CRANIOFIX® ABSORBABLE FF016Recalls & reported problems
Aesculap AG · 6 CRANIOFIX ABSORBABL.CLAMP STERILE 11MM · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04038653348398), received between Feb 11, 2021 and Jan 3, 2022 — classified as 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Device Appears to Trigger Rejection2
- Break1
- Detachment of Device or Device Component1
- Fracture1
- Mechanics Altered1
- Patient Device Interaction Problem1
- Product Quality Problem1
- Unstable1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Failure of Implant1
- Swelling/ Edema1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 3, 2022Product Quality Problem; Product Quality Problem
IT WAS REPORTED TO AESCULAP AG THAT A 6 CRANIOFIX ABSORBABL.CLAMP STERILE 11MM (PART # FF016) WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE SPRING INSIDE OF THE HANDLE WAS NOT RELEASED DURING THE FINAL TIGHTENING OPERATION. REPORTEDLY, THE PRODUCT WAS DAMAGED. THE COMPLAINT DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. THE PATIENT HARM IS UNKNOWN. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4). ASSOCIATED MEDWATCH REPORTS: 9610612-2021-00792 - 400538941, 9610612-2021-00793 - 400538942.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13142136 ↗MalfunctionDec 20, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
CORRECTION/ADDITIONAL INFORMATION: PATIENT HARM CHANGED FROM ADDITIONAL MEDICAL INTERVENTION TO TEMPORARY IMPAIRMENT.
Patient problems: Swelling/ Edema
MAUDE report 13043692 ↗InjuryDec 20, 2021Unstable; Unstable
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH FF016-6 CRANIOFIX ABSORBABL. CLAMP STERILE 11MM. ACCORDING TO THE COMPLAINT DESCRIPTION, THE PATIENT UNDERWENT SURGERY IN (B)(6) 2020 AND WAS RE-FIXED WITH CFA IN (B)(6) OF THE SAME YEAR. A REVISION SURGERY WAS NECESSARY. ADDITIONAL INFORMATION WAS NOT PROVIDED NOR AVAILABLE / WAS NOT AVAILABLE. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4) PATIENT INFO: (B)(6) GIRL.
Patient problems: Failure of Implant
MAUDE report 13043615 ↗InjuryDec 8, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
IT WAS REPORTED TO AESCULAP AG THAT A 6 CRANIOFIX ABSORBABL.CLAMP STERILE 11MM (PART # FF016) WAS USED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THREE (3) MONTHS AFTER THE SURGICAL PROCEDURE, THE SCALP JUST ABOVE THE CFA CONTRACTED AND BECAME THIN, AND EVENTUALLY THE SCALP WAS TORN. THE DOCTOR RE-SUTURED IN THE OUTPATIENT DEPARTMENT, HOWEVER, AFTER APPROXIMATELY A HALF YEAR, THE PATIENT WAS RE-SUTURED WITH GENERAL ANESTHESIA SURGERY. THE COMPLAINT DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. A REVISION SURGERY WAS NECESSARY. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12960559 ↗MalfunctionDec 3, 2021Mechanics Altered; Mechanics Altered
IT WAS REPORTED TO AESCULAP (B)(4) THAT A 6 CRANIOFIX ABSORBABL. CLAMP STERILE 11MM (PART # FF016) WAS USED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE PRODUCT WAS DAMAGED. IN PREPARATION FOR CRANIAL CLOSURE, THE PRODUCT LINE WAS BROKEN DURING PRE-TENSION. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE MALFUNCTION IS FILED UNDER (B)(4) REFERENCE CC (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12924149 ↗MalfunctionMay 7, 2021Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED TO AESCULAP AG THAT A CRANIOFIX ABSORBABL. CLAMP STERILE 11MM (PART # FF016) WAS USED DURING AN UNKNOWN PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, SOME TISSUE PIECES COLLECTED AFTER REOPERATION OF PATIENTS USING CRANIOFIX ABSORBABLE. CRANIOFIX ABSORBABLE HAS BEEN ABSORBED, BUT INVESTIGATION OF THE TISSUE FRAGMENTS FORMED IN THE REMAINING SUTURE PART SHOULD BE INVESTIGATED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE MALFUNCTION IS FILED UNDER (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11793664 ↗MalfunctionApr 20, 2021Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH FF016 - 6 CRANIOFIX ABSORBABL. CLAMP STERILE 11MM. ACCORDING TO THE COMPLAINT DESCRIPTION, THE WIRE BROKE DURING SURGERY. THERE WAS NO DESCRIBED PATIENT HARM. ADDITIONAL INFORMATION WAS NOT PROVIDED. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11693287 ↗InjuryFeb 11, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH FF016 - 6 CRANIOFIX ABSORBABL. CLAMP STERILE 11MM. ACCORDING TO THE COMPLAINT DESCRIPTION, THE DEVICE WAS DAMAGED IN THE LOWER DISC AND THE BONE FLAP WAS LIFTED, SO SURGERY WAS PERFORMED AGAIN. NO INFECTION. A REVISION SURGERY WAS NECESSARY. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT YET RECEIVED AS OF THIS REPORT. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11314669 ↗
FDA clearance and classification
The FDA regulates product code GXR (cover, burr hole) under 21 CFR 882.5250 as a Class II device. 21 CFR 882.5250 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the CRANIOFIX® ABSORBABLE FF016
Has the CRANIOFIX® ABSORBABLE FF016 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04038653348398), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the CRANIOFIX® ABSORBABLE FF016?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Feb 11, 2021 and Jan 3, 2022. In those reports, the most common specific device problems are “Device Appears to Trigger Rejection” (2), “Break” (1) and “Detachment of Device or Device Component” (1); the most common patient problems are “Failure of Implant” (1) and “Swelling/ Edema” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CRANIOFIX® ABSORBABLE FF016?
None of the reports are classified as deaths, 3 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the CRANIOFIX® ABSORBABLE FF016 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the CRANIOFIX® ABSORBABLE FF016, and is it still sold?
The FDA's device database lists Aesculap AG as the company responsible for it (brand CRANIOFIX® ABSORBABLE, model FF016). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other COVER, BURR HOLE devices
- GUARDIAN™ 6010St. Jude Medical, Inc. · COVER, BURR HOLE147 reports
- GUARDIAN™ 6010St. Jude Medical, Inc. · COVER, BURR HOLE1 death25 reports
- "1.5MM" SYSTEM 95-1505Biomet Microfixation, Inc · COVER, BURR HOLE23 reports
- "1.5MM" SYSTEM 91-6105Biomet Microfixation, Inc · COVER, BURR HOLE20 reports
- "1.5MM" SYSTEM 91-6106Biomet Microfixation, Inc · COVER, BURR HOLE7 reports
- 1.5MM SYSTEM 91-6105X-50Biomet Microfixation, Inc · COVER, BURR HOLE6 reports
- TMS SYSTEM 25-2007Biomet Microfixation, Inc · COVER, BURR HOLE3 reports
- NA 446.611Synthes GmbH · COVER, BURR HOLE2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page