PaediGAV® FV288TRecalls & reported problems
Christoph Miethke GmbH & Co. KG · PAEDI-SHUNTASSTISTANT 10 · Class 2 · Implant
All models of PaediGAV (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04041906113001), received between May 7, 2021 and May 18, 2026 — classified as 5 injuries.
The most frequently reported specific device problem is “Infusion or Flow Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Infusion or Flow Problem4
- Mechanical Problem2
Patient problems
- Hydrocephalus5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMay 18, 2026Infusion or Flow Problem
IT WAS REPORTED THAT A PAEDIESA VALVE (#FV288T) WAS IMPLANTED ON (B)(6) 2024. ACCORDING TO THE COMPLAINANT, THE VALVE CAUSED AN OVER-DRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2026. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. SAME EVENT AS #: 3004721439-2026-00141.
Patient problems: Hydrocephalus
MAUDE report 25200135 ↗InjuryJan 27, 2023Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A SHUNTASSISTANT (PART # FV288T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE VALVE IN COMBINATION WITH A PROGAV 2.0 WAS BELIEVED TO BE OPERATING IN OVERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 17 YEARS, HEIGHT: 153 CENTIMETERS (CM), WEIGHT: 57 KILOGRAMS (KG), GENDER: FEMALE. SAME EVENT AS MEDWATCH#3004721439-2022-00435 AND 3004721439-2022-00436.
Patient problems: Hydrocephalus
MAUDE report 16258210 ↗InjuryJan 7, 2022Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PAEDI SHUNTASISSTANT (PART # FV288T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO HAVE A SHUNT DYSFUNCTION. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6), HEIGHT: 148 CENTIMETERS (CM), WEIGHT: (B)(6) GENDER: UNKNOWN. ITS SAME PATIENT AS MDR ID #3004721439-2021-00345 AND MDR ID # 3004721439-2021-00347.
Patient problems: Hydrocephalus
MAUDE report 13193800 ↗InjuryDec 2, 2021Mechanical Problem; Mechanical Problem
ACCORDING TO THE COMPLAINANT A SHUNT ASSISTANT WAS NOT FUNCTIONING AS EXPECTED. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS NOT WORKING PROPERLY WHEN PUT TOGETHER AND INLINE, BUT EACH COMPONENT (RESERVOIR, VALVE, & SHUNT ASSIST) SEEMED TO WORK INDIVIDUALLY. THE SURGEON TOOK OUT THE VALVE SYSTEM AND REPLACED IT WITH ALL NEW COMPONENTS OF THE SAME PRODUCT. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. FURTHER DETAILS WERE NOT PROVIDED. WEIGHT: (B)(6). ASSOCIATED MEDWATCH: 3004721439-2021-00765.
Patient problems: Hydrocephalus
MAUDE report 12919455 ↗InjuryMay 7, 2021Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PAEDISA (#FV288T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN 2011 . ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE OPERATING IN UNDERDRIANAGE. THE VENTRICLE WERE NARROWED. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6), HEIGHT:142 CM, WEIGHT: (B)(6). GENDER: FEMALE. SAME EVENT: MEDWATCH 3004721439-2021-00106.
Patient problems: Hydrocephalus
MAUDE report 11790192 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page