Skip to content

proGAV® FV434TRecalls & reported problems

Christoph Miethke GmbH & Co. KG · PROGAV SYS W/SA 20 A.CONTROL RESERVOIR · Class 2 · Implant

All models of proGAV (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This proGAV shunt drains excess fluid from the brain. Reports mostly describe fluid draining too quickly or too slowly and valves that were hard to adjust, with patients needing another surgery to revise the shunt. There are 24 reports, with no deaths and 24 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
24
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 04041906132460), received between Jan 20, 2021 and Dec 5, 2024 — classified as 24 injuries.

The most frequently reported specific device problem is “Infusion or Flow Problem” (15 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Infusion or Flow Problem15
  • Mechanical Problem14

Patient problems

  • Hydrocephalus24

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 5, 2024Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV SHUNT SYSTEM (#FV434T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2012. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN OVERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2024. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.

    Patient problems: Hydrocephalus

    MAUDE report 20852899 ↗
  • InjurySep 3, 2024Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (#FX434T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2011. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 17 YEARS. GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 20125399 ↗
  • InjuryNov 2, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT A PROGAV SHUNTSYSTEM (#FV434T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2013. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 16 YEARS. HEIGHT: 161 CENTIMETERS. WEIGHT: 54 KILOGRAMS. GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 18055881 ↗
  • InjuryMar 9, 2023Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV SHUNT SYSTEM (#FV434T) WAS IMPLANTED DURING A PROCEDURE ON (B)(6) 2015. ACCORDING TO THE COMPLAINANT, THE PROGAV SHOWED AN UNDER-DRAINGE AND ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2023. THE COMPLAINT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6) WEIGHT: 53.8 KILOGRAMS (KG) HEIGHT: 150 CENTIMETERS (CM) GENDER: FEMALE.

    Patient problems: Hydrocephalus

    MAUDE report 16518446 ↗
  • InjuryOct 26, 2022Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV SHUNTSYSTEM (PART # FV434T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2016. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN UNDERDRAINAGE AND WAS DIFFICULT TO ADJUST. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 9 YEARS AND 2 MONTHS, HEIGHT: 130 CENTIMETERS (CM), WEIGHT: 26 KILOGRAMS (KG), GENDER: FEMALE.

    Patient problems: Hydrocephalus

    MAUDE report 15677118 ↗
  • InjuryJun 14, 2022Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV (#FV434T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2015. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN UNDERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 7 YEARS 11 MONTH, HEIGHT: 128 CM, WEIGHT: 27 KG, GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 14691206 ↗
  • InjuryApr 29, 2022Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV SHUNT SYSTEM (PART # FV434T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2012. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATING IN OVERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6). HEIGHT: 152 CENTIMETERS (CM). WEIGHT: (B)(6) . GENDER: UNKNOWN.

    Patient problems: Hydrocephalus

    MAUDE report 14251818 ↗
  • InjuryDec 13, 2021Mechanical Problem; Mechanical Problem

    ACCORDING TO THE COMPLAINANT A PROGAV WAS NOT ADJUSTABLE POSTOPERATIVELY. THE VALVE WAS IMPLANTED ON (B)(6) 2014. THE VALVE WAS EXPLANTED ON (B)(6) 2017 AND RETURNED TO THE MANUFACTURER. FURTHER DETAILS WERE NOT PROVIDED. HEIGHT: 50 CM.

    Patient problems: Hydrocephalus

    MAUDE report 12990288 ↗
  • InjuryDec 13, 2021Mechanical Problem; Mechanical Problem

    ACCORDING TO THE COMPLAINANT A PROGAV WAS NOT ADJUSTABLE POSTOPERATIVELY. THE VALVE WAS IMPLANTED ON (B)(6) 2015. THE VALVE WAS EXPLANTED ON (B)(6) 2017 AND RETURNED TO THE MANUFACTURER. FURTHER DETAILS WERE NOT PROVIDED. HEIGHT: 57 CM.

    Patient problems: Hydrocephalus

    MAUDE report 12989557 ↗
  • InjuryDec 9, 2021Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV SYS W/SA 20 A.CONTROL RESERVOIR (# FV434T) WAS IMPLANTED DURING THE PROCEDURE ON UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE NOT ADJUSTABLE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2017. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION.

    Patient problems: Hydrocephalus

    MAUDE report 12968976 ↗
  • InjuryDec 8, 2021Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED TO MIETHKE THAT A PROGAV SYS W/SA 20 A.CONTROL RESERVOIR (# FV434T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE OVERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2017. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. FURTHER DETAILS WERE NOT PROVIDED. HEIGHT: 135CM.

    Patient problems: Hydrocephalus

    MAUDE report 12964544 ↗
  • InjuryDec 3, 2021Mechanical Problem; Mechanical Problem

    ACCORDING TO THE COMPLAINANT A PROGAV HAD OVERDRAINAGE POSTOPERATIVELY AND COULD NOT BE ADJUSTED. THE VALVE WAS IMPLANTED ON AN UNSPECIFIED DATE IN 2015. IT WAS REMOVED IN 2017 DUE TO THE MALFUNCTION AND RETURNED TO THE MANUFACTURER. FURTHER DETAILS WERE NOT PROVIDED. HEIGHT: 170 CM.

