Skip to content

AESCULAP FV027TRecalls & reported problems

Christoph Miethke GmbH & Co. KG · BURR HOLE RESERVOIR W/CATHETER · Class 2 · Implant

All models of AESCULAP (18)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04041906116804), received between Oct 28, 2021 and Oct 28, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1
  • Material Integrity Problem1
  • Material Rupture1
  • Mechanical Problem1

Patient problems

  • Hydrocephalus1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 28, 2021Break; Mechanical Problem; Material Rupture; Material Integrity Problem; Break; Mechanical Problem; Material Rupture; Material Integrity Problem

    IT WAS REPORTED THAT A BURR HOLE RESERVOIR W/CATHETER (PART # FV027T) WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, DURING THE OPERATION, WHILE THE SURGEON ATTEMPTED TO ATTACH TO SHUNT, A LACK OF HEAT SEAL WAS NOTED ON END OF CATHETER TO RESERVOIR. SURGEON WAS THEN GOING TO REATTACH CATHETER, BUT FOUND THE INLET WAS LOOSE. THE PROCEDURE WAS THEN CONTINUED AND SUCCESSFULLY COMPLETED USING ANOTHER OF THE SAME TYPE OF DEVICE. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.

    Patient problems: Hydrocephalus

    MAUDE report 12711851 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources and updates