Skip to content

AESCULAP US751Recalls & reported problems

Aesculap, Inc. · ROUND FILTERS W/INDICATOR · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Aesculap round sterilization filter with indicator is used in sterilization trays and containers. The 51 reports mostly describe filters found with pinprick holes, tears, or weak spots, plus brown staining, black marks on the indicator circles, and other printing or color problems, sometimes delaying a case. There are no deaths and one injury, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 04038653473731), received between Feb 5, 2020 and Jun 11, 2026 — classified as 1 injury, 49 malfunctions.

The most frequently reported specific device problem is “Material Puncture/Hole” (41 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 9 reportsDec 2023: 9 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 10 reportsJul 2025: 0 reportsAug 2025: 4 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Puncture/Hole41
  • Product Quality Problem5
  • Material Split, Cut or Torn3
  • Illegible Information2
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Material Discolored1
  • Material Integrity Problem1
  • Residue After Decontamination1

Patient problems

  • No Clinical Signs, Symptoms or Conditions49
  • Failure of Implant1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 11, 2026Material Discolored

    CUSTOMER HAS CONCERN ABOUT BROWN STAINING ON FILTERS. THIS INCIDENT OCCURRED ON (B)(6) 2026 AND CAUSED A DELAY TO A CASE. APPROXIMATELY (3) FILTERS SHOWED THE BROWN STAINS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25470343 ↗
  • MalfunctionMay 28, 2026Material Puncture/Hole

    DISTRIBUTOR STATED CUSTOMER REPORTED THAT FILTERS HAVE BEEN NOTED TO HAVE PIN PRICK HOLES IN THE FILTERS. THE LOT IS NOT KNOWN AS IT IS UNCLEAR WHEN THE PROCESSING OF THE DEVICES OCCURRED - LOT THEREFORE CANNOT BE DETERMINED BASED ON STOCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25317386 ↗
  • MalfunctionMay 6, 2026Material Puncture/Hole

    IT WAS REPORTED THAT THERE WERE PIN HOLES OBSERVED IN THE FILTERS WHILE OPENING THE STERILIZATION CONTAINER FOR A PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25100894 ↗
  • MalfunctionFeb 13, 2026Material Puncture/Hole

    AS REPORTED BY THE FACILITY: "THERE WERE BLACK MARKS ON THE INDICATOR CIRCLES ON THE FILTERS. IT APPEARS THEY WERE PARTIALLY EXPOSED TO HEAT OF SOME TYPE OF MACHINE ERROR.." NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24331787 ↗
  • MalfunctionFeb 12, 2026Material Puncture/Hole

    AS REPORTED BY THE FACILITY: "WE HAD MULTIPLE BAGS OF FILTERS WITH WEAK SPOTS OR PIN HOLES IN THEM. WE DID A ROOT CAUSE ANALYSIS AND COULD LOCATE A CAUSE FROM THE TRAYS INVESTIGATED." NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24323918 ↗
  • MalfunctionAug 18, 2025Product Quality Problem; Residue After Decontamination; Product Quality Problem; Residue After Decontamination

    AS REPORTED BY THE USER FACILITY: AS ADVISED, WE WILL INSPECT AND CONTINUE TO REJECT THE STERILIZATION FILTERS, NOT ONLY FOR THE MICRO HOLES WE ARE FINDING BUT NOW FOR THE NEW IMPERFECTION, THE BLACK INK LINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22826370 ↗
  • MalfunctionAug 7, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT THERE WERE PIN HOLES OBSERVED IN THE FILTERS. WHEN THE FILTERS ARE HELD UP TO THE LIGHT AND EXAMINED WITH THE NAKED EYE, THE BLUE COLOR SEEMS VARIED AND THERE ARE WHITE SPOTS THAT ARE EITHER VARIATION IN THE MATERIAL (IN THE NAP OR FILTER WEAVE PATTERN, A THINNER KNAP) OR HOLES. REPORTEDLY THE HOLE IN THE FILTER LINES UP EXACTLY WITH THE HOLE IN THE LID.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22736755 ↗
  • MalfunctionAug 5, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT THERE WERE PIN HOLES OBSERVED IN THE FILTERS. WHEN THE FILTERS ARE HELD UP TO THE LIGHT AND EXAMINED WITH THE NAKED EYE, THE BLUE COLOR SEEMS VARIED AND THERE ARE WHITE SPOTS THAT ARE EITHER VARIATION IN THE MATERIAL (IN THE NAP OR FILTER WEAVE PATTERN, A THINNER KNAP) OR HOLES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22714255 ↗
  • MalfunctionAug 5, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT THERE WERE PIN HOLES OBSERVED IN THE FILTERS. WHEN THE FILTERS ARE HELD UP TO THE LIGHT AND EXAMINED WITH THE NAKED EYE, THE BLUE COLOR SEEMS VARIED AND THERE ARE WHITE SPOTS THAT ARE EITHER VARIATION IN THE MATERIAL (IN THE NAP OR FILTER WEAVE PATTERN, A THINNER KNAP) OR HOLES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22714254 ↗
  • MalfunctionJun 27, 2025Material Puncture/Hole; Material Puncture/Hole

    ACCORDING TO THE CUSTOMER: " WE ARE HAVING ISSUES WITH HOLES IN AESCULAP FILTERS. THIS IS AN EVERYDAY OCCURRENCE. WE, IN SPD, HAVE RECOMMENDED EVERYTHING POSSIBLE TO AVOID CREATING HOLES IN FILTERS. I DO UNDERSTAND THERE ARE AREAS IN THE FILTERS THAT ARE LIGHT AND CAN LOOK LIKE HOLE IN FILTER BUT WE ARE RECEIVING REPORTS OF TINY HOLES. I DID INSPECT A NEW BAG. I PUT IT UP AGAINST A LIGHTED MAGNIFIER. THERE ARE AREAS THAT LOOK LIKE TINY PIN HOLES."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22344789 ↗
  • MalfunctionJun 26, 2025Material Puncture/Hole; Material Puncture/Hole

    ACCORDING TO THE CUSTOMER: " WE ARE HAVING ISSUES WITH HOLES IN AESCULAP FILTERS. THIS IS AN EVERYDAY OCCURRENCE. WE, IN SPD, HAVE RECOMMENDED EVERYTHING POSSIBLE TO AVOID CREATING HOLES IN FILTERS. I DO UNDERSTAND THERE ARE AREAS IN THE FILTERS THAT ARE LIGHT AND CAN LOOK LIKE HOLE IN FILTER BUT WE ARE RECEIVING REPORTS OF TINY HOLES. I DID INSPECT A NEW BAG. I PUT IT UP AGAINST A LIGHTED MAGNIFIER. THERE ARE AREAS THAT LOOK LIKE TINY PIN HOLES."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22337334 ↗
  • MalfunctionJun 24, 2025Material Puncture/Hole; Material Puncture/Hole

    ACCORDING TO THE EVENT DESCRIPTION: CUSTOMER FOUND SMALL PIN HOLES IN SEVERAL OF THE FILTERS IN ONE OF THEIR BAGS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22322580 ↗
  • MalfunctionJun 18, 2025Material Puncture/Hole; Material Puncture/Hole

    ACCORDING TO THE CUSTOMER: "WE ARE HAVING ISSUES WITH HOLES IN AESCULAP FILTERS. THIS IS AN EVERYDAY OCCURRENCE. WE, IN SPD, HAVE RECOMMENDED EVERYTHING POSSIBLE TO AVOID CREATING HOLES IN FILTERS. I DO UNDERSTAND THERE ARE AREAS IN THE FILTERS THAT ARE LIGHT AND CAN LOOK LIKE HOLE IN FILTER BUT WE ARE RECEIVING REPORTS OF TINY HOLES. I DID INSPECT A NEW BAG. I PUT IT UP AGAINST A LIGHTED MAGNIFIER. THERE ARE AREAS THAT LOOK LIKE TINY PIN HOLES."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22286610 ↗
  • InjuryJun 16, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH A COMPONENT OF AN AS VEGA SYSTEM. ACCORDING TO THE COMPLAINT DESCRIPTION, THIS KNEE IMPLANT WAS "DEFECTIVE AND FAILED PREMATURELY BECAUSE THE CERAMIC COATED SURFACE FAILS TO PROPERLY ADHERE TO THE CEMENT". THERE WAS POSTOPERATIVE MECHANICAL LOOSENING. IT WAS NOTED THAT THE SURGEON EASILY REMOVED THE IMPLANT BY HAND DURING THE REVISION. ADDITIONAL INFORMATION WAS NOT PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 22210709 ↗
  • MalfunctionJun 13, 2025Material Puncture/Hole; Material Puncture/Hole

    ACCORDING TO THE COMPLAINANT THE FILTERS HAVE MICRO PIN HOLES. REPORTEDLY THE PIN HOLES WERE FOUND UPON INSPECTION OF NEW FILTERS FROM THE PACKAGE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22205755 ↗
  • MalfunctionJun 2, 2025Material Puncture/Hole; Product Quality Problem; Material Puncture/Hole; Product Quality Problem

    AS REPORTED BY THE USER FACILITY: EVENT 3. CUSTOMER IS INSPECTING HOLES IN FILTERS WHEN FILTER IS REFLECTING SUN AND LIGHT IN CERTAIN WAYS. CUSTOMER COULD NOT PROVIDE A LOT NUMBER BUT GAVE 4 FILTER SAMPLES. ONE UNUSED AND 3 USED. THE FILTERS AROUND THE WORD AESCULAP SEEM LIGHTER UNDER LIGHT IN CERTAIN AREAS AND CUSTOMER CIRCLED THE HOLES. ONE SAMPLE IS FROM A 2023 FILTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22122458 ↗
  • MalfunctionJun 2, 2025Material Puncture/Hole; Product Quality Problem; Material Puncture/Hole; Product Quality Problem

    AS REPORTED BY THE USER FACILITY: EVENT 4: CUSTOMER IS INSPECTING HOLES IN FILTERS WHEN FILTER IS REFLECTING SUN AND LIGHT IN CERTAIN WAYS. CUSTOMER COULD NOT PROVIDE A LOT NUMBER BUT GAVE 4 FILTER SAMPLES. ONE UNUSED AND 3 USED. THE FILTERS AROUND THE WORD AESCULAP SEEM LIGHTER UNDER LIGHT IN CERTAIN AREAS AND CUSTOMER CIRCLED THE HOLES. ONE SAMPLE IS FROM A 2023 FILTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22122312 ↗
  • MalfunctionJun 2, 2025Material Puncture/Hole; Product Quality Problem; Material Puncture/Hole; Product Quality Problem

    AS REPORTED BY THE USER FACILITY: EVENT 1: CUSTOMER IS INSPECTING HOLES IN FILTERS WHEN FILTER IS REFLECTING SUN AND LIGHT IN CERTAIN WAYS. CUSTOMER COULD NOT PROVIDE A LOT NUMBER BUT GAVE 4 FILTER SAMPLES. ONE UNUSED AND 3 USED. THE FILTERS AROUND THE WORD AESCULAP SEEM LIGHTER UNDER LIGHT IN CERTAIN AREAS AND CUSTOMER CIRCLED THE HOLES. ONE SAMPLE IS FROM A 2023 FILTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22122234 ↗
  • MalfunctionJun 2, 2025Material Puncture/Hole; Product Quality Problem; Material Puncture/Hole; Product Quality Problem

    AS REPORTED BY THE USER FACILITY: EVENT 2: CUSTOMER IS INSPECTING HOLES IN FILTERS WHEN FILTER IS REFLECTING SUN AND LIGHT IN CERTAIN WAYS. CUSTOMER COULD NOT PROVIDE A LOT NUMBER BUT GAVE 4 FILTER SAMPLES. ONE UNUSED AND 3 USED. THE FILTERS AROUND THE WORD AESCULAP SEEM LIGHTER UNDER LIGHT IN CERTAIN AREAS AND CUSTOMER CIRCLED THE HOLES. ONE SAMPLE IS FROM A 2023 FILTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22122233 ↗
  • MalfunctionMay 14, 2025Material Puncture/Hole; Material Puncture/Hole

    SUPPLIER REPORTS HOLES ARE SHOWING UP IN FILTERS. AFTER CUSTOMER EXAMINED THESE FILTERS, THEY HAD WEAK SPOTS IN THEM, AND WHEN THEY WERE PROCESSED THEY CREATED PIN HOLE SIZED HOLES. CUSTOMER FOUND UNUSED FILTERS WITH PIN HOLES AS WELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22019959 ↗
  • MalfunctionApr 21, 2025Material Puncture/Hole; Material Puncture/Hole

    ACCORDING TO THE EVENT DESCRIPTION: THERE ARE HOLES ARE SHOWING UP IN FILTERS AFTER STERILE PROCESSING. AFTER EXAMINING THE FILTERS, THEY HAD WEAK SPOTS IN THEM, AND WHEN THEY WERE PROCESSED, THEY CREATED PIN HOLE SIZED HOLES. REPORTEDLY THERE WERE UNUSED FILTERS WITH PIN HOLES AS WELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21877429 ↗
  • MalfunctionFeb 6, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED TO AESCULAP INC. THAT A ROUND FILTERS W/INDICATOR (PART# US751) WAS USED FOR DURING STERILIZATION ON (B)(6) 2025. ACCORDING TO THE COMPLAINANT THERE IS A STRAIGHT LINE GOING THROUGH THE TOP OF THE FILTER THAT WAS SEEN POST STERILIZATION. REPORTEDLY THERE WAS A DELAY OF 5 MINUTES TO THE PROCEDURE AS ANOTHER SET HAD TO BE PULLED. THE ADVERSE EVENT IS FILED UNDER AIC REFERENCE (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21324239 ↗
  • MalfunctionSep 27, 2024Material Puncture/Hole; Material Puncture/Hole

    ACCORDING TO THE COMPLAINANT THERE WERE HOLES IN THE CONTAINER FILTERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20326128 ↗
  • MalfunctionFeb 27, 2024Illegible Information; Illegible Information

    THE USER FACILITY REPORTED THAT THE INDICATOR DOT ON THEIR ROUND FILTERS WITH INDICATOR ARE NOT "UNIFORMLY DARK BROWN" AFTER PROCESSING. PROCEDURE CANCELLATIONS WERE REPORTED AS THE DEVICES WERE NOT UTILIZED AND WERE REPROCESSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18791195 ↗
  • MalfunctionFeb 27, 2024Illegible Information; Illegible Information

    THE USER FACILITY REPORTED THAT THE FILTER INDICATOR DOT DID NOT FULLY CHANGE IN COLOR. A PROCEDURE DELAY WAS REPORTED AS THE DEVICES WERE NOT UTILIZED AND WERE REPROCESSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18791167 ↗

Other Sterilization wrap containers, trays, cassettes & other accessories devices

All Sterilization wrap containers, trays, cassettes & other accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page