Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04041906132439), received between Jun 22, 2018 and Sep 21, 2021 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device Operates Differently Than Expected1
  • Mechanical Problem1
  • Use of Device Problem1

Patient problems

  • Hydrocephalus1
  • No Code Available1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 21, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED MIETHKE THAT A PROGAV SYSTEM W/CONTROL RESERVOIR (PART # FV431T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2016. ACCORDING TO THE COMPLAINANT, THE VALVE WAS ORIGINALLY IMPLANTED AND PROGRAMMED TO 6 MMHG WITHOUT COMPLICATION. ON (B)(6) 2016, THE VALVE WAS FOUND TO NOT BE PROGRAMMED. THE PATIENT WAS REFERRED TO THE HOSPITAL WHERE AN IMAGING EXAM WAS PERFORMED, WHICH CONFIRMED THE VALVE LOCATION. ADDITIONAL ATTEMPTS TO PROGRAM THE VALVE WERE PERFORMED, BUT WERE UNSUCCESSFUL. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2016. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6), WEIGHT: (B)(6), HEIGHT: (B)(6), GENDER: UNKNOWN.

    Patient problems: Hydrocephalus

    MAUDE report 12510324 ↗
  • MalfunctionJun 22, 2018Use of Device Problem; Device Operates Differently Than Expected

    PATIENT RECEIVED A VENTRICULOPERITONEAL SHUNT LAST YEAR. HE CAME BACK TO THE HOSPITAL WITH AMS AND CT SCAN SHOWED LARGER VENTRICLES. WE TAPPED THE SHUNT WITH NO INFECTION. WANTED TO CHANGE THE VALVE SETTING OF THE PROGRAMMABLE PROGAV VALVE TO A LOWER SETTING TO IMPROVE THE DRAINAGE, HOWEVER, WE COULD NOT CHANGE THE SETTING. PATIENT NEEDED TO BE TAKEN TO THE OR TO CHANGE THE VALVE.

    Patient problems: No Code Available

    MAUDE report 7629460 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →