Skip to content

Biolox® forte NJ108Recalls & reported problems

Aesculap AG · BIOLOX PROSTHESIS HEAD 8/10 32MM L · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04038653304356), received between Dec 1, 2020 and Jun 9, 2021 — classified as 2 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • Failure of Implant2
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 9, 2021Fracture; Fracture

    CORRECTION: NJ108- BIOLOX PROSTHESIS HEAD 8/10 32MM L- AAG REFERENCE (B)(4) IS NO LONGER REPORTABLE AS IT CHANGED FROM LEADING MATERIAL TO INVOLVED COMPONENT. ASSOCIATED MEDWATCH-REPORTS: 9610612-2022-00087 ((B)(4) - NH102) AND 9610612-2021-00426 ((B)(4) - NJ108). INVOLVED COMPONENTS: NH048T - PLASMACUP SC SIZE 48MM - 51407851. NJ011T - BICONTACT N PLASMAPORE 8/10 SIZE 11MM - 51344702A. NJ108 - BIOLOX PROSTHESIS HEAD 8/10 32MM L - 51422587.

    Patient problems: Failure of Implant; Pain

    MAUDE report 11970824 ↗
  • InjuryDec 1, 2020Fracture; Fracture

    IT WAS REPORTED TO AESCULAP AG THAT A BIOLOX PROSTHESIS HEAD (PART # NJ108) WAS USED DURING A TOTAL HIP ARTHROPLASTY PROCEDURE OF THE RIGHT LEG. THE FIRST IMPLANTATION SURGERY WAS PERFORMED ON (B)(6) 2008 AT (B)(6) HOSPITAL. ACCORDING TO THE COMPLAINANT, ON (B)(6) 2020, THE CERAMIC IMPLANT WAS CONFIRMED AS BEING FRACTURED, WHICH RESULTED IN A REVISION SURGERY BEING SCHEDULED, AND THEN PERFORMED ON (B)(6) 2020. BOTH THE CERAMIC INSERT AND THE FEMORAL HEAD WERE FOUND TO BE BROKEN. THE PATIENT HAD UNDERGONE A COMPUTERIZED TOMOGRAPHY (CT) SCAN TO CONFIRM THE BREAK PRIOR TO THE REVISION SURGERY. ADDITIONALLY, THE CERAMIC IMPLANTS WERE NOTED AS HAVING BEEN BROKEN APPROXIMATELY TWO (2) WEEKS BEFORE THE REVISION SURGERY. THE DEVICE IS AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION. ALTHOUGH THE PATIENT EXPERIENCED HIP PAIN AS A RESULT OF THE BREAK, THE PATIENT WAS ABLE TO WALK "AS USUAL." NO FURTHER PATIENT INFORMATION HAS BEEN MADE AVAILABLE. ADDITIONAL INFORMATION HAS BEEN REQUESTED, BUT HAS NOT BEEN MADE AVAILABLE. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4). INVOLVED COMPONENTS: NH102-SC/MSC CERAMICS INSERT 32MM 48/50 SYM - LOT # 51450162.

    Patient problems: Failure of Implant

    MAUDE report 10924885 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page