Skip to content

proGAV® FV424TRecalls & reported problems

Christoph Miethke GmbH & Co. KG · PROGAV SYSTEM W/SPRUNG RESERVOIR · Class 2 · Implant

All models of proGAV (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04041906132217), received between May 13, 2021 and Apr 5, 2022 — classified as 2 injuries.

The most frequently reported specific device problem is “Infusion or Flow Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Infusion or Flow Problem2
  • Mechanical Problem2

Patient problems

  • Hydrocephalus2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 5, 2022Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV (PART # FV424T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN 2007. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN UNDERDRAINAGE AND DIFFICULT TO ADJUST. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6). HEIGHT: 180 CENTIMETERS (CM). WEIGHT: (B)(6). GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 14017421 ↗
  • InjuryMay 13, 2021Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED TO MIETHKE THAT A PROGAV SHUNTSYSTEM (PART # FV424T) AND A PROSA VALVE (PART # FV701T) (REF. MDR ID 3004721439-2021-00121) WERE IMPLANTED DURING A PROCEDURE PERFORMED IN 2010. ACCORDING TO THE COMPLAINANT, THE SHUNTSYSTEM WAS BELIEVED TO BE OPERATING IN UNDER-DRAINAGE AND AN ADJUSTMENT PROBLEM OCCURRED. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2021. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6) YEARS. WEIGHT: (B)(6) KILOGRAMS (KG). HEIGHT: 155 CENTIMETERS (CM). GENDER: MALE. THE SAME EVENT: MEDWATCH # 3004721439-2021-00121.

    Patient problems: Hydrocephalus

    MAUDE report 11816046 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources and updates