    THIS DEVICE WAS EXPLANTED DUE TO LOSS OF SENSING POST CARDIOVERSION AND SUSPECTED DAMAGE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25210991 ↗
  • InjuryMay 12, 2026Adverse Event Without Identified Device or Use Problem

    DEVICE WAS EXPLANTED DUE TO PATIENT WANTED TO REMOVE BECAUSE HE CLAIMED THAT IT INTERFERED WITH HIS THERAPY HE WAS GETTING AND IT WAS PAINFUL.

    Patient problems: Implant Pain

    MAUDE report 25148553 ↗
  • InjuryMay 4, 2026Migration or Expulsion of Device

    PATIENT REPORTED A FALL WHICH RIPPED THEIR SUTURES CAUSING THE LOOP TO FALL OUT. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: Fall; Wound Dehiscence

    MAUDE report 25074852 ↗
  • InjuryMay 1, 2026Migration or Expulsion of Device

    IT WAS REPORTED THAT THIS DEVICE FELL OUT. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: Wound Dehiscence

    MAUDE report 25059599 ↗
  • InjuryMay 1, 2026Adverse Event Without Identified Device or Use Problem

    INCISION DID NOT HEAL AFTER IMPLANT AND DEVICE POKED OUT LEADING TO INFECTION AND WAS EXPLANTED.

    Patient problems: Unspecified Infection; Wound Dehiscence

    MAUDE report 25059595 ↗

Other Recorder, Event, Implantable Cardiac, (With Arrhythmia Detection) devices

All Recorder, Event, Implantable Cardiac, (With Arrhythmia Detection) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page