    Patient problems: Hydrocephalus

    MAUDE report 12925710 ↗
  • InjuryDec 2, 2021Mechanical Problem; Mechanical Problem

    ACCORDING TO THE COMPLAINANT A PROGAV HAD REDUCED DRAINAGE POSTOPERATIVELY. THE VALVE WAS IMPLANTED ON AN UNSPECIFIED DATE. IT WAS REMOVED DUE TO THE MALFUNCTION AND RETURNED TO THE MANUFACTURER. FURTHER DETAILS WERE NOT PROVIDED.

    Patient problems: Hydrocephalus

    MAUDE report 12920223 ↗
  • InjuryDec 1, 2021Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV SHUNTSYSTEM (PART # FV434T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN JANUARY 2012. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN OVERDRAINAGE AND WAS DIFFICULT TO ADJUST. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2021. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6), HEIGHT: 150 CENTIMETERS (CM), WEIGHT: (B)(6) KILOGRAMS (KG), GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 12913758 ↗
  • InjuryNov 18, 2021Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED TO MIETHKE THAT A PROGAV SYS W/SA 20 A.CONTROL RESERVOIR (PART # FV434T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE VALVE WAS SUSPECTED OF BEING OCCLUDED APPROXIMATELY TWO (2) TO THREE (3) MONTHS POST-IMPLANTATION. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2017. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.

    Patient problems: Hydrocephalus

    MAUDE report 12836857 ↗
  • InjuryOct 18, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT A PROGAV SHUNTSYSTEM (#FV434T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN (B)(6) 2014. ACCORDING TO THE COMPLAINANT, THE PROGAV WAS BELIEVED TO BE BLOCKED. TO CHANGE THE PRESSURE SETTING WAS NOT POSSIBLE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2021. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6).

    Patient problems: Hydrocephalus

    MAUDE report 12651701 ↗
  • InjurySep 23, 2021Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV SYS W/SA 20 A.CONTROL RESERVOIR. (#FV434T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE NOT ADJUSTABLE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2016. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. FURTHER DETAILS WERE NOT PROVIDED. PATIENT INFORMATION: AGE: (B)(6).

    Patient problems: Hydrocephalus

    MAUDE report 12519836 ↗
  • InjurySep 23, 2021Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV SYS W/SA 20 A.CONTROL RESERVOIR. (#FV434T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE NOT ADJUSTABLE 1Y 4M POST OPERATIVELY. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2016. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. FURTHER DETAILS WERE NOT PROVIDED. PATIENT INFORMATION: AGE: (B)(6) YEARS.

    Patient problems: Hydrocephalus

    MAUDE report 12519831 ↗
  • InjurySep 13, 2021Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED MIETHKE THAT A PROGAV SYS W/SA 20 A.CONTROL RESERVOIR (PART # FV434T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2016. ACCORDING TO THE COMPLAINANT, THE VALVE WAS OCCLUDED. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2016. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.

    Patient problems: Hydrocephalus

    MAUDE report 12465347 ↗
  • InjurySep 13, 2021Mechanical Problem; Mechanical Problem

    ACCORDING TO THE COMPLAINANT A PROGAV WAS NOT ADJUSTABLE POSTOPERATIVELY. THE PATIENT WAS ORIGINALLY IMPLANTED IN 2008. THE EXPLANT DATE WAS ON (B)(6) 2016. THE VALVE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. FURTHER DETAILS WERE NOT PROVIDED.

    Patient problems: Hydrocephalus

    MAUDE report 12463632 ↗
  • InjurySep 3, 2021Mechanical Problem; Mechanical Problem

    ACCORDING TO THE COMPLAINANT A PROGAV WAS SUSPECTED OF UNDER OR OVER-DRAINAGE POSTOPERATIVELY. THE PATIENT WAS ORIGINALLY IMPLANTED ON (B)(6) 2015. TLHE VALVE WAS EXPLANTED ON (B)(6) 2016 AND RETURNED TO THE MANUFACTURER FOR EVALUATION. FURTHER DETAILS WERE NOT PROVIDED. PATIENT SYMPTOMS DUE TO UNDER-DRAINAGE WERE NOT NOTED. HEIGHT: 150 CM.

    Patient problems: Hydrocephalus

    MAUDE report 12421549 ↗
  • InjuryAug 19, 2021Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV SHUNTSYSTEM (PART # FV434T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2013. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO HAVE BEEN OPERATING IN OVERDRAINAGE AND HAD A PROBLEM WITH ADJUSTABILITY. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2021. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6). HEIGHT: 143 CENTIMETERS (CM). WEIGHT: (B)(6). GENDER: FEMALE.

    Patient problems: Hydrocephalus

    MAUDE report 12341815 ↗
  • InjuryMay 18, 2021Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV SHUNTSYSTEM WAS IMPLANTED DURING A PROCEDURE PERFORMED IN (B)(6) 2012. ACCORDING TO THE COMPLAINANT, THE GRAVITATIONAL UNIT WAS BELIEVED TO BE OPERATED IN UNDER-DRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6) WEIGHT: (B)(6) HEIGHT: 180 CM GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 11841495 ↗
  • InjuryJan 20, 2021Mechanical Problem; Mechanical Problem

    INVESTIGATION IS COMPLETED THE RESULT WILL BE SUBMITTED WITH THIS FOLLOW UP REPORT.

    Patient problems: Hydrocephalus

    MAUDE report 11202285 ↗

Other Shunt, central nervous system and components devices