Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04041906117603), received between Sep 17, 2020 and Jul 3, 2024 — classified as 10 injuries.
The most frequently reported specific device problem is “Infusion or Flow Problem” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Infusion or Flow Problem8
- Complete Blockage1
- Mechanical Problem1
Patient problems
- Hydrocephalus9
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryJul 3, 2024Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PED. PRECHAMBER (#FV035T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT, THE PRECHMABER CAUSED AN OCCLUSION. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 2 YEARS, 6 MONTHS. HEIGHT: 87 CM. WEIGHT: 11 KG. GENDER: FEMALE.
Patient problems: Hydrocephalus
MAUDE report 19670536 ↗InjuryMay 1, 2024Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PEDIATRIC PRECHAMBER (#FV035T) WAS IMPLANTED DURING A PROCEDURE. ACCORDING TO THE COMPLAINANT, THE PRECHAMBER CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 5 MONTHS, HEIGHT: 56 CM, WEIGHT: 4.95 KG, GENDER: MALE. SAME EVENT AS MEDWATCH#: 3004721439-2024-00106 + MEDWATCH#: 3004721439-2024-00107.
Patient problems: Hydrocephalus
MAUDE report 19225917 ↗InjuryMar 15, 2024Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PED.PRECHAMBER (#FV035T) AND A M.BLUE (#FXT802) WERE IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT, THE PRECHAMBER CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 10 YEARS HEIGHT: 145 CM WEIGHT: 35 KG GENDER: FEMALE *SAME EVENT AS MEDWATCH#3004721439-2024-00067
Patient problems: Hydrocephalus
MAUDE report 18911942 ↗InjuryNov 1, 2023Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PAEDI PRECHAMBER (#FV035T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE SHUNT SHOWED A UNDER-DRAINAGE AND BLOCKAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2023. THE COMPLAINANT DEVICE HAS NOT YET RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 2 YEARS, 6 MONTHS. WEIGHT: 10.4 KILOGRAMS. HEIGHT: 82.7 CENTIMETERS. GENDER: MALE. SAME EVENT AS # 3004721439-2023-00311 AND # 3004721439-2023-00312.
Patient problems: Hydrocephalus
MAUDE report 18049335 ↗InjuryJul 12, 2023Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PED. PRECHAMBER (PART # FV035T) WAS BELIEVED TO BE OPERATING IN UNDERDRAINAGE. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE DYSFUNCTION. AGE: 16 YEARS, 11 MONTHS. HEIGHT: 150 CENTIMETERS (CM). WEIGHT: 40 KILOGRAMS (KG). GENDER: FEMALE.
Patient problems: Hydrocephalus
MAUDE report 17312787 ↗InjuryAug 25, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT A PRECHAMBER PED. (PART # FV035T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE VALVE WAS NOT ABLE TO BE ADJUSTED. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. PATIENT DATA: AGE: 8 MONTHS; WEIGHT: 5 KILOGRAMS (KG); HEIGHT: 60 CENTIMETERS (CM); GENDER: MALE. SAME EVENT AS MEDWATCH-NO.:3004721439-2022-00248.
Patient problems: Hydrocephalus
MAUDE report 15297352 ↗InjuryJun 22, 2022Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PEDIATRIC PRECHAMBER (#FV035T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE BLOCKED. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 6 YEARS, HEIGHT: 89 CENTIMETERS (CM), WEIGHT: 10.5 KILOGRAMS (KG), GENDER: FEMALE. SAME EVENT AS MEDWATCH-NO. 3004721439-2022-00173.
Patient problems: Hydrocephalus
MAUDE report 14775442 ↗InjuryNov 7, 2021Infusion or Flow Problem; Infusion or Flow Problem
SAME EVENT AS MEDWATCH # 3004721439-2021-00297.
Patient problems: Hydrocephalus
MAUDE report 12761680 ↗InjuryApr 7, 2021Infusion or Flow Problem; Infusion or Flow Problem
SAME PATIENT/EVENT: MEDWATCH#3004721439-2021-00064.3
Patient problems: Hydrocephalus
MAUDE report 11627494 ↗InjurySep 17, 2020Complete Blockage; Complete Blockage
THE COMPLAINANT REPORTED THAT A PROGAV 2.0 WITH SHUNT ASSISTANCE 20 (PART # FX413T) (REF. ASSOCIATED MDR ID 3004721439-2020-00182) AND PRECHAMBER (PART # FV035T) WERE IMPLANTED IN A PATIENT ON (B)(6) 2019. REPORTEDLY, CONTRAST MEDIUM WAS INJECTED INTO THE DEVICE, WHICH REVEALED THAT THE FLOW WAS BLOCKED AT THE LOCATION OF THE SHUNT VALVE. THE PATIENT UNDERWENT REVISION SURGERY ON (B)(6) 2020. NO KNOWN PATIENT COMPLICATIONS. NO FURTHER ADDITIONAL INFORMATION IS CURRENTLY AVAILABLE. ASSOCIATED MEDWATCH: 3004721439-2020-00182.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10546417 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page